Published online by Cambridge University Press: 16 August 2018
To assess skin dose and incidence of skin reactions in early breast cancer patients treated via Intrabeam™ intraoperative radiation therapy (IORT) device.
In total, 250 breast cancer patients treated with a single fraction of 20 Gy using 50 kV photon were recruited. The applicator to skin distance (ASD) was measured before the initiation of the radiation and the skin dose in each patient was accordingly calculated based on the treatment planning system (TPS).
The average skin doses calculated were equal to 7·91, 5·83, 3·96 and 2·14 Gy for 6–10, 10–15, 15–20 and 20–30 mm ASD values, respectively. It is noticeable that the skin doses could be lower than the TPS measurements up to 45%, mostly due to lack of backscatter radiation in breast tissue compared with the full scatter condition in the Zeiss water phantom. Finally, only three patients showed low-grade skin reactions 1 week after IORT. A review of the related literature also revealed the incidence of lower skin complications among patients treated via Intrabeam™ compared with MammoSite™ machine.
The Intrabeam™ TPS did not seem to be very reliable for accurate skin dosimetry. However, breast cancer treatment using Intrabeam™ could result in fewer incidences of skin reactions than MammoSite™ machine.