This is a phantom study to evaluate the dosimetry effects of using virtual bolus (VB) in TomoTherapy Treatment Planning System (TPS) optimisation for superficial planning target volume (PTV) that extends to the body surface. Without VB, the inverse-planning TPS will continuously boost the photon fluence at the surface of the superficial PTV due to lack of build-up region. VB is used during TPS optimisation only and will not be present in actual treatment delivery.
In this study, a dummy planning target was contoured on a cylindrical phantom which extends to the phantom surface, and VB of various combinations of thickness and density was used in treatment planning optimisation with TomoTherapy TPS. The plans were then delivered with the treatment modality TomoTherapy. Radiochromic films (Gafchromic EBT3) were calibrated and used for dose profiles measurements. TomoTherapy Planned-Adaptive software was used to analyse the delivered Dose-Volume Histograms (DVHs).
The use of 2 mm VB was not providing adequate build-up area and was unable to reduce the hot spots during treatment planning and actual delivery. The use of 4 mm VB was able to negate the photon fluence boosting effect by the TPS, and the actual delivery showed relatively small deviations from the treatment plan. The use of 6 mm VB caused significant dose overestimation by the TPS in the superficial regions resulting in insufficient dose coverage delivered.
VB with the combination of 4 mm thickness and 1·0 g/cc density provides the most robust solution for the TomoTherapy TPS optimisation of superficial PTV.