Study of backward terahertz radiation from intense picosecond laser–solid interactions using a multichannel calorimeter system

Published online by Cambridge University Press:  22 January 2019

H. Liu
Affiliation:
Beijing National Laboratory for Condensed Matter Physics, Institute of Physics, Chinese Academy of Sciences, Beijing 100190, China School of Physical Sciences, University of Chinese Academy of Sciences, Beijing 100049, China
G.-Q. Liao
Affiliation:
Key Laboratory for Laser Plasmas (MoE) and School of Physics and Astronomy, Shanghai Jiao Tong University, Shanghai 200240, China
Y.-H. Zhang
Affiliation:
Beijing National Laboratory for Condensed Matter Physics, Institute of Physics, Chinese Academy of Sciences, Beijing 100190, China School of Physical Sciences, University of Chinese Academy of Sciences, Beijing 100049, China
B.-J. Zhu
Affiliation:
Beijing National Laboratory for Condensed Matter Physics, Institute of Physics, Chinese Academy of Sciences, Beijing 100190, China School of Physical Sciences, University of Chinese Academy of Sciences, Beijing 100049, China
Z. Zhang
Affiliation:
Beijing National Laboratory for Condensed Matter Physics, Institute of Physics, Chinese Academy of Sciences, Beijing 100190, China
Y.-T. Li*
Affiliation:
Beijing National Laboratory for Condensed Matter Physics, Institute of Physics, Chinese Academy of Sciences, Beijing 100190, China School of Physical Sciences, University of Chinese Academy of Sciences, Beijing 100049, China Collaborative Innovation Center of IFSA (CICIFSA), Shanghai Jiao Tong University, Shanghai 200240, China
G. G. Scott
Affiliation:
Central Laser Facility, Rutherford Appleton Laboratory, Didcot, Oxfordshire OX11 0QX, UK
D. Rusby
Affiliation:
Central Laser Facility, Rutherford Appleton Laboratory, Didcot, Oxfordshire OX11 0QX, UK Department of Physics, SUPA, University of Strathclyde, Glasgow G4 0NG, UK
C. Armstrong
Affiliation:
Central Laser Facility, Rutherford Appleton Laboratory, Didcot, Oxfordshire OX11 0QX, UK Department of Physics, SUPA, University of Strathclyde, Glasgow G4 0NG, UK
E. Zemaityte
Affiliation:
Central Laser Facility, Rutherford Appleton Laboratory, Didcot, Oxfordshire OX11 0QX, UK Department of Physics, SUPA, University of Strathclyde, Glasgow G4 0NG, UK
P. Bradford
Affiliation:
Department of Physics, York Plasma Institute, University of York, Heslington, York YO10 5DD, UK
N. Woolsey
Affiliation:
Department of Physics, York Plasma Institute, University of York, Heslington, York YO10 5DD, UK
P. Huggard
Affiliation:
Space Science Department, STFC Rutherford Appleton Laboratory, Didcot OX11 0QX, UK
P. McKenna
Affiliation:
Department of Physics, SUPA, University of Strathclyde, Glasgow G4 0NG, UK
D. Neely*
Affiliation:
Central Laser Facility, Rutherford Appleton Laboratory, Didcot, Oxfordshire OX11 0QX, UK Department of Physics, SUPA, University of Strathclyde, Glasgow G4 0NG, UK
*
Correspondence to: Y.-T. Li, Beijing National Laboratory of Condensed Matter Physics, Institute of Physics, Chinese Academy of Sciences, Beijing 100190, China; D. Neely, Central Laser Facility, STFC Rutherford Appleton Laboratory, Didcot OX11 0QX, UK. Email: [email protected] (Y.-T. Li); [email protected] (D. Neely)
Correspondence to: Y.-T. Li, Beijing National Laboratory of Condensed Matter Physics, Institute of Physics, Chinese Academy of Sciences, Beijing 100190, China; D. Neely, Central Laser Facility, STFC Rutherford Appleton Laboratory, Didcot OX11 0QX, UK. Email: [email protected] (Y.-T. Li); [email protected] (D. Neely)

Abstract

A multichannel calorimeter system is designed and constructed which is capable of delivering single-shot and broad-band spectral measurement of terahertz (THz) radiation generated in intense laser–plasma interactions. The generation mechanism of backward THz radiation (BTR) is studied by using the multichannel calorimeter system in an intense picosecond laser–solid interaction experiment. The dependence of the BTR energy and spectrum on laser energy, target thickness and pre-plasma scale length is obtained. These results indicate that coherent transition radiation is responsible for the low-frequency component (${<}$1 THz) of BTR. It is also observed that a large-scale pre-plasma primarily enhances the high-frequency component (${>}$3 THz) of BTR.

Type
Research Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution, and reproduction in any medium, provided the original work is properly cited.
Copyright
© The Author(s) 2019

1 Introduction

Intense terahertz (THz) radiation sources have attracted increasing research interests due to their applications in coherent and incoherent control of matter, light and electron beam manipulation[Reference Kampfrath, Tanaka and Nelson1Reference Symes, Neely, Collier, Jaroszynski and Dumitras4]. With terawatt and petawatt laser systems[Reference Danson, Hillier, Hopps and Neely5], THz radiation from high intensity laser–solid interactions has been demonstrated as a new intense THz source[Reference Li, Liao, Zhou, Du, Ma, Li, Wang, Sheng, Chen and Zhang6Reference Gopal, Singh, Herzer, Reinhard, Schmidt, Dillner, May, Meyer, Ziegler and Paulus11]. At present, the reported energy of intense laser–solid interaction-based THz sources[Reference Symes, Neely, Collier, Jaroszynski and Dumitras4, Reference Gopal, Singh, Herzer, Reinhard, Schmidt, Dillner, May, Meyer, Ziegler and Paulus11] has been at the same level as sources based on traditional accelerators[Reference Wu, Fisher, Goodfellow, Fuchs, Daranciang, Hogan, Loos and Lindenberg12] and optical crystals[Reference Vicario, Ovchinnikov, Ashitkov, Agranat, Fortov and Hauri13].

