Hostname: page-component-cd9895bd7-q99xh Total loading time: 0 Render date: 2024-12-18T22:49:08.079Z Has data issue: false hasContentIssue false

Can we improve the estimate of arterial PCO2 from end-tidal PCO2?

Published online by Cambridge University Press:  16 August 2006

R. Fletcher
Affiliation:
Directorate of Anaesthesia, Central Manchester Hospitals NHS Trust, Manchester, UK
F. Boris-Möller
Affiliation:
Directorate of Anaesthesia, Central Manchester Hospitals NHS Trust, Manchester, UK
Get access

Abstract

To use end-tidal PCO2 as a non-invasive estimate of arterial PCO2, one adds a nominal value, representing the arterial–end-tidal PCO2 difference. How much does one add? We hypothesized that, halving the ventilator rate and simultaneously doubling tidal volume, the immediate change in end-tidal PCO2 would be proportional to the original arterial-end-tidal PCO2 difference. We ventilated 31 patients at 20 breaths per minute (bpm), sampled arterial blood, and changed the rate to 10 bpm. The change in end-tidal PCO2 was, as hypothesized, positively correlated to the original arterial–end-tidal difference at 20 bpm (r = 0.64). End-tidal PCO2 increased in 23 patients. Thus, in theory, this method could offer some improvement in the estimate of arterial PCO2 from end-tidal. However, because of the considerable spread of values, a separate study is needed for verification

Type
Original Article
Copyright
2000 European Society of Anaesthesiology

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)