An experimental study of the vertical travel accuracy of the treatment table used in radiotherapy

Published online by Cambridge University Press:  09 August 2012

Hans L. Riis*
Affiliation:
Radiofysisk Laboratorium, Odense University Hospital, Odense C, Denmark
Sune J. Zimmermann
Affiliation:
Radiofysisk Laboratorium, Odense University Hospital, Odense C, Denmark
*
Correspondence to: Hans L. Riis, Radiofysisk Laboratorium, Odense University Hospital, Sdr. Boulevard 29, DK–5000 Odense C, Denmark. Tel: + 45 29 79 64 73. E-mail: [email protected]

Abstract

The idea of the present work was to investigate general features of the vertical travel of the treatment table in different configurations and the limitations of an optimised alignment. The investigations were carried out on two tables for different load cases, lateral positions and turntable angles. A wire was held vertically nearby the rotational axis of the table. The wire was used as reference in the investigations. A digital USB-microscope was attached to the tabletop. An orthogonal set of images of the wire was acquired at different vertical table positions. By analysing the images the vertical travel accuracy of the table was extracted. The two tables were found to travel linearly with the same characteristics over the full range of 110 cm with deviations less than 0.5 mm relative to a straight line. A divergence from a vertical travel above 0.5 mm was found to originate from misalignment. An alignment procedure to attain the optimal vertical performance with a minimum of inclination was presented. This procedure must be relayed on objective measured data of the travel with uncertainties far beyond 0.5 mm. Our suggested experimental method was found to have the potential to obtain the required data.

Type
Original Article
Copyright
Copyright © Cambridge University Press 2012

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Karger, CP, Hartmann, GH, Heeg, P, Jäkel, O. A method for determining the alignment accuracy of the treatment table axis at an isocentric irradiation facility. Phys Med Biol 2001; 46:N19N26.CrossRefGoogle ScholarPubMed
IEC. Radiotherapy equipment - Coordinates, movements and scales. IEC 61217; International Standard, International Electrotechnical Commission (IEC); Geneva; 1996. First edition.Google Scholar
Riis, HL, Zimmermann, SJ. Elekta Precise Table characteristics of IGRT remote table positioning. Acta Oncol 2009; 48:267270.CrossRefGoogle ScholarPubMed
Riis, HL, Zimmermann, SJ, Riis, P. Isocentric rotational performance of the Elekta Precise Table studied using a USB-microscope. Phys Med Biol 2010; 55:75977614.CrossRefGoogle ScholarPubMed
Kutcher, GJ, Coia, L, Gillin, M, Hanson, WF, Leibel, S, Morton, RJ, Palta, JR, Purdy, JA, Reinstein, LE, Svensson, GK, Weller, M, Wingfield, L. Comprehensive QA for radiation oncology: report of AAPM radiation therapy committee task group 40. Med Phys 1994; 21:581618.CrossRefGoogle ScholarPubMed