Wearable CNT/PAni/fabric piezoresistive sensor for continuous blood pressure monitoring

Published online by Cambridge University Press:  10 February 2025

Milad Barati
Affiliation:
MEMS & NEMS Lab, Department of Microsystem and Photonics, School of Intelligent Systems Engineering, University of Tehran, Tehran, Iran
Alireza Nikfarjam*
Affiliation:
MEMS & NEMS Lab, Department of Microsystem and Photonics, School of Intelligent Systems Engineering, University of Tehran, Tehran, Iran
*
Corresponding author: Alireza Nikfarjam; Email: [email protected]

Abstract

Wearable pressure sensors with high sensitivity, fast response time, and low detection limit have great potential for blood pressure monitoring and early diagnosis of hypertension. This article introduces a piezoresistive pressure sensor based on carbon nanotubes (CNTs), polyaniline (PAni), and fabric (CNT/PAni/fabric) for health monitoring applications. This sensor is made by using two layers of linen fabric coated with CNT and PAni. These layers are placed on a polyester fabric substrate. One of the coated layers has a mesh structure, which increases the sensitivity of the sensor and lowers its detection limit. The CNT/PAni/fabric sensor has a high sensitivity of 2.035 kPa−1 at pressures from 0 to 0.2 kPa, a response time of 290 ms, and a detection limit of 1.5 Pa. These features make it suitable for measuring blood pressure. The results obtained by measuring blood pressure using the pulse transit time method on four people, compared with the values obtained using the digital sphygmomanometer, show a discrepancy ranging between 0.019% and 1.62%. Also, the average error and standard deviation for the sensor measurement in systolic and diastolic pressures are 0.56 ± 0.33 and 0.57 ± 0.46, respectively, which shows that measurement with this sensor can be an alternative to existing devices.

Type
Research Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2025. Published by Cambridge University Press

1. Introduction

Pressure sensors have a wide range of applications in soft robotic skin (Zhan et al., Reference Zhan, Lin, Tran, An, Wei, Du, Tran and Lu2017), human motions detection (Ge et al., Reference Ge, Cai, Dong, Zhang, Shao, Huang and Dong2018), energy harvesting (Cheng et al., Reference Cheng, Du, Wang, Mao, Xu, Zhang, Zhong, Jiang, Wang and Sui2018), medical diagnostics (Li et al., Reference Li, Fan, Li, Cao, Xiao, Wang, Liang, Wang, Jiang, Wang and Zhu2020), and health monitoring (Huang et al., Reference Huang, Witomska, Aliprandi, Stoeckel, Bonini, Ciesielski and Samorì2019). The key features of these sensors are high sensitivity, low weight, and low-cost fabrication processes. These characteristics demonstrate the high potential of pressure sensors in health monitoring. Blood pressure measurement is one of the major challenges in this field. High blood pressure is a major risk factor for cardiovascular disease, including heart attack, aneurysm, and heart failure. High blood pressure can be controlled by lifestyle changes and medication, but because it has no symptoms, it often goes unnoticed by most people. Therefore, there is a need for a wearable system that can continuously measure blood pressure (Kim et al., Reference Kim, Choi, Jeong, Cho, Kim, Kim, Oh and Park2020; Kumar, Reference Kumar2021).

Pressure sensors are devices that convert mechanical pressure into electrical signals. Parameters such as linearity, sensitivity, detection limit, response time, and stability are used to describe their performance. The operating mechanism of these sensors can be piezoresistive (Cao et al., Reference Cao, Su, Fan, Qiu, Su and Li2021), triboelectric (Chen et al., Reference Chen, Wu, Lin, Duan, Xu, Pan, Zhang, Xu, Huang, Hu and Zhou2020), piezoelectric (Sreeja and Sankararajan, Reference Sreeja and Sankararajan2022), and piezocapacitive (Kwon et al., Reference Kwon, Lee, Shim, Ryu, Kim, Kim, Kim and Park2016). However, piezoresistive and piezocapacitive sensors are the most promising for wearable applications. The piezocapacitance mechanism can be thought of as a parallel plate capacitor. When pressure is applied, the distance between the two plates decreases and the capacitance increases. In this example of these sensors, two parallel silicon plates, on which silver nanowires are placed, are used as capacitor plates. The lines of silver nanowires (AgNWs) are 2 mm long and placed 2 mm apart in the silicon plate, and their ends are made of gallium indium liquid metal for bonding. The lines of AgNWs are superimposed on two planes with an angle difference of 90°. Ecoflex is used as the dielectric layer between the two plates. The sensitivity for pressures below and above 500 kPa is 1.62 Mpa−1 and 0.57 Mpa−1, respectively. The response time and detection limit are 40 ms and 0.6 g, respectively (Yao and Zhu, Reference Yao and Zhu2014). Piezoresistive sensors are devices that change their electrical resistance in response to pressure. Electrical resistance indicates how well a device reduces current. This response is typically caused by a change in contact resistance between two materials or surfaces due to mechanical deflection caused by pressure. Gao et al. (Reference Gao, Yu, Tan and Xuan2018) conducted research by casting PDMS on sandpaper and using it as a mould. They then used the formed PDMS as a mould and added a mixture of PDMS and carbon nanotubes (CNTs). Finally, they placed two samples of the same structure face-to-face to create a low-cost sensor. This sensor has a sensitivity of 0.2 kPa−1, a detection limit of 5 Pa, and a response time of 190 ms. When comparing piezoresistive and piezocapacitive mechanisms, capacitive pressure sensors are highly sensitive to static deformation. In contrast, piezoresistive sensors perform well in detecting transient or dynamic deformation due to the intermittent nature of their piezoelectric effects. Piezoresistive mechanisms also offer advantages such as easy fabrication methods, simple readout mechanisms, fast response times, and easy signal acquisition. Another feature of piezoresistive sensors is their deficient power consumption, making them more popular in recent years (Ding et al., Reference Ding, Xu, Onyilagha, Fong and Zhu2019; Cao et al., Reference Cao, Su, Fan, Qiu, Su and Li2021).