Terahertz radiation is generated from both target front and rear surfaces in intense laser–solid interaction[Reference Li, Liao, Zhou, Du, Ma, Li, Wang, Sheng, Chen and Zhang6, Reference Liao, Li, Zhang, Liu, Ge, Yang, Wei, Yuan, Deng, Zhu, Zhang, Wang, Sheng, Chen, Lu, Ma, Wang and Zhang8]. THz radiation from target front surface propagates backward relative to the incident laser, which is referred as backward THz radiation (BTR) in this paper. And similarly, THz radiation from target rear surface is referred as forward THz radiation (FTR). There are mainly two mechanisms which account for FTR generation, including the target normal sheath acceleration radiation[Reference Gopal, Herzer, Schmidt, Singh, Reinhard, Ziegler, Brommel, Karmakar, Gibbon, Dillner, May, Meyer and Paulus10] and coherent transition radiation[Reference Liao, Li, Zhang, Liu, Ge, Yang, Wei, Yuan, Deng, Zhu, Zhang, Wang, Sheng, Chen, Lu, Ma, Wang and Zhang8]. However, the BTR generation mechanism still needs further clarification. A number of sources and processes such as fast electrons and plasma waves can produce electromagnetic pulses in intense laser–plasma interactions[Reference Li, Liao, Zhou, Du, Ma, Li, Wang, Sheng, Chen and Zhang6, Reference Liao, Li, Li, Su, Zheng, Liu, Wang, Hu, Yan, Dunn, Nilsen, Hunter, Liu, Wang, Chen, Ma, Lu, Jin, Kodama, Sheng and Zhang7, Reference Sheng, Mima, Zhang and Sanuki14Reference Hamster, Sullivan, Gordon, White and Falcone16]. Several experiments have shown that pre-plasma scale length plays an important role in BTR generation[Reference Li, Liao, Zhou, Du, Ma, Li, Wang, Sheng, Chen and Zhang6, Reference Liao, Li, Li, Su, Zheng, Liu, Wang, Hu, Yan, Dunn, Nilsen, Hunter, Liu, Wang, Chen, Ma, Lu, Jin, Kodama, Sheng and Zhang7, Reference Li, Zhou, Ding, Du, Liu, Li, Wang, Sheng, Ma, Chen, Lu, Dong, Wang, Lou, Shi, Wei and Zhang17]. For pre-plasma with large scale length, THz radiation can be generated through the linear mode conversion (LMC) mechanism[Reference Sheng, Mima, Zhang and Sanuki14, Reference Sheng, Mima and Zhang18]. For pre-plasma with small scale length, the time-evolving directional surface fast electron currents (SFE) are thought to contribute to the BTR[Reference Li, Li, Zhou, Wang, Du, Ding, Lin, Liu, Sheng, Peng, Chen, Ma, Lu, Wang, Wei and Zhang19]. Sagisaka et al. proposed the antenna model which attributes the BTR to a fast radial current confined within the target surface[Reference Sagisaka, Daido, Nashima, Orimo, Ogura, Mori, Yogo, Ma, Daito, Pirozhkov, Bulanov, Esirkepov, Shimizu and Hosoda20]. Coherent transition radiation (CTR) by laser-produced currents at the target front surface is also proposed as an effective mechanism for BTR generation[Reference Ding, Sheng and Koh15]. However, no crucial experimental evidence has been found so far. It should be noted that most of the above mechanisms are demonstrated in intense femtosecond laser–solid interactions experiments. Therefore, these mechanisms may not apply to BTR generated in interactions between high-energy picosecond laser pulse and solid targets.

It has been found that the high-frequency component ( ${>}$ 10 THz) is dominant in backward THz radiation due to LMC[Reference Liao, Li, Li, Su, Zheng, Liu, Wang, Hu, Yan, Dunn, Nilsen, Hunter, Liu, Wang, Chen, Ma, Lu, Jin, Kodama, Sheng and Zhang7]. However, the CTR or SFE mainly accounts for the BTR at low frequencies ( ${<}$ 3 THz)[Reference Ding, Sheng and Koh15, Reference Li, Li, Zhou, Wang, Du, Ding, Lin, Liu, Sheng, Peng, Chen, Ma, Lu, Wang, Wei and Zhang19]. Therefore, spectral measurement of the emitted THz wave is necessary to determine the BTR generation mechanism. However, the existing THz spectrum measurement technology is not ideally applicable for THz sources based on intense laser–plasma interactions (ILPI). This is because the repetition rate of many such sources is quite low at present, typically below 1 Hz. Also, the fluctuations between shots can be large in real experiments. As a result, multi-shot scanning-based methods are quite inefficient and potentially inaccurate when used to characterize ILPI-based sources, such as electro-optic sampling[Reference Wu and Zhang21] and autocorrelation measurement based on a Michelson interferometer[Reference Zhang, Wang, Wu, Zhang, Zhang, Zhang, Zhang, Sheng and Zhang22]. Moreover, the working bandwidth of general electro-optic crystals also limits the application of electro-optic sampling techniques. The bandwidth of BTR generated in intense laser–plasma interactions can reach tens of THz[Reference Li, Liao, Zhou, Du, Ma, Li, Wang, Sheng, Chen and Zhang6, Reference Liao, Li, Li, Su, Zheng, Liu, Wang, Hu, Yan, Dunn, Nilsen, Hunter, Liu, Wang, Chen, Ma, Lu, Jin, Kodama, Sheng and Zhang7]. However, the effective bandwidth is only 5.3 THz for ZnTe and 7 THz for GaP[Reference Wu and Zhang23]. In addition, the existing frequency-domain methods also share this disadvantage such as THz spectrometers based on prism[Reference Meijer, Pijpers, Nienhuys, Bonn and van der Zande24] and consecutive dispersion grating[Reference Wesch, Schmidt, Behrens, Delsim-Hashemi and Schmüser25].

