To outline the management options and neurological complications associated with repeated traumatic falls that cause intrusion of bone-anchored hearing aid abutments.
A three-year-old boy with coloboma, heart defects, atresia of nasal choanae, retarded growth, genital abnormalities, ear defects and deafness was fitted with a bone-anchored hearing aid for severe conductive hearing loss and congenital ear malformations. Six months later, a traumatic fall caused an intrusion injury which rendered the bone-anchored hearing aid abutment unusable. Without removing the original abutment, a second abutment was inserted on the same side to aid his hearing. Two years later, the child fell again and damaged his second bone-anchored hearing aid abutment. Having been offered a surgical option to repair the area, the parents opted to keep the abutments in situ.
Direct trauma to the fixture of a bone-anchored hearing aid is a relatively common long-term complication in children which can disrupt osseointegration and disable the implant. For young children who are either prone to falling or have behavioural problems, a bone-anchored hearing aid Softband may be more appropriate to non-invasively aid hearing.