To improve the performance of piezoresistive sensors, various materials have been used in their structure to reduce their initial resistance. Conductive polymers such as polypyrrole (Wang et al., Reference Wang, Ding, Yuan, Cao, He, Peng and Li2016), PEDOT: PSS (Ding et al., Reference Ding, Yang, Tolle and Zhu2018), or polyaniline (PAni; Chen et al., Reference Chen, Liu, Xu, Ye, Li, Qin and Zhang2022) have proven to be superior alternatives to other expensive conductive materials (Kannichankandy et al., Reference Kannichankandy, Pataniya, Narayan, Patel, Sumesh, Patel, Solanki and Pathak2021). PAni can be an excellent alternative due to its favorable properties. It has an inexpensive monomer and a convenient synthesis method. In addition, features such as high thermal, electrical and chemical stability, environmental compatibility, and the ability to protect against corrosion are among its advantages (Visakh et al., Reference Visakh, Della and Falletta2017; Mozafari and Chauhan, Reference Mozafari and Chauhan2019). Zheng et al. (Reference Zheng, Wu, Huang, Xu, Yang, Liu, Huang, Xie and Yang2019) reported microcracked and cavitated PAni foams supported by polydimethylsiloxane (PDMS), which showed a sensitivity of 0.055 kPa−1 in the range of 0–5 kPa and its detection limit was 4 Pa. PAni is combined with other materials in newer technologies due to its fragile structure, medium conductivity, and short charge/discharge cycle life. Carbon materials such as graphene (Huang et al., Reference Huang, Wang, Li, Wu, Wang and Yang2019), reduced graphene oxide (rGo; Zheng et al., Reference Zheng, Wu, Huang, Xu, Yang, Liu and Yang2020), and CNTs (Hong et al., Reference Hong, Oh, Park, Yun, Jin, Sun, Zi and Ha2017) are ideal candidates for integration with PAni due to their exceptional electron mobility, mechanical stability, and thermal stability. The exceptional mechanical properties of CNTs, including their high Young’s modulus and tensile strength, make them optimal and promising reinforcements for significantly improving the mechanical properties of polymer–CNT composites (Arash et al., Reference Arash, Wang and Varadan2014). The addition of CNTs to PAni not only significantly increases its electrical conductivity but also improves its stability, effective surface area, and brittleness (Zhang et al., Reference Zhang, Cao, Wang, Yang, Shi and Gu2008). Xiao et al. (Reference Xiao, Sun and Li2023) prepared a 3D conductive foam from polyurethane, CNT, and PAni. This foam showed impressive electrical conductivity and durability, resulting in a significant response over a wide range of 0–30,000 kPa.

Here, we have demonstrated a sensitive piezoresistive sensor based on a mesh structure of linen fabric coated with PAni and CNT. The sensor consists of two linen fabrics coated on a polyester fabric substrate. One of the linen fabrics has a mesh structure which increases the air space between the two layers, increasing the sensitivity of the sensor and reducing its detection limit. The two-layer PAni/CNT/fabric sensor has a sensitivity of (2.035 kPa−1) in the range (0–200 Pa) and a detection limit of (1.5 Pa), making it suitable for measuring pulse waves and blood pressure. Considering the high risk of hypertension disease and the importance of its prevention, it seems necessary to develop a blood pressure monitoring system that can be used to measure blood pressure at home without the need for a monitor.

2. Experimental section

2.1. Fabrication of CNT/PAni/fabric layer and pressure sensor

The fabrication process of the CNT/PAni/fabric layer is shown in Figure 1(a). Linen fabrics are cut into 1 × 1 cm2 pieces. Before coating, they are washed three times with deionized water (DIW) and ethanol. The fabric is washed in an ultrasonic bath. The washing time is set to 480 s. To prepare the CNT solution, 80 mL of DIW is added to 28.4 mg of CNT. The linen fabric is immersed in the solution and an ultrasonic probe is used to distribute the CNT evenly on the fabric. The mixture is stirred several times for 10 min each, to ensure complete coverage of the canvas. Then the fabrics from the solution are removed and allowed to dry completely in an oven for 10 min.

Figure 1. Fabrication process of CNT/PAni/fabric. (a) Overview of the process to fabricate the sensing layer (CNT/PAni/fabric). (b) Different parts of the sensor. (c) The final shape of the sensor.

The next step is to make two separate solutions, A and B, to create a PAni solution. Solution A is made by combining 60 mL of DIW with 4.8 mL of hydrochloric acid (HCl). After stirring for 15 min, 4 mL of aniline is added and the solution is placed at 0°C for polymerization. Solution B is prepared by combining 24 mL of DIW with 1.95 mL of HCl and then adding 0.7 g of ammonium persulfate (APS). Similar to Solution A, it is stirred for 15 min and placed at 0°C. The coated fabrics are immersed in Solution A, followed by the dropwise addition of Solution B. To complete the polymerization process, the resulting solution is kept at 0°C for 20 hr. The tissues are then removed from the solution and washed several times with DIW.

