Hostname: page-component-78c5997874-g7gxr Total loading time: 0 Render date: 2024-11-11T07:22:31.935Z Has data issue: false hasContentIssue false

Evaluation of AFM Probes and Instruments with Dynamic Cantilever Calibrator

Published online by Cambridge University Press:  30 January 2018

John Alexander
Affiliation:
SPM Labs LLC, TempeAZ85283USA
Sergey Belikov
Affiliation:
SPM Labs LLC, TempeAZ85283USA
Sergei Magonov*
Affiliation:
SPM Labs LLC, TempeAZ85283USA
Mark Smith
Affiliation:
AFM Services LLC, Santa BarbaraCA93110USA
*
Get access

Abstract

The main functions of the dynamic cantilever calibrator (DCC), which are related to characterization of AFM probes and instruments, are demonstrated on a variety of probes. The resonant frequency, Q-factor and spring constant of the rectangular and V-shaped probes were evaluated by thermal tune method. The inverse optical sensitivity and optical beam deflection noise, which define performance of AFM microscopes, were extracted from DCC data. Peculiarities of thermal tune studies and the use of DCC for advanced applications are discussed.

Type
Articles
Copyright
Copyright © Materials Research Society 2018 

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

REFERENCES

Butt, H.-J. and Jaschke, M., Nanotechnology 6, 1 (1995).Google Scholar
Kobayashi, K., Yamada, H. and Matsushige, K., Rev. Sci. Instrum. 80, 043708 (2009).CrossRefGoogle Scholar
Belikov, S., Alexander, J., Wall, C., Yermolenko, I., Magonov, S., and Malovichko, I., Proceed. American Control. Conf. 1009 (2014).Google Scholar
Timoshenko, S., Young, D. and Weaver, W. Jr., Vibration Problems in Engineering, 4th ed. (Wiley & Sons, New York, 1974).Google Scholar
Tikhonov, A. and Samarskii, A., Equations of Mathematical Physics, (Dover Publications, Meniola, New York, 2011).Google Scholar
Sader, J. E., Chon, J. W. M. and Mulvaney, P., Rev. Sci. Instrum. 70, 3967 (1999).Google Scholar
Belikov, S., Alexander, J. and Magonov, S., Proceed. Am. Contr. Conf. (2018), submitted.Google Scholar
Sader, J. E., Borgani, R., Gibson, C., Haviland, D. B., Higgins, M., Kirpatrick, J., Lu, J., Mulvaney, P., Shearer, C. J., Slattery, A. D., Thoren, P. A., Tran, J., Zhang, H., Zhang, H. and Zheng, T., Rev. Sci. Instrum. 87, 093711 (2016).Google Scholar