In this paper, an eight-channel calorimeter system has been designed and constructed to deliver single-shot spectral measurements of BTR generated in an intense picosecond laser–solid interaction experiment. The system has components including pyroelectric detectors, high resistance float zone silicon (HRFZ-Si) THz beam splitters and THz band pass filters. In the experiment, the driving laser energy, target thickness and pre-pulse condition are varied. It is found that the transition radiation by refluxing fast electrons contributes greatly to the low-frequency component ( ${<}$ 1 THz) of BTR. The large-scale pre-plasma plays an important role in the generation of the high-frequency component ( ${>}$ 3 THz).

2 Design and calibration of multichannel calorimeter system

A multichannel calorimeter system consisting of pyroelectric detectors has been used to obtain spectral information of broad-band plasma-based THz sources in one laser shot[Reference Li, Liao and Li26]. Coherent THz radiation in two filtered bands was measured to the bunch length in Tsinghua Thomson scattering X-ray source[Reference Su, Yan, Du, Zhang, Zhou, Wang, Zheng, Tian, Huang and Tang27]. In this paper, the calorimeter system has eight channels and each channel contains an HRFZ-Si THz beam splitter, a THz band pass filter and a pyroelectric detector, as shown in Figure 1. The incoming THz beam is divided and guided into each channel by seven HRFZ-Si THz beam splitters. In each channel, a THz band pass filter is set in front of the pyroelectric detector. As a result, the calorimeter system can sample the THz radiation in eight different bands in a single laser shot.

Figure 1. Schematic layout of eight-channel THz calorimeter system.

The flat transmission curves of the HRFZ-Si beam splitter and responsivity of the pyroelectric detector enable the calorimeter system to have a quite broad sensitivity in the THz band. HRFZ-Si is an isotropic crystalline material which is suitable to serve as a THz beam splitter material[28]. In the calorimeter, 1 mm-thick HRFZ-Si wafers with both sides polished are used as the single-pass THz beam splitters. The transmission of 1 mm-thick HRFZ-Si wafer was calibrated using a Fourier transform infrared (FTIR) spectrometer in the previous work[Reference Li, Liao, Zhou, Du, Ma, Li, Wang, Sheng, Chen and Zhang6Reference Liao, Li, Zhang, Liu, Ge, Yang, Wei, Yuan, Deng, Zhu, Zhang, Wang, Sheng, Chen, Lu, Ma, Wang and Zhang8]. According to the calibration results, it maintains around 54% transmission from 1.2 to $1000~\unicode[STIX]{x03BC}\text{m}$ . The pyroelectric detector in our calorimeter system consists of a single lithium tantalite ( $\text{LiTaO}_{3}$ ) sensor of surface area $4~\text{mm}^{2}$ , which has a broad spectral response from 0.1 to $1000~\unicode[STIX]{x03BC}\text{m}$ [29]. The original signal from the sensor is coupled to an amplifying and readout circuit.

The sensitivity of the pyroelectric detector in the near-infrared band is calibrated with a Ti:sapphire laser, which outputs 100 fs pulses at 10 Hz with a wavelength centered around 800 nm. The energy of the laser pulse incident onto the sensor of the detector is varied and measured by a calibrated power meter. As shown in Figure 2, there is a high linearity between the output signal and the input laser energy, with a sensitivity of the detector being $1666\pm 10~\text{mV}/\unicode[STIX]{x03BC}\text{J}$ for the 800 nm laser. The detector’s sensitivity is also calibrated in the THz band. THz pulses are emitted from a lithium niobate ( $\text{LiNbO}_{3}$ ) crystal when pumped with 800 nm pulse-front-tilted laser[Reference Hafez, Chai, Ibrahim, Mondal, Férachou, Ropagnol and Ozaki2]. The spectrum of emitted THz waves is broad from 0.1 to 1 THz as measured with an electro-optic sampler. The energy of emitted THz pulses is measured using a calibrated pyroelectric detector[Reference Wu, Ma, Zhang, Chai, Fang, Xia, Kong, Wang, Liu, Zhu, Wang, Ruan and Li30], whose sensitivity is $25\pm 0.875~\text{mV}/\unicode[STIX]{x03BC}\text{J}$ . With cross calibration, the sensitivity of our detector at 0.1–1 THz is measured to be $898\pm 11~\text{mV}/\unicode[STIX]{x03BC}\text{J}$ . A ratio of sensitivity in near-infrared and THz band of 1.86 is calculated. The relatively low sensitivity of the pyroelectric detector in the THz band is probably due to the decreased efficiency of power absorption in the coating on the pyroelectric crystal. Huggard et al. [Reference Huggard, Zeuner, Goller, Lengfellner and Prettl31] show that the coating can have an increased reflectivity in the far-infrared and THz band.

