Published online by Cambridge University Press: 23 August 2017
To characterise small photon beams using the Monte Carlo dose calculation algorithm for small field ranges in a heterogeneous medium.
An in-house phantom constructed with three different mediums, foam, polymethyl methacrylate and delrin resembling the densities of lung, soft tissue and bone respectively, was used in this study. Photon beam energies of 6 and 15 MV and field sizes of 8×8, 16×16, 24×24, 32×32 and 40×40 mm using X-ray voxel Monte Carlo (XVMC) algorithm using different detectors were validated. The relative output factor was measured in three different mediums having six different tissue interfaces; at the depth of 0, 1, 2 and 3 cm. The planar dose verification was undertaken using gafchromic films and considered dose at the lung and bone medium interfaces. For all the measurements, 104×104 mm was taken as the reference field size. The relative output factor for all other field sizes was taken and compared with planning system calculated values.
From field size 16×16 mm and above, the relative output factors were analysed in bone and soft tissue medium having lung as first medium. The maximum deviations were observed as 1·8 and 1·3% for 6 MV and 2·5 and 1·1% for 15 MV photon beams for bone and soft tissue, respectively. For lung as measurement medium, the maximum deviation of 14·8 and 19·2% were observed and having bone as first medium with 8×8 mm for 6 and 15 MV photon beams, respectively. The fluence verification of dose spectrum for the lung–bone interface scenarios with smaller field sizes were found within 2% of deviation with treatment planning system (TPS).
The accuracy of dose calculations for small field sizes in XVMC-based treatment planning algorithm was studied in different inhomogeneous mediums. It was found that the results correlated with measurement data for field size 16×16 mm and above. Noticeable deviation was observed for the smallest field size of 8×8 mm with interfaces of significant change in density. The observed results demands further analysis of work with smaller field sizes.