To make the sensor substrate, the polyester fabric is cut into 5 × 2 cm2 pieces and then washed in an ultrasonic bath for 480 s, first with DIW and then with ethanol. This process is repeated several times to ensure the cleanliness of the fabric. To dry, the fabric is placed in an oven at 80°C for 10 min. Once the polyester fabric is dry, the double-sided copper tape is cut to a size of 1 × 5 cm2 and placed vertically on both sides of the polyester fabric. The copper tape plays the role of the electrode in two layers and the entire sensor output, which is connected to the polyester fabric. To make the final sensor, the pre-prepared linen fabric coated with PAni and CNT is divided into smaller dimensions of 0.3 × 1 cm2. Five pieces of fabric are selected and placed on the polyester fabric. Three pieces of coated fabric are placed horizontally on top of the sensor, between two copper strips, at a distance of 5 mm. The remaining two pieces of fabric are placed vertically on top of these three pieces. Next, another 1 × 1 cm2 piece of linen fabric coated with PAni and CNT is placed on these parts. Finally, the sensor is encapsulated on both sides with masking tape.

2.2. Characterization of CNT/PAni/fabric layer and measurements

An electron microscope is used to observe the morphology of the sample surface using scanning electron microscopy (SEM) and energy-dispersive X-ray spectroscopy (EDS) analyses. The SEM analysis was performed at 20 kV voltage and the EDS measurement was also performed with energy between 0 and13 keV. CNT and PAni coating on linen fabric can also be confirmed by Fourier transform infrared spectroscopy (FTIR) analysis. For FTIR, an instrument (Perkin Elmer spectrometer) was used and the wavelength between 400 and 4,000 cm−1 was analyzed.

A digital multimeter (Victor VC97) was used to measure the resistance of the sensor. A digital oscilloscope (RIGOL DS 1054) was used to measure the voltage changes due to the application of pressure and a voltage source (PS-405 U2) was used to supply the input voltage. A resistive divider circuit is used to convert changes in sensor resistance due to pressure into changes in voltage. To have a digital output of the circuit and finally to display the voltage changes resulting from pressure changes as changes in blood pressure, an Arduino board (Arduino UNO R3) is used and the output signal is connected to one of the analog-to-digital conversion pins.

To calibrate the sensors, it is necessary to use a device that can measure blood pressure in real time. After recording the signals from two sensors using the Arduino board and its digital display, the final step is to use a cuff-based sphygmomanometer (Omron M6 Comfort) to obtain the people’s actual blood pressure.

3. Results and discussion

3.1. Pressure sensor performance

First, the linen fabric was coated with CNT solution to create the pressure-sensitive layer. The CNT-coated fabric was then immersed in the PAni solution during polymerization. To fabricate the sensor, a mesh-like CNT/PAni/fabric structure is placed on polyester and covered with a 1 × 1 cm2 square fabric, as shown in Figure 1. The detailed fabrication process of the designed device is given in the experimental section. The CNT and PAni coating on the linen fabric was identified by SEM images, as shown in Figure 2. Figure 2(a) shows the SEM image of the linen fabric before coating. The fibers of this fabric are irregularly distributed, creating a large air gap between them. The surface of the uncoated fabric is completely smooth and the average length of each fiber is 150–200 μm. After CNTs are added to the fabric, a curved tubular structure appears on its surface (Figure 2(b)). Figure 2(c) shows that PAni forms a cluster-like network of nanowires on the surface of the linen fabric. This dense network of PAni fibers increases the roughness and electrical conductivity of the fabric. Table 1 shows the percentage by weight of the elements in the CNT/PAni/fabric based on EDS analysis and also gives the percentage by mass of each element. The chemical structure of linen fabric is similar to that of cellulose cotton and the presence of cellulose ((C6H10O5) n) has caused the presence of carbon and oxygen in the EDS analysis. Also, after the addition of CNT, the percentage of carbon in the structure has increased compared to other elements. According to Table 1, there has been a definite absorption of PAni on the fabric, which can be detected by the presence of elements such as nitrogen, sulfur, and chlorine. Nitrogen and sulfur are related to the presence of APS ((NH4)2S2O8) in the formation of PAni and Cl is formed by the addition of HCl (Muthukumar et al., Reference Muthukumar, Thilagavathi and Kannaian2016).

Figure 2. SEM images: (a) Linen Fabric at 3,000x magnification, (b) CNT/fabric at 15,000x magnification, and (c) CNT/PAni/fabric at 6,000x magnification.

Table 1. Weight percentage of elements in CNT/PAni/fabric based on EDS analysis

Figure 3 shows the results of the FTIR analysis for CNT/PAni/fabric. The long peak at wave number 3039.5 cm−1 is the characteristic peak of the hydroxyl groups (OH) of cellulose. A peak at 2,804.5 cm−1 is seen, which is related to the stretching vibration of the C-H bond in cellulose. The absorption at 1312.8 cm−1 is related to the bending vibration of the C-O group, which is related to the aromatic ring of the polysaccharide. The strong peak observed at 1,019.19 cm−1 is related to the stretching vibration of CO with OH in cellulose polysaccharides (Chung et al., Reference Chung, Lee and Choe2004; Portella et al., Reference Portella, Romanzini, Angrizani, Amico and Zattera2016). The peak in the range of 1,100–1,200 cm−1 (1,101.6 cm−1) confirms the hexagonal structure of CNT, which is related to the C=C bond of the CNT skeletal vibration mode. The peak at 1,054 cm−1 is also related to the C-O vibration of the ester, phenol, or carboxyl group, which makes it difficult to give a definite opinion due to overlap (Lin et al., Reference Lin, Xiu, Zhu, Moon and Wong2008; Ţucureanu et al., Reference Ţucureanu, Matei and Avram2016). The two peaks at 1,575 cm−1 and 1,493.3 cm−1 are related to the stretching vibration of the quinoid ring and the benzenoid ring. The peak at 740.59 cm−1 characterizes the out-of-plane C-H bond in the aromatic ring and the stretching vibration of the S-O bond of the sulfonate groups attached to the aromatic ring causes the absorption at 682.69 cm−1 (Shao et al., Reference Shao, Jamal, Xu, Ubul and Abdiryim2012; Sharma and Sharma, Reference Sharma and Sharma2013).

