Intraoperative herniation of a tracheostomy tube cuff

Published online by Cambridge University Press:  29 June 2007

Yaron Bar-Lavie
Affiliation:
Department of Anaesthesia, Soroka Medical Centre, Faculty of Health Sciences, Ben-Gurion University of the Negev, Beer-Sheva, Israel
Albert Gatot
Affiliation:
Department Otolaryngology, Soroka Medical Centre, Faculty of Health Sciences, Ben-Gurion University of the Negev, Beer-Sheva, Israel
Ferit Tovi*
Affiliation:
Department Otolaryngology, Soroka Medical Centre, Faculty of Health Sciences, Ben-Gurion University of the Negev, Beer-Sheva, Israel
*
Address for correspondence: Dr Ferit Tovi, Department of Otolaryngology, Soroka Medical Centre, PO Box 151, Beer-Sheva 84101, Israel

Abstract

A case of intraoperative tracheostomy tube obstruction is reported. The clinical features and the chain of events leading to the diagnosis of cuff herniation are presented. The different mechanisms of herniation are discussed. In the present case we speculate that a manufacturing defect together with nitrous oxide diffusion into the cuff caused dilatation and herniation of the latter which led to obstruction of the distal tube lumen. We draw attention to this rare but life-threatening complication.

Type
Clinical Records
Copyright
Copyright © JLO (1984) Limited 1995

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Davidson, I., Zimmer, S. (1989) Cuff herniation. Anaesthesia 44: 938939.CrossRefGoogle ScholarPubMed
Dorsch, J., Dorsch, E. (1984) Understanding anesthesia equipment, 2nd Edition, Williams and Wilkins, Baltimore, USA, p 380.Google Scholar
Partridge, B. (1988) Nitrous oxide and endotracheal tube cuff leaks. Anesthesiology 68: 167168.CrossRefGoogle ScholarPubMed
Patterson, K., Keane, P. (1990) Missed diagnosis of cuff herniation in a modern nasal endotracheal tube. Anesthesia and Analgesia 71: 563564.CrossRefGoogle Scholar
Roland, P., Stovner, J. (1975) Brain damage following collapse of a ‘polyvinyl’ tube: elasticity and permeability of the cuff. Acta Anaesthesiologica Scandinavica 19: 303309.CrossRefGoogle ScholarPubMed
Stanley, T., Kawamura, R., Graves, C. (1974) Effects of nitrous oxide on volume and pressure of endotracheal tube cuffs. Anesthesiology 41: 256262.CrossRefGoogle ScholarPubMed
Viallard, M., Sok, S., Olivera, C., Dehghani, J., Riou, B. (1990) Respiratory tract obstruction caused by ballooning of an intubation tube cuff during nitrous oxide inhalation. Annales Francoises d'Anesthesie et de Reanimation 9(5): 460462.CrossRefGoogle ScholarPubMed