Hostname: page-component-586b7cd67f-rcrh6 Total loading time: 0 Render date: 2024-11-24T04:05:10.745Z Has data issue: false hasContentIssue false

A HOMEMADE SEMIAUTOMATIC GRAPHITIZATION DEVICE FOR AMS 14C DATING AT NTUAMS LAB

Published online by Cambridge University Press:  29 April 2024

Tzu-Tsen Shen
Affiliation:
Department of Geosciences, National Taiwan University, Taipei 10617, Taiwan Instrumentation Center, National Taiwan University, Taipei 10617, Taiwan
Hong-Chun Li*
Affiliation:
Department of Geosciences, National Taiwan University, Taipei 10617, Taiwan
Rick Qiu
Affiliation:
Department of Geosciences, National Taiwan University, Taipei 10617, Taiwan
*
*Corresponding author. Email: [email protected]
Rights & Permissions [Opens in a new window]

Abstract

A low-cost and computer-controlled graphitization system connected to an elemental analyzer (EA) has been designed and built at the NTUAMS Lab. This semiautomatic system equips 6-unit reactors for the graphitization of CO2 with H2 on the iron catalyst. The entire procedure takes about 7 hours for iron conditioning, sample combustion and loading, and graphitization. The system can produce good-quality graphite for samples containing 0.5–1.6 mg carbon mass, with the pressure yield of graphitization ranging from 57.7% to 87.1%. The average values of OXI and OXII agree well with the consensus value, but the result of ANU sucrose was observed to be slightly higher than the reported one. The background samples of anthracite over ten months yielded an average of 0.38±0.10 pMC (n=21) corresponding to a 14C age of 45 kyr BP. Intercomparison samples L and M of FIRI exhibit that the measured 14C ages are almost identical to the consensus values and have a small spread in these values. The system has been carrying out graphitization for total organic carbon (TOC) of peat samples, and providing a more efficient and convenient way for AMS 14C dating.

Type
Conference Paper
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2024. Published by Cambridge University Press on behalf of University of Arizona

INTRODUCTION

The HVEE 1.0 MV Tandetron Model 4110 BO-accelerate mass spectrometry (AMS) for radiocarbon (14C) dating has been performed at the NTUAMS Lab of National Taiwan University (NTU) since 2013. To eliminate Li2 ++ interference, the AMS measures 14C3+ ions with a transmission rate of ∼18%. Every sample measurement is set up for four cycles and analyzed for 1500 seconds in every cycle, so that measurement time of a sample is normally 6000 seconds in total. During the analysis for every sample, the beam current of 12C3+ should remain steady and higher than 1.0 μA. Finally, the weighted mean and uncertainty values of 14C/12C and 13C/12C ratios are taken to calculate the sample age and the associated error (Li et al. Reference Li, Chang, Berelson, Zhao, Misra and Shen2022).

In order to convert sample CO2 into graphite, three vacuum lines for CO2 extraction and purification were installed with the TiH2/Zn reduction method. At first, the quartz tube is purified in the vacuumed system under a pressure of 10-5–10-6 mbar. The organic sample reacts with CuO and Ag in the evacuated quartz tube at 850°C for 6 hours to acquire CO2 gas (Barile et al. Reference Barile, Barone, Fedi, Liccioli, Schiavulli, Casamassima, Valentini, Pastore and Paticchio2019; Li et al. Reference Li, Chang, Berelson, Zhao, Misra and Shen2022). The produced CO2 in the quartz tube is purified cryogenically on the vacuum line. The next step is to transfer and seal the pure CO2 into a combination glass tube (Xu et al. Reference Xu, Trumbore, Zheng, Southon, McDuffee, Luttgen and Liu2007). The combination glass tube contains a 9-mm diameter 6-inch long tube in which the reagents of TiH2 and Zn exist at the bottom, with a small 6-mm diameter 2-inch long tube inside which has iron powder/catalyst. The CO2 reduction to graphite is carried out for 6 hours in a muffle furnace at 550°C. Finally, the graphite is pressed into a target holder and measured with the AMS for 14C dating. It is a highly time-consuming experiment process and a sample takes four separate days.