Figure 3. FTIR analysis for CNT/PAni/fabric.

The sensor mechanism is based on resistance changes caused by pressure changes. When pressure is applied to the sensor, the fabric on top makes more contact with the linen fabric below. This contact increases the conductivity between the two copper electrodes. In addition, applying pressure reduces the amount of air trapped between the mesh and the upper fabric, thereby reducing the resistance. When the pressure is released, the sensor returns to its original state and resistance level.

The combination of CNT and PAni on fabric results in a highly sensitive sensor that operates in two linear ranges. The sensitivity of the sensor can be calculated using Equation (1):

(1) $$ \mathrm{Sensitivity}=\frac{\Delta R/R}{\Delta P}, $$

where ΔR and R are the changes in resistance and initial resistance respectively, and ΔP is the applied pressure. The sensitivity of the sensor is shown in Figure 4. Figure 4 shows that in the low-pressure range (0–0.2 kPa), the sensitivity has increased much faster. The rapid increase in conductive pathways between the two electrodes is the primary reason for this effect. The current flow from one electrode to the other electrode on both sides of the polyester fabric is mostly through the lower fabric, which has a mesh-like structure, which has empty holes in its structure. In pressures <0.2 kPa, these holes are gradually filled by the upper fabric, creating a noticeable change in the overall resistance of the path. By increasing the pressure to 0.2 kPa, almost all of the empty holes in the mesh-like structure of the lower fabric are filled by the upper fabric, and the two surfaces are completely attached and applying more pressure does not change the overall resistance of the path, and the sensitivity is noticeably reduced. The reason for the decrease in sensitivity is also the saturation of the number of conductive paths between the two copper electrodes (Luo et al., Reference Luo, Huang, Liu, Chen, Wong and Zhao2017; Chen et al., Reference Chen, Li, Cheng, He, Li, Zhang, Li, Feng, Wei, Li, Zhong and Yang2018; Gao et al., Reference Gao, Yu, Tan and Xuan2018; Park et al., Reference Park, Kim, Chu and Khine2018; Xia et al., Reference Xia, Wang, Jian, Wang and Zhang2018; He et al., Reference He, Xiao, Lu, Shi, Zhang, Liang, Pan, Kuo and Chen2019). This pressure range is critical for blood pressure measurement. Figure 4(a) shows a sensitivity diagram for three different sensors. The CNT/PAni/fabric sensor has a sensitivity of 2.035 kPa−1, which is much better than the PAni/fabric (0.32 kPa−1) and CNT/fabric (0.12 kPa−1) sensors. This indicates that the combination of these two materials has been very effective in achieving high sensitivity at low pressures. Figure 4(b) shows the measuring range of the sensor, which is 0–50 kPa, together with the sensitivity for the whole range. The CNT/PAni/Fabric sensor has a higher sensitivity than the PAni/Fabric and CNT/Fabric sensors over the entire range. The sensitivities of the CNT/PAni/fabric, PAni/fabric, and CNT/fabric sensors are 0.0152 kPa−1, 0.0095 kPa−1, and 0.0057 kPa−1 respectively.

Figure 4. Diagram of changes in resistance (sensitivity) of sensors made of a combination of fabric with CNT, PAni, and a combination of CNT/PAni by applying a pressure of (a) 0–0.2 kPa and (b) 0–50 kPa. (c) Response time of CNT/PAni/fabric sensor by applying pressure of 0.1 kPa. (d) Detection limit of CNT/PAni/fabric sensor by applying 1.5 Pa pressure. (e) CNT/PAni/fabric sensor response by applying 1 kPa pressure at 1 Hz, 2 Hz, and 3 Hz frequencies. (f) CNT/PAni/fabric sensor response by applying a constant pressure of 1 kPa for 650 cycles.

To calculate the response time of the CNT/PAni/fabric sensor, the resistance changes of the sensor were measured by applying a certain pressure. According to Figure 4(c), by applying a pressure of 0.1 kPa, the response time and recovery time of the sensor were 290 ms and 510 ms, respectively.

The detection limit was obtained by placing a piece of aluminium sheet measuring 1.5 × 2 cm2 and weighing 45 mg, which applied a pressure of 1.5 Pa to the sensor (Figure 4(d)).

Figure 4(e) shows the changes in resistance of the fabric/CNT/PAni sensor at frequencies of 1 Hz, 2 Hz, and 3 Hz. In particular, when the sensor is subjected to the same pressure of 1 kPa, the output signal curve remains constant. This observation suggests that the response of the sensor to constant pressure is independent of frequency. At a frequency of 1 Hz, the initial range of the sensor showed a change of 48.6%. This range remains constant at 1 Hz, while at frequencies of 2–3 Hz, the maximum deviation from the initial range is 4.32%, which is within an acceptable range. Figure 4(f) shows the output of the sensor when a constant pressure of 1 kPa is applied for 650 cycles. The cycle ranges show a maximum percentage difference of 1.5%.