Figure 2. Linear response of the detector to energy of near-infrared and THz wavelengths.

The band pass filters used in the calorimeter are fabricated from gold-mesh frequency-selective surfaces[32]. Eight filters were installed with center frequencies equal to 0.5 THz, 1 THz, 1.5 THz, 3 THz, 6 THz, 10 THz, 20 THz and 30 THz, respectively. Each filter has its own working band with peak transmission above 70% at the center wavelength which has also been calibrated using FTIR spectrometer. The eight filters share negligible transmission in the overlapping regions between each other.

3 BTR measurements in an intense picosecond laser–solid interaction experiment

With all components ready, the eight-channel calorimeter system is constructed as shown in Figure 3(a). An eight-channel power supply is made for the pyroelectric detectors. The signals from the detectors are recorded by two 500 MHz oscilloscopes. The eight-channel calorimeter system is used in an intense laser–solid interaction experiment to characterize the THz radiation spectrum from the target front surface. The experiment is performed at the Vulcan laser facility[Reference Danson, Collier, Neely, Barzanti, Damerell, Edwards, Hutchinson, Key, Norreys and Pepler33] of the Rutherford Appleton Laboratory, UK. A 1053 nm laser pulse with energy up to 100 J and duration 1 ps is best focused onto copper foil targets at an intensity of $10^{20}~\text{W}/\text{cm}^{2}$ at an incident angle of $30^{\circ }$ , as shown in Figure 3(a). THz emission generated from the front surface of the target is collected by a 50 mm diameter TPX lens placed at an angle of $60^{\circ }$ from the target normal direction. The beam is reduced to 25 mm diameter using two off-axis parabolic mirrors with focal length 200 mm and 100 mm, respectively. The reduced beam is guided into the eight-channel calorimeter through a 50 mm TPX window. The transmission of the TPX lens and window is also measured with the FTIR spectrometer. In one of the eight channels, none filter was added so that the total energy of the THz radiation could be monitored. The remaining seven channels are equipped with band pass filters whose central frequencies are 0.5 THz, 1 THz, 1.5 THz, 3 THz, 6 THz, 10 THz and 20 THz, respectively. To avoid any disturbance of the calorimeters by pickup of electromagnetic pulses[Reference Mead, Neely, Gauoin, Heathcote and Patel34, Reference Bradford, Woolsey, Scott, Liao, Liu, Zhang, Zhu, Armstrong, Astbury, Brenner, Brummitt, Consoli, East, Gray, Haddock, Huggard, Jones, Montgomery, Musgrave, Oliveira, Rusby, Spindloe, Summers, Zemaityte, Zhang, Li, McKenna and Neely35] and high-energy ionizing radiation such as X-rays from the laser–plasma interaction, the calorimeter system is well shielded with a metal box and lead sheets.

Figure 3. (a) Experimental setup of THz radiation spectrum measurement in ultra-intense laser–plasma interaction experiment. (b) Spectral sensitivity of the multichannel calorimeter system.

The spectral sensitivity of the system is necessary to obtain the THz spectrum from the measured signals. The spectral sensitivity mainly depends on the transmission properties of the THz optics used in the system and the spectral responsivity of the pyroelectric detector. Each band pass filter has a narrow passing band. In the narrow band, the transmissions of the band pass filter, HRFZ-Si beam splitter, TPX window and lens, as well as the spectral sensitivity of the pyroelectric detector, are simplified to be constant. Considering all the factors, the spectral sensitivity of the multichannel calorimeter system $S(f_{i})$ can be written as

(1) $$\begin{eqnarray}\displaystyle S(f_{i}) & = & \displaystyle T_{\text{lens}}(f_{i})\cdot T_{\text{window}}(f_{i})\cdot E_{i}\nonumber\\ \displaystyle & & \displaystyle \cdot T_{\text{BPF}}(f_{i})\cdot S_{\text{detector}}(f_{i}).\end{eqnarray}$$

Here, $i$ is the channel number from 1 to 8, $f_{i}$ is the central frequency of band pass filter located in the $i$ th channel, $T_{\text{lens}}(f_{i})$ and $T_{\text{window}}(f_{i})$ are the transmissions of the TPX lens and window at $f_{i}$ , respectively, $E_{i}$ is the relative energy portion allocated to the $i$ th channel determined by the layout of HRFZ-Si beam splitters, $T_{\text{BPF}}(f_{i})$ is the transmission of the band pass filter at $f_{i}$ , and $S_{\text{detector}}(f_{i})$ is the relative sensitivity of the detector at $f_{i}$ . Table 1 lists the factors for the system spectral sensitivity calculation which is based on the calibration of the THz detectors and other components. With the system spectral sensitivity and measured signals, the THz radiation spectral intensity $I\!(f_{i})$ is inversed as

(2) $$\begin{eqnarray}\displaystyle I\!(f_{i})=V_{i}/[E_{\text{laser}}\cdot S(f_{i})\cdot W_{i}]. & & \displaystyle\end{eqnarray}$$

Here, the $E_{\text{laser}}$ is shot energy of the intense laser which is used to normalize the spectrum to reduce the effects of shot-to-shot fluctuations. $W_{i}$ is the bandwidth of the pass filter in the $i$ th channel.

Table 1. Factors for the spectral sensitivity of the multichannel system.