It is also widely applicable that iron powder is used to catalyze graphite production from CO2 with H2 (Vogel et al. Reference Vogel, Southon, Nelson and Brown1984). Moreover, the CO2 gas from organic sample combustion using an EA, and then coupled to the graphite reduction system is developed universally (Aerts-Bijma et al. Reference Aerts-Bijma, van der Plicht and Meijer2001; Wacker et al. Reference Wacker, Němec and Bourquin2010). Many AMS 14C labs are equipped with automatic graphitization system (e.g., AGE by Ionplus). However, the cost of such a system is a concern for our lab. Here we present the first establishment of a graphitization system with the H2 reduction method and its operating results from the past a year or so.

Configuration

Our homemade graphitization system assembled by NTUAMS Lab has a compact design at a low cost (US$27,000 not including the EA). All devices, including six reactors with oven, vacuum lines, turbo pumps, electronics components, etc., are built up on a 3-tier wheeled rack (500×900×1100 mm) (Figure 1). The system is connected with a EURO EA made in Germany, and also CO2 can be brought into the system from an independent gas inlet. Six reactors are aligned parallel and each one is comprised of a “glass tube” with iron powder placed at the bottom (Figure 1c-10), a “reused tube” for the water trap (Figure 1b-6), and a DCbox BVS-N11 pressure transducer (0–2000 mbar). Two tubes are made of SIMAX 70 mm long tube with 6 mm outer diameter (OD), and 4 mm inner diameter (ID). The volume of each reactor including two tubes is estimated at about 5.2 mL based on the ideal gas law. The “glass tube” is horizontally placed in an oven (Figure 1c-9), and the “reused tube” is vertically connected with the “glass tube” for purification purpose. A temperature sensor is in the oven to record the actual temperature during the reaction process. A vacuum line is built up from 1/4” OD 316L stainless steel and connected to the reactors with FITOK® pneumatic diaphragm valves. The vacuum system contains two low vacuum stations with a Pfeiffer Vacuum MVP 015-4 diaphragm pump and an EDWARDS nXDS10i dry scroll pump, and a high vacuum pumping system using an EDWARDS nExt240 turbo molecular pump. The pressure measurement is read by an EDWARDS Wide Range Gauge that offers the capability range from the atmosphere to 10–8 mbar. The background pressure of the graphitization system is achieved around low 10–7 mbar. The H2-injected flow is controlled by Swagelok NUPRO SS-4BMG-VCR Metering Bellows Sealed valve. All valves as well as temperature and pressure electronic sensors are automatically controlled and monitored by HAPS (HQT Auto Process System) program that is commercially available from HQT Inc.

Figure 1 Photographs of the semiautomatic graphitization system and its devices. (a) The vacuum system included a diaphragm pump (1), dry scroll pump (2), and one high vacuum turbo molecular pump (3). (b) Picture shows the components of temperature electrics (4), CO2 trap (5), reused tube (6), and computer for monitoring and controlling (7). (c) Picture shows EA (8) connected with the graphitization system which is comprised of six-unit reactors. Each reactor has an oven (9), a glass tube (10), and a pressure transducer (11). Gas inlet of CO2 (12) and H2 (13). (d) Diagram shows the schematic graphitization system.

GRAPHITIZATION PROCEDURES

Iron Conditioning

The iron catalyst (-325 mesh, Alfa Aesar®) is weighed in a range of 3–4 mg and then placed into the bottom of a “glass tube”. At first, the iron powder is heated with air for 150 s at 500°C; afterwards wasted gases are pumped away until the system reached below 35°C. Then, to inject pure H2 with 800 mbar into the reactor, baking at 500°C for 30 min. The process would be effective in activating the catalyst to remove any present carbon-bearing contaminations and iron oxide by reduction mechanism (Němec et al. Reference Němec, Wacker and Gäggeler2010; Wacker et al. Reference Wacker, Němec and Bourquin2010).