3.2. Application: blood pressure monitoring system

To accurately measure blood pressure, the sensor must be highly sensitive and have a very low detection limit for pulse pressure measurement. In addition, the response time of the sensor is crucial for this measurement. Since the interval between two heartbeats is ~900 ms, the sensor must have a total response and recovery time of <900 ms to effectively detect a heartbeat. A CNT/PAni/fabric sensor with a total response and recovery time of 800 ms and a detection limit of 1.5 Pa is a suitable sensor for blood pressure measurement. This sensor can measure the smallest dynamic information of changes in the arteries of the human body, to monitor the pulse signal and then calculate the blood pressure. As shown in Figure 5, due to the high sensitivity of the sensor to external pressure changes, all characteristic points of the pulse waveform are measured correctly. The CNT/PAni/fabric sensor can only detect mechanical changes associated with blood pressure. A measurement kit is required to measure blood pressure. This set consists of a sensor to receive a mechanical signal, a power supply for the sensor, a circuit to manage and process the signal, and a system to display the output signal. All the components in this set work together to provide continuous blood pressure monitoring. The average heart rate for an individual is around 1 Hz, so the circuitry has been optimized to filter out signals at other frequencies. The signal is then sent to the Arduino board and its analog-to-digital conversion (ADC) section. The ADC operates at a sampling rate of 15k samples/s. The highly accurate 10-bit analog-to-digital converter and the appropriate sampling rate ensure that the information is stored in great detail. The microcontroller unit (MCU) on the Arduino board is responsible for processing the data. Once the information has been processed, the results are transmitted.

Figure 5. Waveform of the pulse signal obtained from the CNT/PAni/fabric sensor on the neck.

To calculate blood pressure using the Mons–Korteweg formula, it is important to determine the time difference between the blood reaching two different points in the body, assuming that the elastic modulus of the vessel walls remains constant. In humans, the elastic modulus of the vessel walls does not change significantly over short periods of time, so the pulse transit time (PTT) is closely related to blood pressure and is influenced by physiological characteristics. Therefore, Equation (2), derived from the Mons–Korteweg formula, can be used to calculate blood pressure in humans (Chen et al., Reference Chen, Kobayashi, Ichikawa, Takeuchi and Togawa2000; Shriram et al., Reference Shriram, Wakankar, Daimiwal and Ramdasi2010).

(2) $$ {P}_e={P}_b-\alpha \frac{\Delta T}{T_b}. $$

In Equation (2), ΔT refers to the change in time taken for the pulse to travel from Point A to Point B. Tb represents the amount of PTT related to the Pb pressure. In addition, Pe represents the secondary pressure. The α coefficient is also a unique value determined by calibration and is different for each person.

For blood pressure monitoring, the functionality of the sensor was tested on healthy subjects according to standard criteria. First, the stability of the sensor was assessed on three individuals. Subsequently, five different sensors were used on a single individual to assess the reproducibility of sensor manufacture and to ensure consistent results. To demonstrate the ability of the sensor to monitor pressure, two sensors were placed on the subject’s wrist and neck, where the radial artery passes and the received signal is stronger. The subject was then asked to hold the palm of their hand up and stretch a little more than usual to make the radial artery more visible on the surface of the skin. During the measurement, the person held the pressure sensor close to their heart. No allergic reaction or pain was reported by the subject. Adhesive tape is used to attach the sensor to the skin of the hand, and the pressure on the sensor when placed on the skin is high to ensure good contact with the skin. This pressure reduces the initial resistance of the sensor. As blood pressure increases, the radial artery expands and the surrounding tissue deforms, causing the sensor to deform and change its resistance. The transit time is obtained by measuring the change between two pulses.

The blood pressure is first measured by an electronic sphygmomanometer and then the PTT is obtained directly from the sensors. Important components in measuring blood pressure are systolic pressure and diastolic pressure. These are among the most commonly tested components when assessing a person’s cardiovascular health.

To obtain the most accurate results, three consecutive blood pressure readings were taken with the electronic sphygmomanometer and the fourth reading was used to calibrate the sensor.

Figure 6 shows a diagram of the sensors on the subject. The time difference between the two pulse signals reaching the sensors was between 90 and 120 ms. To ensure accurate readings, the signals were averaged over four cardiac cycles and compared with an electronic sphygmomanometer. In this case, the average PTT was 112.5 ms. According to Equation (2), the systolic and diastolic blood pressures obtained by the sensor were 131.96 and 64.98, respectively. The systolic and diastolic blood pressures were measured using an OMRON cuff blood pressure monitor, which indicated 131 mmHg and 65 mmHg, respectively. The measurement error in this case is 0.7% for systolic pressure and 0.019% for diastolic pressure. To demonstrate the feasibility of the wearable CNT/PAni/fabric sensor system for noninvasive continuous biomedical monitoring, we also performed cardiovascular parameter (BP) measurements on three additional subjects. These subjects ranged in age from 26 to 34 years and had different health conditions. For each participant, we first used our sensor system to measure their blood pressure. Then, to validate the accuracy of the BP measurements, we used a cuff-based sphygmomanometer (OMRON) to assess the BP of the same individual. A statistical result of the systolic and diastolic BP measurements is shown in Table 2.

