Nosocomial Outbreak of Pseudomonas cepacia Associated with Contamination of Reusable Electronic Ventilator Temperature Probes

Published online by Cambridge University Press:  21 June 2016

Abstract

Objective:

The investigation and control of an outbreak of nosocomial Pseudomonas cepacia respiratory tract colonization and infection.

Design:

Epidemiologic investigation based on infection control surveillance data, including definition and characterization of case patients, environmental cultures based on epidemiologic information, and institution of control measures based on study results.

Setting:

A 1,171 -bed, university-affiliated tertiary care hospital.

Results:

Between January 1, 1988, and June 30, 1989, 127 patients were culture-positive for P cepacia, 117 (92%) of whom were culture-positive from sputum and were treated in the intensive care unit. Review of respiratory care procedures revealed that when mechanical ventilators were serviced between patients, the electronic temperature probes used with servo-controlled humidifiers were wiped with the same odor-counteractant cleaning solution used on ventilator cabinets, P cepacia was isolated from nine of 12 in-use temperature probe tips, three of which were from patients with negative sputum cultures for P cepacia, one of whom subsequently developed culture positivity for P cepacia.

P cepacia also was isolated from the diluted odor-counteractant solution. Following the institution of a disinfection procedure for temperature probe tips, the incidence of P cepacia sputum culture positivity in ICU patients fell significantly compared to the outbreak period (138 of 5,225 discharges versus 52 of 3,678 discharges, P<0.01).

Conclusions:

This investigation identified contamination of reusable electronic temperature probes as a source of nosocomial respiratory infection due to P cepacia and emphasizes the need to carefully evaluate disinfection practices for reusable patient care equipment.

Type
Original Articles
Copyright
Copyright © The Society for Healthcare Epidemiology of America 1993 

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Craven, DE, Steger, KA. Nosocomial pneumonia in the intubated patient. New concepts on pathogenesis and prevention. Infect Dis Clin North Am 1989;3:843866.10.1016/S0891-5520(20)30310-XCrossRefGoogle Scholar
2. Goldmann, DA, Klinger, JD. Pseudomonas cepacia: Biology, mechanisms of virulence, epidemiology. J Pediatr 1986;108:806812.10.1016/S0022-3476(86)80749-1CrossRefGoogle ScholarPubMed
3. Gamer, JS, Favero, MS. Guideline for handwashing and hospital environmental control, 1985. Am J Infect Control 1986;14:110126.Google Scholar
4. Kahn, HA, Sempos, CT Statistical methods in epidemiology. In: MacMahon, B, ed. Monographs in Epidemiology and Biostatistics. Vol. 12. New York, NY: Oxford University Press; 1989:100102.Google Scholar
5. Conly, JM, Hass, L, Larson, L, et al. Pseudomonas cepacia colonization and infection in intensive care units. Can Med Assoc J 1986;134:363366.Google ScholarPubMed
6. Simmons, BP, Wong, ES. Guidelines for prevention of nosocomial pneumonia. Am J Infect Control 1983;11:230243.10.1016/0196-6553(83)90006-8CrossRefGoogle Scholar