Figure 3(b) shows the calculated spectral sensitivity of the calorimeter system. The system has a broad sensitivity from 0.5 to 20 THz, which is advantageous for spectral measurement of the THz radiation. Combining all the factors, the system has relatively high sensitivity at 3 THz and 6 THz. The low sensitivity at 0.5 THz is mainly due to the low responsivity of the pyroelectric detector to low-frequency part. And the system is less sensitive at high frequencies such as 10 THz and 20 THz because of the absorption of TPX material. There are several factors which may cause inaccuracy of the THz spectrum measurement using the system. The fluctuations of shot spots could lead to a direction instability of THz beam. This will lead to the deviation of a portion of the original THz beam into each channel. This deviation can be reduced by applying a collimating lens with longer focal length or decreasing the distance between the instrument and the target. The use of TPX lens and window causes huge absorption for frequencies large than 10 THz, which would modify the original spectrum. It is preferred to utilize materials with flat transmission such as HRFZ-Si or reflection-style optics like off-axis parabola mirror. In addition, other factors like the diffraction and air absorption loss also affect the system spectral sensitivity. Careful calibration for these factors contributes to highly accurate system sensitivity.

4 Experimental results

Figure 4(a) shows the dependence of the BTR energy from a $5~\unicode[STIX]{x03BC}\text{m}$ copper target driven with a laser energy varying from 0 to 54 J. The BTR energy increases linearly as the driven energy increases. Figure 4(b) shows the measured BTR spectrum using the multichannel calorimeter system under different pump laser energy conditions. The spectrum of BTR is quite broad and the low-frequency component ( ${<}$ 3 THz) has a relatively high spectral intensity. This helps us to conclude that the low-frequency component is laser intensity-dependent, especially those lower than 1 THz. The higher driven energy means higher focal intensity on the target surface, which enhances the fast electrons contributing to terahertz radiation. But the spectral intensity of the high-frequency component changes little, indicating the high frequency THz radiation is not sensitive to the laser intensity. This is an evidence that the generation mechanism of the low- and high-frequency components is different.

Figure 4. (a) The measured BTR energy from a $5~\unicode[STIX]{x03BC}\text{m}$ copper target as a function of the pump laser energy. (b) The measured BTR spectrum with driven laser energy on target varying from 14 J to 54 J from a $5~\unicode[STIX]{x03BC}\text{m}$ copper target.

The dependence of BTR energy and spectrum on the thickness of copper foil target is shown in Figures 5(a) and 5(b). The intensity of low-frequency components increases as the target thickness rises from $1~\unicode[STIX]{x03BC}\text{m}$ to $20~\unicode[STIX]{x03BC}\text{m}$ . As the thickness is further increased to $100~\unicode[STIX]{x03BC}\text{m}$ , the low-frequency component ( ${<}$ 1 THz) decreases significantly. The total BTR energy shows a similar dependence on target thickness.

Figure 5. (a) The measured BTR energy as a function of copper foil target thickness. (b) The measured BTR spectrum with copper foil target thickness varying from $1~\unicode[STIX]{x03BC}\text{m}$ to $100~\unicode[STIX]{x03BC}\text{m}$ .

The effect of pre-plasma on the BTR generation is also studied during our experiment. A 1% portion of the main laser beam is extracted and co-linearly injected as a pre-pulse with a variable delay between the pre-pulse and main pulse. The pre-pulse will generate pre-plasma and the plasma will expand during the delay time. With longer delay between pre-pulse and main pulse, pre-plasma with larger scale length is formed. The dependence of BTR energy on the delay is shown as Figure 6(a). The BTR energy increases as the pre-plasma scale length grows. The BTR spectrum with different delays is as shown in Figure 6(b). When the delay increases from 0 to 400 ps, more BTR of low frequency is generated. When the delay further increases to 1000 ps, the interesting thing is that the intensity of the low-frequency component ( ${<}$ 1 THz) decreases slightly and the higher-frequency component ( ${>}$ 3 THz) is enhanced.

Figure 6. (a) The measured BTR energy from a $100~\unicode[STIX]{x03BC}\text{m}$ copper target as a function of delay between pre-pulse and main pulse. (b) The measured BTR spectrum from a $100~\unicode[STIX]{x03BC}\text{m}$ copper target with different delays between pre-pulse and main pulse, no pre-pulse, 400 ps and 1000 ps, respectively.

5 Discussion

The dominant component of backward terahertz radiation in our experiment is lower than 1 THz in frequency, which can be interpreted with the coherent transition radiation mechanism. According to the CTR theory, the wavelength $\unicode[STIX]{x1D706}$ of coherent THz radiation is longer than the longitudinal length of electron bunches $\unicode[STIX]{x1D70E}_{ez}$ . In our case, $\unicode[STIX]{x1D70E}_{ez}$ is estimated to be $c\cdot \unicode[STIX]{x1D70F}_{L}\sim 300~\unicode[STIX]{x03BC}\text{m}$ , where $\unicode[STIX]{x1D70F}_{L}$ is the laser pulse length[Reference Liao, Li, Zhang, Liu, Ge, Yang, Wei, Yuan, Deng, Zhu, Zhang, Wang, Sheng, Chen, Lu, Ma, Wang and Zhang8]. As the driving laser energy increases, the number of fast electrons increases, which enhances the CTR[Reference Schroeder, Esarey, Van Tilborg and Leemans36]. PIC simulation has shown that both the laser-accelerated backward electron bunches and refluxing electrons contribute to the BTR generated through CTR[Reference Ding, Sheng and Koh15]. The laser-produced electron bunches account for the primary THz pulses. In intense laser–solid interactions, some of laser-accelerated electrons are recirculated inside the target by the sheath field which exists both on the target front and rear surfaces[Reference Quinn, Yuan, Lin, Carroll, Tresca, Gray, Coury, Li, Li, Brenner, Robinson, Neely, Zielbauer, Aurand, Fils, Kuehl and McKenna37]. And subsequent THz pulses are emitted when these refluxing electrons cross the vacuum-target interfaces.