Sample Preparation and Combustion

In order to test the new H2 reduction graphitization system, the organic carbon background made from anthracite (BKG), oxalic acid standards (OXII, NIST SRM 4990C), and intercomparison samples L and M of FIRI (distributed by the University of Glasgow during the fourth international radiocarbon intercomparison) are adopted. All samples containing about 0.3–1.5 mg of carbon mass were wrapped in tin capsules (6×4 mm, OEA labs, United Kingdom). The wrapped samples are combusted using the EURO EA in CNH mode. Pure helium is used as the carrier gas in the graphitization system to avoid air in the system. The temperature of 1030°C in the front furnace and oxygen injection for 8.7 s are set up to ensure complete decomposition of the samples. Excess oxygen is consumed in the reduction column with copper at 650°C.

Loading CO2 Sample into the Reactor

The combustion gases, such as N2, CO2 and H2O, are produced by EA and flush along with helium carrier flow into the CO2 trap of the graphitization system. The abundant nitrogen is pumped away and the resultant CO2 and H2O are trapped with liquid nitrogen (LN2) cooled trap. Next step, the dry ice/ethanol slurry with a temperature of about –60°C replaces LN2 to keep the trapped H2O frozen. Finally, the pure CO2 is transferred into the “reused tube” of a selected reactor cooled by LN2. The pressure of CO2 sample is recorded when the CO2 is filled into the entire reactor at room temperature. After the reactor is cooled again with LN2 to freeze CO2 sample, inject H2 of about 2.4 times the recorded CO2 pressure.

The CO2 trap is utilized for transferring CO2 into six reactors. To prevent cross contamination, the CO2 trap is heated to 120°C before processing the next sample. The system is then evacuated by the high vacuum pump for 10 min to ensure that the graphitization system is clean. To examine the memory effect, two sample batches, including 6 BKG samples and 6 OXII standards, were tested in July 2022. The arrangement was that every OXII standard was followed by a BKG sample. If the cleaning procedure is proper, the background results should not be affected by OXII (modern carbon). Such a test shows that the graphite samples produced from BKG samples displayed a low background with an average of 0.49±0.22 pMC, indicating that the former sample was unable to pollute the next one. After all of six reactors are loaded with CO2 and H2, the system will process graphitization.

Graphitization and Its Pressure Yield (%)

After loading the CO2 sample into the reactor, the graphite reduction is carried out at the temperature of 550°C for 3 hours. Meanwhile, the “reused tube” is immersed in the dry ice/ethanol slurry at about –60°C to keep the reduction product water frozen and ensure the good efficiency of the reduction process. Once the reduction time is finished, the ovens are turned off and the system is evacuated automatically. The temperature and pressure of the six reactors are respectively monitored and logged data once in 2 seconds. For each reactor, CO2 pressure of about 390 mbar corresponds to 1 mg carbon mass. Figure 2 displays a case of the temperature-pressure-time record for 0.36–1.55 mg carbon samples during the graphitization. All pressure lines decreased gradually and then stabilized within about 2.5 hours, indicating that three hours of is sufficient for completed graphitization.

Figure 2 Reduction performance of the graphitization system. Samples with carbon weight of 0.36–1.55 mg show that the graphitization reaction can be finished within 3 hours, but samples with carbon weight less than 0.36 mg are unable to provide reliable graphite due to low yield.

The pressure yield (%) is an indicator of the efficiency of graphitization suggested by Rinyu et al. (Reference Rinyu, Futó, Kiss, Molnár, Svingor, Quarta and Calcagnile2007). The case results (Figure 2) showed that the pressure yield is estimated to be 40.7% when the sample contains only 0.36 mg carbon. The produced graphite has an unacceptable 12C3+ current and a relatively large error in 14C counts. In contrast, the 0.51–1.55 mg carbon samples which have better pressure yield in a range of 57.7–87.1%. The graphite can present an optimal 12C3+ current of 1.59–5.47 µA and provides reliable statistical results for AMS measurement. Therefore, our H2 reduction graphitization system is appropriate for samples containing a range of 0.5–1.6 mg carbon and capable of obtaining good quality graphite.

