Implementation of multidrug-resistant bacterial testing to prioritize duodenoscope sterilization: Experience from a high-volume health system

Published online by Cambridge University Press:  10 November 2023

Jad AbiMansour
Affiliation:
Division of Gastroenterology and Hepatology, Mayo Clinic, Rochester, Minnesota
John A. Martin
Affiliation:
Division of Gastroenterology and Hepatology, Mayo Clinic, Rochester, Minnesota
Bret T. Petersen*
Affiliation:
Division of Gastroenterology and Hepatology, Mayo Clinic, Rochester, Minnesota
*
Corresponding author: Bret T. Petersen; Email: [email protected]
Rights & Permissions [Opens in a new window]

Abstract

Type
Research Brief
Copyright
© The Author(s), 2023. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America

Reusable duodenoscopes are used to perform >650,000 endoscopic retrograde cholangiopancreatography (ERCP) procedures annually in the United States. Reference Yachimski and Ross1 These devices contain small working parts, which makes cleaning and disinfection challenging compared to other devices. Multiple outbreaks have been reported over the past decade, most commonly from multidrug-resistant organisms (MDROs), prompting numerous safety communications and mandates for instrument redesign. Reference Heuvelmans, Wunderink, van der Mei and Monkelbaan2,3 Device sterilization, often performed with ethylene oxide gas (ETO), is the highest level of disinfection available. Data are limited regarding the incremental benefit of sterilization, but it remains the gold standard for disinfection. Reference Day, Muthusamy and Collins4,Reference Snyder, Wright and Smithey5 However, due to cost, environmental impact, need for specialized facilities, and supply shortages, ETO is not used routinely for reprocessing. In this report, we describe the feasibility, implementation, and impact of a systematic testing protocol among patients undergoing ERCP to detect MDROs in duodenoscope reprocessing.

We performed a retrospective review of all patients who underwent screening for MDROs via polymerase chain reaction (PCR) testing of rectal swabs at the time of ERCP (TJF-Q180V duodenoscope, Olympus, Tokyo, Japan) between January 2018 and May 2022 at a single healthcare system, including a tertiary-care center and community practices. Beginning in 2018, PCR testing of rectal swabs was routinely performed at 2 community practices and at the tertiary-care center when daily sterilization capacity was exceeded during a national shortage of ETO in 2022. PCR targets included oxacillin-hydrolyzing β-lactamase (Oxa-48), Klebsiella pneumoniae carbapenemase (KPC), New Delhi metallo-β-lactamase (NDM), and Verona metallo-β-lactamase (VIM). All duodenoscopes underwent double high-level disinfection [ie, wash plus high-level disinfection (HLD), followed by wash plus HLD], with subsequent diversion to ETO sterilization if point-of-care testing returned positive (Fig. 1). HLD was performed on an automated processor with a peracetic acid–based disinfectant (Rapicide PA, Steris, Mentor, OH). Patients were monitored via retrospective chart review and hospital-wide surveillance and reporting systems for MDRO infections.

Figure 1. Schematic outlining MDR testing paradigm to prioritize duodenoscopes for sterilization with ethylene oxide.

In total, 557 patients underwent testing, and only 1 (0.2%) result was positive. However, 2 (0.4%) tests were processed incorrectly, prompting precautionary sterilization. All duodenoscopes used in patients with negative PCR tests were reprocessed with double washing and high-level disinfection. No patients developed a healthcare-associated infection or an MDR-related disease.

The implementation of this protocol maintained procedural throughput while minimizing MDRO transmission risk. We utilized established mechanisms and infrastructure for PCR testing at our institution, which generally yielded results in 12 hours at the tertiary-care center and within 24–48 hours for all sites. Point-of-care testing for similar applications has been used to provide rapid detection of MDROs within 1 hour, allowing for even more rapid triage of high-risk devices. This technology was described in a series of 201 patients, among whom carbapenemase resistance was detected in 0.5% of patients. Reference Pannala, Baldwin and Aluru6 Another study reported a positive or indeterminate rectal swab in 0.6% of patients. Reference Smith, Dua and Saeian7 Alternative approaches, including routine duodenoscope cultures, are more costly and have significant turnaround times. A financial analysis determined that a culture-based approach would only be cost-effective if the probability of MDRO infection rose to >24%. Reference Almario, May and Shaheen8 The rate of colonization reported here is consistent with previous surveillance data reporting CRE rates <1.1%. Reference Ma, Pegues and Kochman9 The low colonization rate would challenge the cost efficacy of any universal screening program. However, this same approach could be targeted to high-risk populations such as inpatients, prolonged hospitalizations, antibiotic exposure, nursing home residents, and outbreak scenarios.