Figure 6. Two waveforms were received from two sensors, and the time difference between two peaks in the signals was calculated for a blood pressure reading of 131/65 mmHg.

Table 2. Results of systolic and diastolic blood pressure measurement for four participants

Note. SBP monitoring discrepancy = |SBPWCSPS − SBPcuff-based|/SBPcuff-based × 100%; DBP monitoring discrepancy = |DBPWCSPS − DBPcuff-based|/DBPcuff-based × 100%).

Among the persons tested, Subject C has high blood pressure, and according to Table 2, he has the highest discrepancy in general, a low discrepancy (between 0.019% and 1.62%) between the measured blood pressure and the results obtained from the sensor shows that this sensor can be used as a device for blood pressure measurement.

The average measurement error of the sensor and the standard deviation were also calculated and their value for systolic and diastolic pressure were 0.056 ± 0.33 mmHg and 0.057 ± 0.46 mmHg, respectively. According to ISO 81060-2, noninvasive blood pressure can be replaced by an arterial catheter if the mean distortion is <5 mmHg and the standard deviation is <8 mmHg according to AAMI (Kim et al., Reference Kim, Chou, Le, Wong, Chu and Khine2019).

4. Conclusion

In this study, we have developed a wearable piezoresistive sensor using CNT and PAni as the pressure-sensitive layer. This sensor has a wide measurement range of 0–50 kPa and a high sensitivity of 2.035 kPa−1 within the range of 0–0.2 kPa. It has a fast response time of 290 ms, a recovery time of 510 ms, and an extremely low detection limit of 1.5 Pa. The pressure sensor has also demonstrated good stability over time and with the application of specific pressure at different frequencies, confirming the sensor’s capability.

We have also demonstrated that this sensor can be used as a blood pressure measurement system. By placing two CNT/PAni/fabric sensors on the wrist and neck of the subjects, where the strongest pulses were received from these points, we received the pulse signal well. Then, using the PTT method and data transmission, we were able to measure the blood pressure using Arduino and display the output on the laptop‥ The blood pressure values obtained using the CNT/PAni/Fabric sensor were very close compared to the blood pressure values shown on the digital sphygmomanometer, indicating the effectiveness of this sensor for measuring blood pressure. One of the advantages of this system is the continuous measurement of blood pressure and its portability, which promises to be part of the new generation of blood pressure measurement systems. Given the asymptomatic nature of a disease such as hypertension, a system capable of continuously measuring blood pressure can be very useful in its prevention.

Data availability statement

Data sets generated during the current study are available from the corresponding author on reasonable request.

Competing interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this article.

Ethical standard

The pressure signals were sourced directly from the author, Milad Barati. In this way, Milad Barati declares consent for all aspects of the experiment, including the acquisition of signals. Ethical approval and consent are not relevant to this article type and is not applicable.