RESULTS AND DISCUSSION

Uncertainty Test

To ensure the reliability of the graphitization system, standards of 6 OXII, 4 OXI, and 2 ANU sucrose were processed by our new device and carried out in the same wheel for AMS measurement. The average values of OXIs and ANUs are 104.47 pMC and 153.18 pMC, respectively, when using OXII results as a primary standard. The measured value of OXI is in good agreement with the accepted values, but the ANU result is slightly higher than the consensus value of 150.61 pMC (Rozanski et al. Reference Rozanski, Stichler, Gonfiantini, Scott, Beukens, Kromer and van der Plicht1992). The activity ratio of OXII/OXI for our measurement is 1.2833±0.0009 which is close to the reported ratio of 1.2933±0.0004 (Stuiver, Reference Stuiver1983).

Background

From December 2021 to October 2022, a total of 32 BKG samples were run by the new H2-reduced graphitization system. The results showed a declining trend in pMC values over time (Figure 3b). At the beginning of the test (before March 2022), the samples showed higher values with an average of 2.05±1.71 pMC (n=7). It is mainly attributed to the weak 12C3+ current (below 1 µA) due to the poor quality of graphite (Figure 3a). Those graphite samples contained less than 0.4 mg of carbon mass. After the initial test, 25 BKG samples generated relatively lower values, ranging from 0.86 pMC to 0.22 pMC with an average of 0.44±0.18 pMC. The 12C3+ current of those samples is higher than 1.0 µA resulted in a relatively small statistical error, representing good quality graphite. The background samples with the carbon mass of 0.5–1.6 mg yielded an average of 0.38±0.10 pMC, which implies that our new H2 reduction system is able to date samples within 14C age of 45,000 yr BP.

Figure 3 (a) AMS 12C3+ currents of the graphite, produced by the H2 reduction method, from the anthracite background (BKG) samples. (b) The pMC results using two reduction methods are shown in different colors. The higher pMC values are due to the weak 12C3+ current. The blank of the AMS is an average of ∼0.08 pMC.

For comparison, 74 BKG samples were produced to graphite through the TiH2/Zn reduction method during the same period in the NTUAMS Lab. The results showed a stable range with an average of 0.43±0.25 pMC, which is equivalent to a 14C age of 44,000 yr BP (Figure 3b). The comparable results from two different reduction methods suggest that the background test of our new H2 reduction graphitization system is reliable.

OXII

Figure 4 shows the comparison of the graphite of OXII standards produced using H2 (red color) and TiH2/Zn (black color) reduction methods respectively. The produced graphite of 17 OXII standards by the H2 reduction method, ranged from 127.98 pMC to 138.38 pMC with a relative standard deviation of 2.57% (Figure 4a). For the results of the TiH2/Zn reduction method, 39 OXII standards were graphited and showed between 124.6 pMC and 147.48 pMC with a scatter of 4.61%. Besides, the isotope correction using measured AMS δ13C, showing the δ13C of –21.57±0.26‰ for H2 reduction method as well as –21.09±0.50‰ for TiH2/Zn reduction method (Figure 4b). The δ13C results of H2-reduced graphite are slightly lighter and less scattered than those of TiH2/Zn reduction graphite. It indicates that the carbon isotope would occur differently fractionated during the graphitization reaction due to the different reduction processes. The degree of fractionation is associated with the carbon content in a sample, the amounts of reagents for reduction, temperature, and time of the process (Xu et al. Reference Xu, Trumbore, Zheng, Southon, McDuffee, Luttgen and Liu2007; Marzaioli et al. Reference Marzaioli, Borriello, Passariello, Lubritto, Cesare, D’Onofrio and Terrasi2008; Macario et al. Reference Macario, Alves, Moreira, Oliveira, Chanca, Jou and Diaz2017). The similar observation regarding the lighter δ13C graphite generated by the zinc reduction was also documented by Xu et al. (Reference Xu, Trumbore, Zheng, Southon, McDuffee, Luttgen and Liu2007). Although the corrected δ13C from both methods are lighter than OXII standard (–17.8‰), the constant AMS δ13C over a range suggested that the fractionation effect can be evaluated as long as the samples are processed in the steady condition.

