Hostname: page-component-586b7cd67f-dlnhk Total loading time: 0 Render date: 2024-11-28T04:05:25.770Z Has data issue: false hasContentIssue false

Use of a-SiC:H multilayer transducers for detection of fluorescence signals from reactive cyan and yellow fluorophores

Published online by Cambridge University Press:  10 August 2011

P. Louro
Affiliation:
Electronics Telecommunications and Computer Dept, ISEL, Lisbon, Portugal. CTS-UNINOVA, Lisbon, Portugal.
M. Vieira
Affiliation:
Electronics Telecommunications and Computer Dept, ISEL, Lisbon, Portugal. CTS-UNINOVA, Lisbon, Portugal. DEE-FCT-UNL, Quinta da Torre, Monte da Caparica, 2829-516, Caparica, Portugal
M. A. Vieira
Affiliation:
Electronics Telecommunications and Computer Dept, ISEL, Lisbon, Portugal. CTS-UNINOVA, Lisbon, Portugal.
J. Costa
Affiliation:
Electronics Telecommunications and Computer Dept, ISEL, Lisbon, Portugal. CTS-UNINOVA, Lisbon, Portugal.
M. Fernandes
Affiliation:
Electronics Telecommunications and Computer Dept, ISEL, Lisbon, Portugal. CTS-UNINOVA, Lisbon, Portugal.
A. Karmali
Affiliation:
CIEB-ISEL, Lisbon, Portugal
Get access

Abstract

The transducer consists of a p-i’(a-SiC:H)-n/p-i(a-Si:H)-n heterostructures produced by PECVD and optimized for the detection of the fluorescence resonance energy transfer between fluorophores with excitation in the violet(400 nm) and emissions in the cyan (470 nm) and yellow (588 nm) range of the spectrum. The thickness and the absorption coefficient of the i’- and i- layers were tailored for cyan and yellow optical confinement, respectively in the front and back photodiodes acting both as optical filters. The devices were characterized through transmittance and spectral response measurements and under different electrical.

To simulate the FRET pairs and the excitation light a chromatic time dependent combination of violet, cyan and yellow wavelengths was applied to the device. The generated photocurrent was measured under negative and positive bias to readout the combined spectra. The independent test signals were chosen in order to sample all the possible chromatic. Different wavelength backgrounds were also superimposed.

Results show that under negative bias the phorocurrent signal presents eight separate levels each one assigned to the different polychromatic mixtures. If a blue background is superimposed the yellow channel is enhanced and the cyan suppressed while under red irradiation the opposite behavior occurs. So under appropriated steady state optical bias the sensor will detect separately the cyan and yellow fluorescence pairs. An electrical model, supported by a numerical simulation, gives insight into the transduction mechanism.

Type
Research Article
Copyright
Copyright © Materials Research Society 2011

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

REFERENCES

1. LaVan, D.A, McGuire, Terry, Langer, Robert, Nat. Biotechnol. 21(10), 2003, 11841191.10.1038/nbt876Google Scholar
2. Grace, D., Medical Product Manufacturing News, 12, 2008, 2223.Google Scholar
3. Karmali, K., Karmali, A., Teixeira, A., Marcelo Curto, M. J.Assay of glucose oxidase from Penicillium amagasakiense and Aspergillus niger by Fourier Transform Infrared Spectroscopy”. (2004) Analytical Biochemistry 333, 320327.10.1016/j.ab.2004.06.025Google Scholar
4. Louro, P., Vieira, M., Vieira, M.A., Fernandes, M., Fantoni, A., Francisco, C., Barata, M., Physica E: Low-dimensional Systems and Nanostructures, 41 (2009) 10821085.10.1016/j.physe.2008.08.029Google Scholar
5. Louro, P., Vieira, M., Fernandes, M., Costa, J., Vieira, M. A., Caeiro, J., Neves, N., Barata, M., Phys. Status Solidi C 7, No. 3–4, 11881191 (2010).Google Scholar