References

Arash, B, Wang, Q and Varadan, V (2014) Mechanical properties of carbon nanotube/polymer composites. Scientific Reports 4(1), 6479.CrossRefGoogle ScholarPubMed
Cao, M, Su, J, Fan, S, Qiu, H, Su, D and Li, L (2021) Wearable piezoresistive pressure sensors based on 3D graphene. Chemical Engineering Journal 406, 126777.CrossRefGoogle Scholar
Chen, F, Liu, H, Xu, M, Ye, J, Li, Z, Qin, L and Zhang, T (2022) Fast-response piezoresistive pressure sensor based on polyaniline cotton fabric for human motion monitoring. Cellulose 29(12), 69836995.CrossRefGoogle Scholar
Chen, M, Li, K, Cheng, G, He, K, Li, W, Zhang, D, Li, W, Feng, Y, Wei, L, Li, W, Zhong, G and Yang, C (2018) Touchpoint-tailored ultrasensitive piezoresistive pressure sensors with a broad dynamic response range and low detection limit. ACS Applied Materials & Interfaces 11(2), 25512558.CrossRefGoogle Scholar
Chen, S, Wu, N, Lin, S, Duan, J, Xu, Z, Pan, Y, Zhang, H, Xu, Z, Huang, L, Hu, B and Zhou, J (2020) Hierarchical elastomer tuned self-powered pressure sensor for wearable multifunctional cardiovascular electronics. Nano Energy 70, 104460.CrossRefGoogle Scholar
Chen, W, Kobayashi, T, Ichikawa, S, Takeuchi, Y and Togawa, T (2000) Continuous estimation of systolic blood pressure using the pulse arrival time and intermittent calibration. Medical and Biological Engineering and Computing 38, 569574.CrossRefGoogle ScholarPubMed
Cheng, H, Du, Y, Wang, B, Mao, Z, Xu, H, Zhang, L, Zhong, Y, Jiang, W, Wang, L and Sui, X (2018) Flexible cellulose-based thermoelectric sponge towards wearable pressure sensor and energy harvesting. Chemical Engineering Journal 338, 17.CrossRefGoogle Scholar
Chung, C, Lee, M and Choe, EK (2004) Characterization of cotton fabric scouring by FT-IR ATR spectroscopy. Carbohydrate Polymers 58(4), 417420.CrossRefGoogle Scholar
Ding, Y, Xu, T, Onyilagha, O, Fong, H and Zhu, Z (2019) Recent advances in flexible and wearable pressure sensors based on piezoresistive 3D monolithic conductive sponges. ACS Applied Materials & Interfaces 11(7), 66856704.CrossRefGoogle ScholarPubMed
Ding, Y, Yang, J, Tolle, CR and Zhu, Z (2018) Flexible and compressible PEDOT: PSS@ melamine conductive sponge prepared via one-step dip coating as piezoresistive pressure sensor for human motion detection. ACS Applied Materials & Interfaces 10(18), 1607716086.CrossRefGoogle ScholarPubMed
Gao, Y, Yu, G, Tan, J and Xuan, F (2018) Sandpaper-molded wearable pressure sensor for electronic skins. Sensors and Actuators A: Physical 280, 205209.CrossRefGoogle Scholar
Ge, G, Cai, Y, Dong, Q, Zhang, Y, Shao, J, Huang, W and Dong, X (2018) A flexible pressure sensor based on rGO/polyaniline wrapped sponge with tunable sensitivity for human motion detection. Nanoscale 10(21), 1003310040.CrossRefGoogle ScholarPubMed
He, J, Xiao, P, Lu, W, Shi, J, Zhang, L, Liang, Y, Pan, C, Kuo, S-W and Chen, T (2019) A universal high accuracy wearable pulse monitoring system via high sensitivity and large linearity graphene pressure sensor. Nano Energy 59, 422433.CrossRefGoogle Scholar
Hong, SY, Oh, JH, Park, H, Yun, JY, Jin, SW, Sun, L, Zi, G and Ha, JS (2017) Polyurethane foam coated with a multi-walled carbon nanotube/polyaniline nanocomposite for a skin-like stretchable array of multi-functional sensors. NPG Asia Materials 9(11), e448e448.CrossRefGoogle Scholar
Huang, CB, Witomska, S, Aliprandi, A, Stoeckel, M-A, Bonini, M, Ciesielski, A and Samorì, P (2019) Molecule–graphene hybrid materials with tunable mechanoresponse: Highly sensitive pressure sensors for health monitoring. Advanced Materials 31(1), 1804600.CrossRefGoogle ScholarPubMed
Huang, J, Wang, H, Li, Z, Wu, X, Wang, J and Yang, S (2019) Improvement of piezoresistive sensing behavior of graphene sponge by polyaniline nanoarrays. Journal of Materials Chemistry C 7(24), 73867394.CrossRefGoogle Scholar
Kannichankandy, D, Pataniya, PM, Narayan, S, Patel, V, Sumesh, CK, Patel, KD, Solanki, GK and Pathak, VM (2021) Flexible piezo-resistive pressure sensor based on conducting PANI on paper substrate. Synthetic Metals 273, 116697.CrossRefGoogle Scholar
Kim, J, Chou, EF, Le, J, Wong, S, Chu, M and Khine, M (2019) Soft wearable pressure sensors for beat-to-beat blood pressure monitoring. Advanced Healthcare Materials 8(13), 1900109.CrossRefGoogle ScholarPubMed
Kim, K, Choi, J, Jeong, Y, Cho, I, Kim, M, Kim, S, Oh, Y and Park, I (2020) Highly sensitive and wearable liquid metal-based pressure sensor for health monitoring applications: Integration of a 3D-printed microbump array with the microchannel. Advanced Healthcare Materials 9(7), 2000313.CrossRefGoogle ScholarPubMed
Kumar, A (2021) Flexible and wearable capacitive pressure sensor for blood pressure monitoring. Sensing and Bio-Sensing Research 33, 100434.Google Scholar
Kwon, D, Lee, T-I, Shim, J, Ryu, S, Kim, MS, Kim, S, Kim, T-S and Park, I (2016) Highly sensitive, flexible, and wearable pressure sensor based on a giant piezocapacitive effect of three-dimensional microporous elastomeric dielectric layer. ACS Applied Materials & Interfaces 8(26), 1692216931.CrossRefGoogle ScholarPubMed
Li, X, Fan, YJ, Li, HY, Cao, JW, Xiao, YC, Wang, Y, Liang, F, Wang, HL, Jiang, Y, Wang, ZL and Zhu, G (2020) Ultracomfortable hierarchical nanonetwork for highly sensitive pressure sensor. ACS Nano 14(8), 96059612.CrossRefGoogle ScholarPubMed
Lin, W, Xiu, Y, Zhu, L, Moon, K-S and Wong, C (2008) Assembling of carbon nanotube structures by chemical anchoring for packaging applications. In 2008 58th Electronic Components and Technology Conference. Lake Buena Vista, FL: IEEE, pp. 421426.CrossRefGoogle Scholar
Luo, N, Huang, Y, Liu, J, Chen, SC, Wong, CP and Zhao, N (2017) Hollow-structured graphene–silicone-composite-based piezoresistive sensors: Decoupled property tuning and bending reliability. Advanced Materials 29(40), 1702675.CrossRefGoogle ScholarPubMed
Mozafari, M and Chauhan, NPS (2019)Fundamentals and Emerging Applications of Polyaniline. Elsevier. ISBN 9780128179161.Google Scholar
Muthukumar, N, Thilagavathi, G and Kannaian, T (2016) Polyaniline-coated polyurethane foam for pressure sensor applications. High Performance Polymers 28(3), 368375.CrossRefGoogle Scholar
Park, SJ, Kim, J, Chu, M and Khine, M (2018) Flexible piezoresistive pressure sensor using wrinkled carbon nanotube thin films for human physiological signals. Advanced Materials Technologies 3(1), 1700158.CrossRefGoogle Scholar
Portella, EH, Romanzini, D, Angrizani, CC, Amico, SC and Zattera, AJ (2016) Influence of stacking sequence on the mechanical and dynamic mechanical properties of cotton/glass fiber reinforced polyester composites. Materials Research 19, 542547.CrossRefGoogle Scholar
Shao, W, Jamal, R, Xu, F, Ubul, A and Abdiryim, T (2012) The effect of a small amount of water on the structure and electrochemical properties of solid-state synthesized polyaniline. Materials 5(10), 18111825.CrossRefGoogle Scholar
Sharma, AK and Sharma, Y (2013) P-toluene sulfonic acid doped polyaniline carbon nanotube composites: Synthesis via different routes and modified properties. Journal of Electrochemical Science and Engineering 3(2), 4756.Google Scholar
Shriram, R, Wakankar, A, Daimiwal, N and Ramdasi, D (2010) Continuous cuffless blood pressure monitoring based on PTT. In 2010 International Conference on Bioinformatics and Biomedical Technology. Chengdu, China: IEEE, pp. 5155.CrossRefGoogle Scholar
Sreeja, B and Sankararajan, R (2022) Bio-compatible piezoelectric material based wearable pressure sensor for smart textiles. Smart Materials and Structures 31(12), 125015.Google Scholar
Ţucureanu, V, Matei, A and Avram, AM (2016) FTIR spectroscopy for carbon family study. Critical Reviews in Analytical Chemistry 46(6), 502520.CrossRefGoogle ScholarPubMed
Visakh, P, Della, C and Falletta, E (2017) Polyaniline Blends, Composites, and Nanocomposites. Amsterdam, The Netherlands: Elsevier.Google Scholar
Wang, C, Ding, Y, Yuan, Y, Cao, A, He, X, Peng, Q and Li, Y (2016) Multifunctional, highly flexible, free-standing 3D Polypyrrole foam. Small 12(30), 40704076.CrossRefGoogle ScholarPubMed
Xia, K, Wang, C, Jian, M, Wang, Q and Zhang, Y (2018) CVD growth of fingerprint-like patterned 3D graphene film for an ultrasensitive pressure sensor. Nano Research 11, 11241134.CrossRefGoogle Scholar
Xiao, Y, Sun, P and Li, H (2023) Flexible pressure sensor based on polystyrene foam with superelasticity and ultra-wide range. Smart Materials and Structures 32(4), 045006.CrossRefGoogle Scholar
Yao, S and Zhu, Y (2014) Wearable multifunctional sensors using printed stretchable conductors made of silver nanowires. Nanoscale 6(4), 23452352.CrossRefGoogle ScholarPubMed
Zhan, Z, Lin, R, Tran, V-T, An, J, Wei, Y, Du, H, Tran, T, and Lu, W (2017) Paper/carbon nanotube-based wearable pressure sensor for physiological signal acquisition and soft robotic skin. ACS Applied Materials & Interfaces 9(43), 3792137928.CrossRefGoogle ScholarPubMed
Zhang, H, Cao, G, Wang, Z, Yang, Y, Shi, Z and Gu, Z (2008) Tube-covering-tube nanostructured polyaniline/carbon nanotube array composite electrode with high capacitance and superior rate performance as well as good cycling stability. Electrochemistry Communications 10(7), 10561059.CrossRefGoogle Scholar
Zheng, S, Wu, X, Huang, Y, Xu, Z, Yang, W, Liu, Z, Huang, S, Xie, B and Yang, M (2019) Highly sensitive and multifunctional piezoresistive sensor based on polyaniline foam for wearable human-activity monitoring. Composites Part A: Applied Science and Manufacturing 121, 510516.CrossRefGoogle Scholar
Zheng, S, Wu, X, Huang, Y, Xu, Z, Yang, W, Liu, Z and Yang, M (2020) Multifunctional and highly sensitive piezoresistive sensing textile based on a hierarchical architecture. Composites Science and Technology 197, 108255.CrossRefGoogle Scholar
Figure 0