MDR pathogens are not the only microorganisms transmitted through reusable endoscopes, and they are likely overrepresented given established hospital surveillance programs. However, outbreaks of MDROs have been closely associated with duodenoscopes despite following disinfection protocols and carry significant clinical implications. This characteristic has made them particularly concerning to regulatory agencies, manufacturers, and professional organizations. There has been significant advancement in the development of single-use duodenoscopes, which are now available from several manufacturers, and in reuseable duodenoscopes with single-use components, which are the only reusable versions available for purchase in the United States. Data assessing the impact of these innovations are very limited. One study evaluated contamination in duodenoscopes reprocessed with or without a removeable cap in place and showed significantly lower ATP activity in the end cap–detached group. However, it remains unclear how this surrogate outcome translates to transmissibility. Reference Ridtitid, Pakvisal and Chatsuwan10 Although these devices have the potential to reduce the risk of pathogen transmission, many questions remain related to safety, implementation strategy, cost, and environmental impact. Reference Trindade, Copland and Bhatt11 The sole use of these devices will make economics of building and maintaining infrastructure to perform ETO sterilization even less favorable. Utilizing the method outlined in this report may help further streamline and focus the sterilization process to cases with the greatest risk of pathogen transmission. Innovations in device cleaning and reprocessing are likely to be as important as duodenoscope design.

This study had several limitations. This study was retrospective in design, with a relatively small sample size and a low number of events. The study cohort originated from a single institution with an established processes in place to perform rapid PCR testing. Institutions without this infrastructure would face upfront costs not well described in this report.

In summary, this point-of-care testing model was efficient and feasible and may help optimize resource utilization while minimizing the risk of interpatient pathogen transmission. This approach to device reprocessing was successfully implemented across tertiary and regional community sites. Although additional studies are needed to evaluate the model’s cost-effectiveness, environmental impact, and performance in a broader patient population, it provides a robust framework that could be implemented easily in similar healthcare systems.

Acknowledgments

Financial support

No financial support was provided relevant to this article.

Competing interests

J.P.A. has a licensing agreement with Ruhoff Corporation. B.T.P. receives research support from Ambu and Boston Scientific and is a consultant for Olympus America. J.A.M. has no potential conflicts to disclose.

References

Yachimski, PS, Ross, A. The future of endoscopic retrograde cholangiopancreatography. Gastroenterology 2017;153:338344.CrossRefGoogle ScholarPubMed
Heuvelmans, M, Wunderink, HF, van der Mei, HC, Monkelbaan, JF. A narrative review on current duodenoscope reprocessing techniques and novel developments. Antimicrob Resist Infect Control 2021;10:171.CrossRefGoogle ScholarPubMed
FDA safety communication. Use duodenoscopes with innovative designs to enhance safety. US Food and Drug Administration website. https://www.fda.gov/medical-devices/safety-communications/use-duodenoscopes-innovative-designs-enhance-safety-fda-safety-communication. Published online June 29, 2022. Accessed October 30, 2023.Google Scholar
Day, LW, Muthusamy, VR, Collins, J, et al. Multisociety guideline on reprocessing flexible GI endoscopes and accessories. Gastrointestinal Endoscopy 2021;93:1133.CrossRefGoogle Scholar
Snyder, GM, Wright, SB, Smithey, A, et al. Randomized comparison of 3 high-level disinfection and sterilization procedures for duodenoscopes. Gastroenterology 2017;153:10181025.CrossRefGoogle ScholarPubMed
Pannala, R, Baldwin, B, Aluru, V, et al. Prospective study of the feasibility of point-of-care testing strategy for carbapenem-resistant organism detection. Endosc Int Open 2018;06:E58E63.Google Scholar
Smith, ZL, Dua, A, Saeian, K, et al. A novel protocol obviates endoscope sampling for carbapenem-resistant Enterobacteriaceae: experience of a center with a prior outbreak. Dig Dis Sci 2017;62:31003109.CrossRefGoogle ScholarPubMed
Almario, CV, May, FP, Shaheen, NJ, et al. Cost utility of competing strategies to prevent endoscopic transmission of carbapenem-resistant Enterobacteriaceae. Am J Gastroenterol 2015;110:16661674.CrossRefGoogle ScholarPubMed
Ma, GK, Pegues, DA, Kochman, ML, et al. Implementation of a systematic culturing program to monitor the efficacy of endoscope reprocessing: outcomes and costs. Gastrointestinal Endoscopy 2018;87:104109.CrossRefGoogle ScholarPubMed
Ridtitid, W, Pakvisal, P, Chatsuwan, T, et al. A newly designed duodenoscope with detachable distal cap significantly reduces organic residue contamination after reprocessing. Endoscopy 2020;52:754760.Google ScholarPubMed
Trindade, AJ, Copland, A, Bhatt, A, et al. Single-use duodenoscopes and duodenoscopes with disposable end caps. Gastrointestinal Endoscopy 2021;93:9971005.CrossRefGoogle ScholarPubMed
Figure 0

Figure 1. Schematic outlining MDR testing paradigm to prioritize duodenoscopes for sterilization with ethylene oxide.