Figure 4 (a) The pMC results from the OXII standards using H2 and TiH2/Zn reduction methods respectively. The pMC results of OXII standards were subtracted by a value of AMS background (0.08 pMC). (b) The difference in δ13C suggests that the carbon isotope fractionation during the graphitization resulted from the different reduction methods. The isotope fractionation using AMS δ13C corrected to –17.8‰.

Intercomparison Sample

In this study, the sample M and sample L of the fourth international radiocarbon intercomparison (FIRI) were adopted for graphitization system testing. FIRI-M is a whole peat sample from a coastal cliff deposit at St Bees Head in northwestern England which had been exposed by erosion. The consensus 14C age is 11139±49 yr BP (Naysmith et al. Reference Naysmith, Scott, Dunbar and Cook2019). FIRI-L is a wood sample from the burial mound of Dogee Barrow, grave 8 (the Tuva king barrows from Scythia), and its reported 14C value is 2505±39 yr BP (Scott et al. Reference Scott, Cook, Naysmith and Staff2019).

The initial test result on March 2022, three measured 14C age of sample FIRI-M was in an average of 11711±233 yr BP which has a large variation to the reported age (Table 1). Those samples produced poor-quality graphite with a weak 12C3+ current below 1.0 μA. Two measurements in August of 2022 were significantly improved, yielding an average 14C age of 11169±57 yr BP which is identical to the consensus value. In addition, the average measured 14C age of sample FIRI-L is 2526±28 yr BP based on four measurements. The measured 14C ages of intercomparison samples achieved a comparable result with the consensus values, indicating that our graphitization system is capable of providing a valid dating quality on organic material.

Table 1 14C dating results of samples from the fourth international radiocarbon intercomparison (FIRI) through the H2 reduction method. The age is within 1σ uncertainties.

a The consensus values of intercomparison samples are from Naysmith et al. (Reference Naysmith, Scott, Dunbar and Cook2019) and Scott et al. (Reference Scott, Cook, Naysmith and Staff2019)

Case Study of JCE Peat Core Samples

A case study of JCE core drilled in Jinchuan Mire of Jilin Province, northeastern China was carried out with the semiautomatic graphitization system. Four moss samples were picked up from the depths of 68–70, 78–80, 88–90, and 98–100 cm of the core. Those organic samples were graphitized by H2 reduction method using the new graphitization system and subsequent AMS 14C dating in the NTUAMS Lab. The calibrated 14C ages are processed on Bchron program (Haslett and Parnell Reference Haslett and Parnell2008) which is using IntCal20 calibration curve for calendar age calibration. The AMS results show an increasing trend in 14C ages with depth (Table 2). However, the older 14C age of about 1315±70 cal BP at the depth of 88–89 cm is out of depositional sequence. The old carbon influence on the 14C age of the peat samples has been studied by Li et al. (Reference Li, Wang, Sun, Chou, Li, Xia, Zhao, Yang and Kashyap2019) and Misra et al. (Reference Misra, Kashyap, Chou, Chang, Li, Ning, Sun, Wang and Zhao2023): dissolved CO2 in peat water produced by the decomposition of older peat plants can be uptake by some peat roots during photosynthesis.

Table 2 Results of 14C ages from the JCE core, Jinchuan Mire of NE China. The error for pMC (percent Modern Carbon) and 14C age is given 1σ uncertainties. The ages were calibrated by IntCal20.

A 92-cm-long JCA core was taken nearby the JCE core site in the same peatland. Li et al. (Reference Li, Wang, Sun, Chou, Li, Xia, Zhao, Yang and Kashyap2019) proposed the AMS calibrated 14C ages of the JCA core, ranging from the modern time at the topmost to 880 cal BP at the depth of 87.5 cm. The AMS 14C dating results of the four moss samples from the JCE core could correspond to the JCA core below 60 cm depth. Therefore, the depositions and chronologies between the JCA and JCE profiles should be comparable.