The observed dependence of backward THz energy on the target thickness suggests the CTR by the refluxing electrons contributes to the BTR in our experiment. It has been found that the target thickness has a significant effect on the recirculation of fast electrons inside the conductive foil. As the target thickness increases, the time to complete a pass through the target will gradually increase and the total number of recirculations will decrease[Reference Quinn, Yuan, Lin, Carroll, Tresca, Gray, Coury, Li, Li, Brenner, Robinson, Neely, Zielbauer, Aurand, Fils, Kuehl and McKenna37, Reference Sentoku, Cowan, Kemp and Ruhl38]. In our experiment, the $20~\unicode[STIX]{x03BC}\text{m}$ thick copper target is optimum for BTR generation. The pre-pulse will ionize the target rear surface and affect the formation of sheath field when thin targets are used. This reduces the reflection effect of the sheath field. As a result, less BTR is measured with $1~\unicode[STIX]{x03BC}\text{m}$ and $5~\unicode[STIX]{x03BC}\text{m}$ thick targets. For thicker targets such as $50~\unicode[STIX]{x03BC}\text{m}$ and $100~\unicode[STIX]{x03BC}\text{m}$ , the decrease of BTR energy is due to the decrease of the number of transits of the refluxing fast electrons across the target.

Other mechanisms such as surface fast electron currents and LMC cannot account for the dependence of BTR on target thickness. In these models, only sources like currents and plasma waves related to the pre-plasma are considered. As the target gets thicker, BTR should not change because of the relatively constant pre-plasma condition according to the surface fast electron currents and LMC model. However, the energy of the BTR decreases a lot with the $100~\unicode[STIX]{x03BC}\text{m}$ foil target, especially the low-frequency component. In addition, THz radiation by LMC mechanism features a high-frequency ( ${>}$ 10 THz) component which disagrees with the measured BTR spectrum.

However, the LMC mechanism will become important as pre-plasma scale length increases. With a large-scale pre-plasma, laser absorption efficiency will be significantly improved by laser self-focusing and filament effects[Reference McKenna, Carroll, Lundh, Nürnberg, Markey, Bandyopadhyay, Batani, Evans, Jafer, Kar, Neely, Pepler, Quinn, Redaelli, Roth, Wahlström, Yuan and Zepf39, Reference Esirkepov, Koga, Sunahara, Morita, Nishikino, Kageyama, Nagatomo, Nishihara, Sagisaka, Kotaki, Nakamura, Fukuda, Okada, Pirozhkov, Yogo, Nishiuchi, Kiriyama, Kondo, Kando and Bulanov40]. In our experiment, the low-frequency component is significantly enhanced when a pre-pulse is added 400 ps ahead of the main pulse. This is attributed to the increase of the number of fast electrons due to improved laser absorption efficiency. When the scale length of pre-plasma further increases, the enhanced CTR by more fast electrons cannot come out of the large-scale plasma. However, large-scale pre-plasma is helpful for LMC[Reference Liao, Li, Li, Su, Zheng, Liu, Wang, Hu, Yan, Dunn, Nilsen, Hunter, Liu, Wang, Chen, Ma, Lu, Jin, Kodama, Sheng and Zhang7]. And this is why the low-frequency component ( ${<}$ 1 THz) is weakened and high- frequency component ( ${>}$ 3 THz) is enhanced when the delay is further increased to 1000 ps. However, considering the contribution of the high-frequency component ( ${>}$ 3 THz), the total energy of BTR still increases. This phenomenon indicates that the LMC mechanism is starting to make a measurable contribution under this condition.

6 Conclusion

A simple multichannel calorimeter system is designed and constructed, which consists of HRFZ-Si THz beam splitters, THz band pass filters and pyroelectric detectors. It provides a convenient single-shot solution to characterize the spectrum of broad-band THz radiation generated in an intense laser–plasma interaction. Spectral measurements of backward THz radiation are performed using the multichannel calorimeter system in an intense picosecond laser–solid interaction experiment. The low-frequency component ( ${<}$ 1 THz) is dominant for the BTR measured in the experiment. The BTR is attributed to coherent transition radiation by laser-accelerated fast electrons as they cross the plasma–vacuum surface. The dependence of BTR energy and spectrum on the target thickness indicates that CTR by refluxing fast electrons contributes to the BTR. The LMC mechanism starts to work with a large-scale pre-plasma, which enhances the intensity of the high-frequency components ( ${>}$ 3 THz) of the signal.

Acknowledgements

This work is supported by the Newton China–UK joint research grant on laser–ion acceleration and novel terahertz radiation, and EPSRC grant EP/K022415/1 on advanced laser–ion acceleration strategies toward next generation healthcare and EPSRC grant EP/R006202/1. This work is also supported by the National Natural Science Foundation of China (Nos. 11520101003 and 11861121001), the Strategic Priority Research Program of the Chinese Academy of Sciences (Nos. XDB16010200 and XDB07030300). G.Q.L. would like to acknowledge support from the National Postdoctoral Program for Innovative Talents (No. BX201600106). Data associated with research published in this paper can be accessed at: http://dx.doi.org/10.5286/edata/715.

