Hostname: page-component-78c5997874-ndw9j Total loading time: 0 Render date: 2024-11-02T22:30:44.147Z Has data issue: false hasContentIssue false

Installing and Operating Fegtem's

Published online by Cambridge University Press:  10 February 2011

C. J. D. Hetherington
Affiliation:
Sheffield University, Department of Electronic and Electrical Engineering, Mappin Street, Sheffield, SI 3JD, UK
A. G. Cullis
Affiliation:
Sheffield University, Department of Electronic and Electrical Engineering, Mappin Street, Sheffield, SI 3JD, UK
S. Walker
Affiliation:
Sheffield University, Department of Electronic and Electrical Engineering, Mappin Street, Sheffield, SI 3JD, UK
J. Turner
Affiliation:
Lawrence Berkeley National Laboratory, National Center for Electron Microscopy, MS 72-150, Berkeley, CA 94720, USA
E.C. Nelson
Affiliation:
Lawrence Berkeley National Laboratory, National Center for Electron Microscopy, MS 72-150, Berkeley, CA 94720, USA
M.A. O'Keefe
Affiliation:
Lawrence Berkeley National Laboratory, National Center for Electron Microscopy, MS 72-150, Berkeley, CA 94720, USA
Get access

Abstract

The latest transmission electron microscopes with field emission guns and imaging filters now provide much of the microanalysis and imaging necessary in applications such as ULSI device development. The installation and operating environment of the instruments are critical to their successful operation. Information from two such installations is presented here, one in a purpose built facility and the other in an existing building. Ground vibration, acoustic noise, stray electromagnetic fields, air flow and temperature variation are considered, and the measures implemented to achieve desirable levels of each parameter are discussed. The physical layout of an installation is also shown.

Type
Research Article
Copyright
Copyright © Materials Research Society 1998

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Anderson, R.H., Design of the Electron Microscope Laboratory, volume 4 of Practical Methods in Electron Microscopy, edited by Glauert, A. M., North-Holland, Amsterdam, 1975.Google Scholar
2. O'Keefe, M. A., Ultramicroscopy 47 (1992) pp 2 82–29710.1016/0304-3991(92)90203-VGoogle Scholar
3. O'Keefe, M. A., Microscopy and Microanalysis, vol. 3, supp. 2, (1997) pp 11651166 Google Scholar
4. Turner, J.H., O'Keefe, M.A., Mueller, R., Microscopy and Microanalysis, vol. 3, supp. 2, (1997) pp 11771178 Google Scholar
5. Buffat, P.A., private communicationGoogle Scholar