CONCLUSION AND OUTLOOK

A semiautomatic graphitization system has been well-developed at the NTUAMS Lab which has been running steadily for over a year. The reliability of the instrument is well-proven by the accuracy and repeatability in AMS 14C ages and isotope ratios, using the backgrounds, intercomparison samples and standards. The entire procedure of iron conditioning, sample combustion, gas loading, and graphitization takes about 7 hours for six samples. The operation of graphite reduction with H2 allows for the efficient production of high-quality graphite from organic materials containing 0.5–1.6 mg carbon mass.

It is optimistic to make more improvements in the future. The production of high-quality graphite requires a sufficient carbon content (>0.5 mg), therefore it is inconvenient for materials with low-carbon content such as sediments, which demand a large amount of sample weight. To address this issue, a smaller reactor volume will be designed in the future to enhance the reaction pressure and increase graphitization efficiency. In addition, the graphitization system is anticipated to be fully automated, and will be tested on carbonate samples.

ACKNOWLEDGMENTS

Thanks to Instrumentation Center of National Taiwan University for support. This study was funded from Ministry of Science and Technology of Taiwan (MOST 108-2116-M-002-012 and MOST 109-2116-M-002-018) and The National Science and Technology Council of Taiwan (NSTC 111-2116-M-002-020) to H-CL. The authors would like to thank Ms. Chun-Yen Chou and Ms. Ting-Yi Chang at the NTUAMS Lab for operating the AMS 14C dating.

Footnotes

Selected Papers from the 24th Radiocarbon and 10th Radiocarbon & Archaeology International Conferences, Zurich, Switzerland, 11–16 Sept. 2022.

References

REFERENCES

Aerts-Bijma, AT, van der Plicht, J, Meijer, HAJ. 2001. Automatic AMS Sample Combustion and CO2 Collection. Radiocarbon 43(2A):293298.CrossRefGoogle Scholar
Barile, F, Barone, S, Fedi, M, Liccioli, L, Schiavulli, L, Casamassima, G, Valentini, A, Pastore, C, Paticchio, V. 2019. Preparation of graphite targets for Accelerator Mass Spectrometry (AMS) radiocarbon measurements at INFN-Bari Laboratory. 2019 IEEE 8th International Workshop on Advances in Sensors and Interfaces (IWASI); 13–14 June 2019. p. 285–290.CrossRefGoogle Scholar
Haslett, J, Parnell, A. 2008. A simple monotone process with application to radiocarbon-dated depth chronologies. Journal of the Royal Statistical Society: Series C (Applied Statistics) 57(4):399418.Google Scholar
Li, H-C, Wang, J, Sun, J-J, Chou, C-Y, Li, H-K, Xia, Y-Y, Zhao, H-Y, Yang, Q-N, Kashyap, S. 2019. Study of Jinchuan Mire in NE China I: AMS 14C, 210Pb and 137Cs dating on peat cores. Quaternary International 528:917.CrossRefGoogle Scholar
Li, H-C, Chang, Y, Berelson, WM, Zhao, M, Misra, S, Shen, T-T. 2022. Interannual variations of D14CTOC and elemental contents in the laminated sediments of the santa barbara basin during the past 200 years. Frontiers in Marine Science 9:823793.CrossRefGoogle Scholar
Macario, KD, Alves, EQ, Moreira, VN, Oliveira, FM, Chanca, IS, Jou, RM, Diaz, M. 2017. Fractionation in the graphitization reaction for 14C-AMS analysis: the role of Zn×the role of TiH2 . International Journal of Mass Spectrometry 423:3945.CrossRefGoogle Scholar
Marzaioli, F, Borriello, G, Passariello, I, Lubritto, C, Cesare, ND, D’Onofrio, A, Terrasi, F. 2008. Zinc reduction as an alternative method for AMS radiocarbon dating: process optimization at Circe. Radiocarbon 50(1):139149.CrossRefGoogle Scholar
Misra, S, Kashyap, S, Chou, CY, Chang, TY, Li, H-C, Ning, XY, Sun, JJ, Wang, J, Zhao, MX. 2023. The influence of plant species and pretreatment on the 14C age of Carex-dominated peat plants of a peat core from Jinchuan Mire, NE China. Radiocarbon:1-21. doi: 10.1017/RDC.2023.112CrossRefGoogle Scholar
Naysmith, P, Scott, EM, Dunbar, E, Cook, GT. 2019. Humics—their history in the radiocarbon intercomparison studies. Radiocarbon 61(5):14131422.CrossRefGoogle Scholar
Němec, M, Wacker, L, Gäggeler, H. 2010. Optimization of the graphitization process at age-1. Radiocarbon 52(3):13801393.CrossRefGoogle Scholar
Rinyu, L, Futó, I, Kiss, ÁZ, Molnár, M, Svingor, É, Quarta, G, Calcagnile, L. 2007. Performance test of a new graphite target production facility in atomki. Radiocarbon 49(2):217224.CrossRefGoogle Scholar
Rozanski, K, Stichler, W, Gonfiantini, R, Scott, EM, Beukens, RP, Kromer, B, van der Plicht, J. 1992. The iaea 14C intercomparison exercise 1990. Radiocarbon 34(3): 506519.CrossRefGoogle Scholar
Scott, EM, Cook, GT, Naysmith, P, Staff, RA. 2019. Learning from the Wood Samples in ICS, TIRI, FIRI, VIRI, and SIRI. Radiocarbon 61(5):12931304.CrossRefGoogle Scholar
Stuiver, M. 1983. International agreements and the use of the new oxalic acid standard. Radiocarbon 25(2):793795.CrossRefGoogle Scholar
Vogel, JS, Southon, JR, Nelson, DE, Brown, TA. 1984. Performance of catalytically condensed carbon for use in accelerator mass spectrometry. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms 5(2):289293.CrossRefGoogle Scholar
Wacker, L, Němec, M, Bourquin, J. 2010. A revolutionary graphitisation system: Fully automated, compact and simple. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms 268(7):931934.CrossRefGoogle Scholar
Xu, X, Trumbore, SE, Zheng, S, Southon, JR, McDuffee, KE, Luttgen, M, Liu, JC. 2007. Modifying a sealed tube zinc reduction method for preparation of AMS graphite targets: Reducing background and attaining high precision. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms 259(1):320329.CrossRefGoogle Scholar
Figure 0