References

Kampfrath, T. Tanaka, K. and Nelson, K. A. Nat. Photon. 7, 680 (2013).Google Scholar
Hafez, H. A. Chai, X. Ibrahim, A. Mondal, S. Férachou, D. Ropagnol, X. and Ozaki, T. J. Opt. 18, 093004 (2016).Google Scholar
Nanni, E. A. Huang, W. R. Hong, K. H. Ravi, K. Fallahi, A. Moriena, G. Miller, R. J. and Kartner, F. X. Nature Commun. 6, 8486 (2015).Google Scholar
Symes, D. R. Neely, D. Collier, J. L. Jaroszynski, D. A. and Dumitras, D. AIP Conf. Proc. 1228, 444 (2010).Google Scholar
Danson, C. Hillier, D. Hopps, N. and Neely, D. High Power Laser Sci. Eng. 3, e3 (2015).Google Scholar
Li, C. Liao, G. Q. Zhou, M. L. Du, F. Ma, J. L. Li, Y. T. Wang, W. M. Sheng, Z. M. Chen, L. M. and Zhang, J. Opt. Express 24, 4010 (2016).Google Scholar
Liao, G. Q. Li, Y. T. Li, C. Su, L. N. Zheng, Y. Liu, M. Wang, W. M. Hu, Z. D. Yan, W. C. Dunn, J. Nilsen, J. Hunter, J. Liu, Y. Wang, X. Chen, L. M. Ma, J. L. Lu, X. Jin, Z. Kodama, R. Sheng, Z. M. and Zhang, J. Phys. Rev. Lett. 114, 255001 (2015).Google Scholar
Liao, G. Q. Li, Y. T. Zhang, Y. H. Liu, H. Ge, X. L. Yang, S. Wei, W. Q. Yuan, X. H. Deng, Y. Q. Zhu, B. J. Zhang, Z. Wang, W. M. Sheng, Z. M. Chen, L. M. Lu, X. Ma, J. L. Wang, X. and Zhang, J. Phys. Rev. Lett. 116, 205003 (2016).Google Scholar
Liao, G. Q. Li, Y. T. Li, C. Liu, H. Zhang, Y. H. Jiang, W. M. Yuan, X. H. Nilsen, J. Ozaki, T. Wang, W. M. Sheng, Z. M. Neely, D. McKenna, P. and Zhang, J. Plasma Phys. Control. Fusion 59, 014039 (2017).Google Scholar
Gopal, A. Herzer, S. Schmidt, A. Singh, P. Reinhard, A. Ziegler, W. Brommel, D. Karmakar, A. Gibbon, P. Dillner, U. May, T. Meyer, H. G. and Paulus, G. G. Phys. Rev. Lett. 111, 074802 (2013).Google Scholar
Gopal, A. Singh, P. Herzer, S. Reinhard, A. Schmidt, A. Dillner, U. May, T. Meyer, H. G. Ziegler, W. and Paulus, G. G. Opt. Lett. 38, 4705 (2013).Google Scholar
Wu, Z. Fisher, A. S. Goodfellow, J. Fuchs, M. Daranciang, D. Hogan, M. Loos, H. and Lindenberg, A. Rev. Sci. Instrum. 84, 022701 (2013).Google Scholar
Vicario, C. Ovchinnikov, A. V. Ashitkov, S. I. Agranat, M. B. Fortov, V. E. and Hauri, C. P. Opt. Lett. 39, 6632 (2014).Google Scholar
Sheng, Z. M. Mima, K. Zhang, J. and Sanuki, H. Phys. Rev. Lett. 94, 095003 (2005).Google Scholar
Ding, W. J. Sheng, Z. M. and Koh, W. S. Appl. Phys. Lett. 103, 204107 (2013).Google Scholar
Hamster, H. Sullivan, A. Gordon, S. White, W. and Falcone, R. W. Phys. Rev. Lett. 71, 2725 (1993).Google Scholar
Li, C. Zhou, M. L. Ding, W. J. Du, F. Liu, F. Li, Y. T. Wang, W. M. Sheng, Z. M. Ma, J. L. Chen, L. M. Lu, X. Dong, Q. L. Wang, Z. H. Lou, Z. Shi, S. C. Wei, Z. Y. and Zhang, J. Phys. Rev. E 84, 036405 (2011).Google Scholar
Sheng, Z.-M. Mima, K. and Zhang, J. Phys. Plasmas 12, 123103 (2005).Google Scholar
Li, Y. T. Li, C. Zhou, M. L. Wang, W. M. Du, F. Ding, W. J. Lin, X. X. Liu, F. Sheng, Z. M. Peng, X. Y. Chen, L. M. Ma, J. L. Lu, X. Wang, Z. H. Wei, Z. Y. and Zhang, J. Appl. Phys. Lett. 100, 254101 (2012).Google Scholar
Sagisaka, A. Daido, H. Nashima, S. Orimo, S. Ogura, K. Mori, M. Yogo, A. Ma, J. Daito, I. Pirozhkov, A. S. Bulanov, S. V. Esirkepov, T. Z. Shimizu, K. and Hosoda, M. Appl. Phys. B 90, 373 (2008).Google Scholar
Wu, Q. and Zhang, X. C. Appl. Phys. Lett. 67, 3523 (1995).Google Scholar
Zhang, L.-L. Wang, W.-M. Wu, T. Zhang, R. Zhang, S.-J. Zhang, C.-L. Zhang, Y. Sheng, Z.-M. and Zhang, X.-C. Phys. Rev. Lett. 119, 235001 (2017).Google Scholar
Wu, Q. and Zhang, X. C. Appl. Phys. Lett. 70, 1784 (1997).Google Scholar
Meijer, A. S. Pijpers, J. J. H. Nienhuys, H. K. Bonn, M. and van der Zande, W. J. J. Opt. A 10, 095303 (2008).Google Scholar
Wesch, S. Schmidt, B. Behrens, C. Delsim-Hashemi, H. and Schmüser, P. Nucl. Instrum. Methods Phys. Res. A 665, 40 (2011).Google Scholar
Li, C. Liao, G.-Q. and Li, Y.-T. High Power Laser Sci. Eng. 3, e15 (2015).Google Scholar
Su, X. Yan, L. Du, Y. Zhang, Z. Zhou, Z. Wang, D. Zheng, L. Tian, Q. Huang, W. and Tang, C. Nucl. Instrum. Methods Phys. Res. B 402, 157 (2017).Google Scholar
Wu, X.-J. Ma, J.-L. Zhang, B.-L. Chai, S.-S. Fang, Z.-J. Xia, C.-Y. Kong, D.-Y. Wang, J.-G. Liu, H. Zhu, C.-Q. Wang, X. Ruan, C.-J. and Li, Y.-T. Opt. Express 26, 7107 (2018).Google Scholar
Huggard, P. G. Zeuner, S. Goller, K. Lengfellner, H. and Prettl, W. J. Appl. Phys. 75, 616 (1994).Google Scholar
Danson, C. N. Collier, J. Neely, D. Barzanti, L. J. Damerell, A. Edwards, C. B. Hutchinson, M. H. R. Key, M. H. Norreys, P. A. and Pepler, D. A. Opt. Acta Int. J. Opt. 45, 1653 (1998).Google Scholar
Mead, M. J. Neely, D. Gauoin, J. Heathcote, R. and Patel, P. Rev. Sci. Instrum. 75, 4225 (2004).Google Scholar
Bradford, P. Woolsey, N. C. Scott, G. G. Liao, G. Liu, H. Zhang, Y. Zhu, B. Armstrong, C. Astbury, S. Brenner, C. Brummitt, P. Consoli, F. East, I. Gray, R. Haddock, D. Huggard, P. Jones, P. J. R. Montgomery, E. Musgrave, I. Oliveira, P. Rusby, D. R. Spindloe, C. Summers, B. Zemaityte, E. Zhang, Z. Li, Y. McKenna, P. and Neely, D. High Power Laser Sci. Eng. 6, e21 (2018).Google Scholar
Schroeder, C. B. Esarey, E. Van Tilborg, J. and Leemans, W. P. Phys. Rev. E 69, 016501 (2004).Google Scholar
Quinn, M. N. Yuan, X. H. Lin, X. X. Carroll, D. C. Tresca, O. Gray, R. J. Coury, M. Li, C. Li, Y. T. Brenner, C. M. Robinson, A. P. L. Neely, D. Zielbauer, B. Aurand, B. Fils, J. Kuehl, T. and McKenna, P. Plasma Phys. Control. Fusion 53, 025007 (2011).Google Scholar
Sentoku, Y. Cowan, T. E. Kemp, A. and Ruhl, H. Phys. Plasmas 10, 2009 (2003).Google Scholar
McKenna, P. Carroll, D. C. Lundh, O. Nürnberg, F. Markey, K. Bandyopadhyay, S. Batani, D. Evans, R. G. Jafer, R. Kar, S. Neely, D. Pepler, D. Quinn, M. N. Redaelli, R. Roth, M. Wahlström, C. G. Yuan, X. H. and Zepf, M. Laser Part Beams 26, 591 (2008).Google Scholar
Esirkepov, T. Z. Koga, J. K. Sunahara, A. Morita, T. Nishikino, M. Kageyama, K. Nagatomo, H. Nishihara, K. Sagisaka, A. Kotaki, H. Nakamura, T. Fukuda, Y. Okada, H. Pirozhkov, A. S. Yogo, A. Nishiuchi, M. Kiriyama, H. Kondo, K. Kando, M. and Bulanov, S. V. Nucl. Instrum. Methods Phys. Res. A 745, 150 (2014).Google Scholar
Figure 0

