Hostname: page-component-586b7cd67f-t8hqh Total loading time: 0 Render date: 2024-11-24T07:57:19.418Z Has data issue: false hasContentIssue false

Embolisation of ceramic-coated PDA devices into the descending thoracic aorta: probable mechanisms and retrieval strategies

Published online by Cambridge University Press:  06 June 2019

Supratim Sen*
Affiliation:
Department of Pediatric Cardiology, SRCC Children’s Hospital, Mumbai, India
Sneha Jain
Affiliation:
Department of Pediatric Cardiology, SRCC Children’s Hospital, Mumbai, India
Bharat Dalvi
Affiliation:
Department of Pediatric Cardiology, SRCC Children’s Hospital, Mumbai, India Glenmark Cardiac Center, Mumbai, India
*
Author for correspondence: Supratim Sen, Department of Pediatric Cardiology, SRCC Children’s Hospital, 1-1A Haji Ali Park, Mahalaxmi, Mumbai 400034, India. Mobile: +91 8080329189; E-mail: [email protected]

Abstract

We describe two cases of spontaneous embolisation and successful retrieval of ceramic-coated patent arterial duct devices. In both, the device embolised to the descending aorta in the absence of pulmonary hypertension and despite optimum placement. We have discussed possible mechanisms for embolisation in these patients and suggested alternative methods for device retrieval. Based on this limited experience, we conclude that for tubular ducts, ceramic-coated devices should be oversized to form a tighter waist or alternate devices may be considered.

Type
Brief Report
Copyright
© Cambridge University Press 2019 

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Boehm, W, Emmel, M, Sreeram, N. The Amplatzer duct occluder for PDA closure: indications, technique of implantation and clinical outcome. Images Paediatr Cardiol 2007; 9: 1626.Google ScholarPubMed
Chamié, F, Simões, LC, Chamié de Queiroz, DS, Mattos, R. Percutaneous closure of a patent ductus arteriosus with the Cera™ PDA Occluder: another good option in the toolbox. Rev Bras Cardiol 2012; 20: 7781.Google Scholar
Bilkis, AA, Alwi, M, Hasri, S, et al. The Amplatzer duct occluder: experience in 209 patients. JACC 2001; 37: 258261.CrossRefGoogle ScholarPubMed
El-Said, HG, Bratincsak, A, Foerster, SR, et al. Safety of percutaneous patent ductus arteriosus closure: an unselected multicenter population experience. J Am Heart Assoc: Cardiovasc Cerebrovasc Dis 2013; 2: e000424.CrossRefGoogle ScholarPubMed
Backes, CH, Rivera, BK, Bridge, JA, et al. Percutaneous Patent Ductus Arteriosus (PDA) closure during infancy: a meta-analysis. Pediatrics 2017; 139: e20162927 CrossRefGoogle ScholarPubMed
McMullan, DM, Moulick, A, Jonas, RA. Late embolization of Amplatzer patent ductus arteriosus occlusion device with thoracic aorta embedment. Ann Thorac Surge 2007; 83: 11771179.CrossRefGoogle ScholarPubMed
Vijayalakshmi, IB, Setty, N, Narasimhan, C, Singla, V, Manjunath, CN. Percutaneous device closure of patent ductus arteriosus with pulmonary artery hypertension: long-term results. J Interv Cardiol 2014; 27: 563569.CrossRefGoogle ScholarPubMed
Khan, MA, Almashham, YH, Almoukirish, ARS, Momenah, TS. Embolized Amplatzer duct occluder to aorta: retrieval technique. J Saudi Heart Assoc 2016; 28: 116118.10.1016/j.jsha.2015.11.002CrossRefGoogle ScholarPubMed
Koneti, NR, Bakhru, S, Penumatsa, RR, Lalukota, KM. Rescrewing the embolized duct occluder using the delivery cable. Ann Pediatr Card 2014; 7: 103106.CrossRefGoogle ScholarPubMed
Supplementary material: Image

Sen et al. supplementary material

Sen et al. supplementary material 1

Download Sen et al. supplementary material(Image)
Image 1.1 MB