Figure 1 Photographs of the semiautomatic graphitization system and its devices. (a) The vacuum system included a diaphragm pump (1), dry scroll pump (2), and one high vacuum turbo molecular pump (3). (b) Picture shows the components of temperature electrics (4), CO2 trap (5), reused tube (6), and computer for monitoring and controlling (7). (c) Picture shows EA (8) connected with the graphitization system which is comprised of six-unit reactors. Each reactor has an oven (9), a glass tube (10), and a pressure transducer (11). Gas inlet of CO2 (12) and H2 (13). (d) Diagram shows the schematic graphitization system.

Figure 1

Figure 2 Reduction performance of the graphitization system. Samples with carbon weight of 0.36–1.55 mg show that the graphitization reaction can be finished within 3 hours, but samples with carbon weight less than 0.36 mg are unable to provide reliable graphite due to low yield.

Figure 2

Figure 3 (a) AMS 12C3+ currents of the graphite, produced by the H2 reduction method, from the anthracite background (BKG) samples. (b) The pMC results using two reduction methods are shown in different colors. The higher pMC values are due to the weak 12C3+ current. The blank of the AMS is an average of ∼0.08 pMC.

Figure 3

Figure 4 (a) The pMC results from the OXII standards using H2 and TiH2/Zn reduction methods respectively. The pMC results of OXII standards were subtracted by a value of AMS background (0.08 pMC). (b) The difference in δ13C suggests that the carbon isotope fractionation during the graphitization resulted from the different reduction methods. The isotope fractionation using AMS δ13C corrected to –17.8‰.

Figure 4

Table 1 14C dating results of samples from the fourth international radiocarbon intercomparison (FIRI) through the H2 reduction method. The age is within 1σ uncertainties.

Figure 5

Table 2 Results of 14C ages from the JCE core, Jinchuan Mire of NE China. The error for pMC (percent Modern Carbon) and 14C age is given 1σ uncertainties. The ages were calibrated by IntCal20.