Hostname: page-component-cd9895bd7-7cvxr Total loading time: 0 Render date: 2024-12-12T11:41:56.041Z Has data issue: false hasContentIssue false

Some Factors Affecting the Accuracy of Stellar Photometry with CCDs (And some ways of dealing with them)

Published online by Cambridge University Press:  14 August 2015

Peter B. Stetson*
Affiliation:
Dominion Astrophysical Observatory, Herzberg Institute of Astrophysics, National Research Council, 5071 West Saanich Road, Victoria, British Columbia V8X 4M6 / Canada

Abstract

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the ‘Save PDF’ action button.

There are many factors which make it difficult to relate instrumental CCD photometry to a fundamental standard system with an accuracy much better than about 1%. Here I will address only three of them: (1) infrared leaks in the filters; (2) the finite opening and closing times of mechanical shutters; and (3) changes in the air mass for long integrations. I will be approaching these subjects from the point of view of a visiting astronomer at someone else’s observatory, who gets three or four nights of observing time a year, and has only the afternoons preceding those nights to perform whatever tests can be carried out while the equipment sits on the telescope.

Type
Joint Commission Meeting
Copyright
Copyright © Kluwer 1989