Figure 1. Fabrication process of CNT/PAni/fabric. (a) Overview of the process to fabricate the sensing layer (CNT/PAni/fabric). (b) Different parts of the sensor. (c) The final shape of the sensor.

Figure 1

Figure 2. SEM images: (a) Linen Fabric at 3,000x magnification, (b) CNT/fabric at 15,000x magnification, and (c) CNT/PAni/fabric at 6,000x magnification.

Figure 2

Table 1. Weight percentage of elements in CNT/PAni/fabric based on EDS analysis

Figure 3

Figure 3. FTIR analysis for CNT/PAni/fabric.

Figure 4

Figure 4. Diagram of changes in resistance (sensitivity) of sensors made of a combination of fabric with CNT, PAni, and a combination of CNT/PAni by applying a pressure of (a) 0–0.2 kPa and (b) 0–50 kPa. (c) Response time of CNT/PAni/fabric sensor by applying pressure of 0.1 kPa. (d) Detection limit of CNT/PAni/fabric sensor by applying 1.5 Pa pressure. (e) CNT/PAni/fabric sensor response by applying 1 kPa pressure at 1 Hz, 2 Hz, and 3 Hz frequencies. (f) CNT/PAni/fabric sensor response by applying a constant pressure of 1 kPa for 650 cycles.

Figure 5

Figure 5. Waveform of the pulse signal obtained from the CNT/PAni/fabric sensor on the neck.

Figure 6

Figure 6. Two waveforms were received from two sensors, and the time difference between two peaks in the signals was calculated for a blood pressure reading of 131/65 mmHg.

Figure 7

Table 2. Results of systolic and diastolic blood pressure measurement for four participants