Published online by Cambridge University Press: 17 May 2021
CyberKnife is the most advanced form of stereotactic body radiotherapy (SBRT) system that uses a robotic arm to deliver highly focused beams of radiation; however, a limitation is that it only irradiates from ceiling to floor direction. In patients with posterior lungs tumour who are positioned supine, normal lung tissue may suffer undesirable radiation injuries. This study compares the treatment planning between the prone set-up and the supine set-up for lung cancer in CyberKnife SBRT to decrease normal lung dose to avoid radiation side effects.
A human phantom was used to generate 108 plans (54 for prone and 54 for supine) using the CyberKnife planning platform. The supine and prone plans were compared in terms of the dosimetric characteristics, delivery efficiency and plan efficiency.
For posterior targets, the area of low-dose exposure to normal lungs was smaller in the prone set-up than in the supine set-up. V10 of the lungs was 7·53% and 10·47% (p < 0·001) in the anterior region, and 10·78% and 8·03% (p < 0·001) in the posterior region in the supine and prone set-up plans, respectively.
The comparison between the prone set-up and the supine set-up was investigated with regard to target coverage and dose to organs at risk. Our results may be deployed in CyberKnife treatment planning to monitor normal tissue dose by considering patient positioning. This may assist in the design of better treatment plans and prevention of symptomatic radiation pneumonitis in lung cancer patients.