Suction Regulators: A Potential Vector for Hospital-Acquired Pathogens

Published online by Cambridge University Press:  02 January 2015

Keith S. Kaye*
Affiliation:
Detroit Medical Center and, Wayne State University Health Center, Detroit, Michigan
Dror Marchaim
Affiliation:
Detroit Medical Center and, Wayne State University Health Center, Detroit, Michigan
Chester Smialowicz
Affiliation:
Deborah Heart and Lung Center, Browns Mills, New Jersey
Lauren Bentley
Affiliation:
Boehringer Laboratories, Phoenixville, Pennsylvania
*
Infection Prevention, Epidemiology and Antimicrobial Stewardship, Detroit Medical Center, University Health Center, 4201 Saint Antoine, Suite 2B, Box 331, Detroit, MI 48201 ([email protected])

Extract

The role of suction regulators in nosocomial infections has, to our knowledge, not been studied. A sampling of devices used in hospitals was conducted. Many regulators (173 [37%] of 470) were found to be colonized. A suction circuit model revealed that pathogens can disseminate throughout the circuit (retrograde and antegrade), colonizing an experimental patient stomach.

Type
Concise Communication
Copyright
Copyright © The Society for Healthcare Epidemiology of America 2010

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1.Vincent, JL, Relio, J, Marshall, J, et al. International study of the prevalence and outcomes of infection in intensive care units. JAMA 2009;302(21): 23232329.CrossRefGoogle Scholar
2.Clinical and Laboratory Standards Institute (CLSI). Performance Standards for Antimicrobial Susceptibility Testing: Sixteenth Informational Supplement. Wayne, PA: CLSI, 2006. CLSI document M100-S16.Google Scholar
3.Heyland, D, Mandeli, LA. Gastric colonization by gram-negative bacilli and nosocomial pneumonia in the intensive care unit patient: evidence for causation. Chest 1992;101(1): 187193.CrossRefGoogle ScholarPubMed
4.Edwards, JR, Peterson, KD, Andrus, ML, Dudeck, MA, Pollock, DA, Horan, TC. National Healthcare Safety Network (NHSN) report, data summary for 2006 through 2007, issued November 2008. Am J Infect Control 2008; 36(9):609626.CrossRefGoogle ScholarPubMed
5.Weber, DJ, Raasch, R, Rutala, WA. Nosocomial infections in the ICU: the growing importance of antibiotic-resistant pathogens. Chest 1999;115(suppl 3):34S41S.CrossRefGoogle ScholarPubMed
6.Hidron, AI, Edwards, JR, Patel, J, et al. NHSN annual update: antimicrobial-resistant pathogens associated with healthcare-associated infections: annual summary of data reported to the National Healthcare Safety Network at the Centers for Disease Control and Prevention, 2006-2007. Infect Control Hosp Epidemiol 2008;29(11): 9961011.CrossRefGoogle Scholar
7.Craven, DE, Steger, KA, Barat, LM, Duncan, RA. Nosocomial pneumonia: epidemiology and infection control. Intensive Care Med 1992;18(suppl 1): S3S9.CrossRefGoogle ScholarPubMed
8.Rutala, WA, Weber, DJ. Guidelines for Disinfection and Sterilization in Healthcare Facillities, 2008. Atlanta, GA: Centers for Disease Control and Prevention, 2008.Google Scholar