Figure 1. Schematic layout of eight-channel THz calorimeter system.

Figure 1

Figure 2. Linear response of the detector to energy of near-infrared and THz wavelengths.

Figure 2

Figure 3. (a) Experimental setup of THz radiation spectrum measurement in ultra-intense laser–plasma interaction experiment. (b) Spectral sensitivity of the multichannel calorimeter system.

Figure 3

Table 1. Factors for the spectral sensitivity of the multichannel system.

Figure 4

Figure 4. (a) The measured BTR energy from a $5~\unicode[STIX]{x03BC}\text{m}$ copper target as a function of the pump laser energy. (b) The measured BTR spectrum with driven laser energy on target varying from 14 J to 54 J from a $5~\unicode[STIX]{x03BC}\text{m}$ copper target.

Figure 5

Figure 5. (a) The measured BTR energy as a function of copper foil target thickness. (b) The measured BTR spectrum with copper foil target thickness varying from $1~\unicode[STIX]{x03BC}\text{m}$ to $100~\unicode[STIX]{x03BC}\text{m}$.

Figure 6

Figure 6. (a) The measured BTR energy from a $100~\unicode[STIX]{x03BC}\text{m}$ copper target as a function of delay between pre-pulse and main pulse. (b) The measured BTR spectrum from a $100~\unicode[STIX]{x03BC}\text{m}$ copper target with different delays between pre-pulse and main pulse, no pre-pulse, 400 ps and 1000 ps, respectively.