Hostname: page-component-cd9895bd7-7cvxr Total loading time: 0 Render date: 2024-12-26T09:43:46.272Z Has data issue: false hasContentIssue false

New statistical calibration approach for Bruker AXS D8 Discover microdiffractometer with Hi-Star detector using GADDS software

Published online by Cambridge University Press:  29 February 2012

Ralph Rowe*
Affiliation:
Research Division, Canadian Museum of Nature, P.O. Box 3443, Stn. D, Ottawa, Ontario K1P 6P4, Canada
*
a)Electronic mail: [email protected]

Abstract

An additional statistical calibration for the Bruker D8 Discover microdiffractometer is necessary to obtain accurate reproducible 2θ data for cell-refinement work. This new approach uses a graphical mapping method of the 2θ error versus the location of a selected diffraction peak on the detector surface to describe the separate roles of different calibration procedures (rebiasing, flood field, and spatial corrections) and parameters (sample-to-detector distance, x-y center coordinate) in minimizing the error. Optimized parameters are used to obtain the lowest achievable Δ2θ with this setup. Intensity error relative to the position of the diffracted line on the detector was found to be consistent at up to 20% and could not be reduced using any of the investigated techniques and parameters.

Type
Crystallography Education
Copyright
Copyright © Cambridge University Press 2009

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Bruker AXS, Inc. (2001). GADDS Service/Administrator’s Manual (Manual Kit 662-002400) (Madison, Wisconsin).Google Scholar
He, B. B., Preckwinkel, U., and Smith, K. L. (2000). “Fundamentals of two-dimensional X-ray diffraction (XRD2)” in Advances in X-ray Analysis (JCPDS, International Center for Diffraction Data, Swarthmore, PA), Vol. 43, pp. 273280.Google Scholar
ICDD (1960). Powder Diffraction File No. 10-0173, Natl. Bur. Stand (U.S.), Circ. No. 539 (U.S. GPO, Washington, D.C.), Vol. 9, p. 3.Google Scholar
Jones, R. and Young, R. (2007). “Monthly Calibration of the Hi-Star Detector” (Bruker AXS, Madison, WI), Part No. M88-E01118.Google Scholar
Smith, K. L. (2001). “Pole Figures Using an Area Detector, Revision 2.1” (Bruker AXS, Madison, WI).Google Scholar
Zoltai, T. and Stout, J. H. (1984). ”Mineralogy: Concepts and Principles” (Burgess, Minneapolis), p. 505.Google Scholar