Hostname: page-component-586b7cd67f-dlnhk Total loading time: 0 Render date: 2024-11-24T11:17:46.760Z Has data issue: false hasContentIssue false

Transcatheter closure of coronary arterial fistulas using the new Amplatzer® vascular plug

Published online by Cambridge University Press:  20 April 2007

Gunther Fischer
Affiliation:
Klinik für Kinderkardiologie, Universitätsklinikum Schleswig-Holstein, Campus Kiel, Germany
Sotiria C. Apostolopoulou
Affiliation:
Onassis Cardiac Surgery Center, Athens, Greece
Spyros Rammos
Affiliation:
Onassis Cardiac Surgery Center, Athens, Greece
Maria Kiaffas
Affiliation:
Onassis Cardiac Surgery Center, Athens, Greece
Hans Heiner Kramer
Affiliation:
Klinik für Kinderkardiologie, Universitätsklinikum Schleswig-Holstein, Campus Kiel, Germany

Abstract

We report our initial experience in using the Amplatzer® vascular plug for closure of coronary arterial fistulas. The self-expanding, cylindrical, device is made from Nitinol wire mesh, and is available from 4 to 16 millimetres in diameter. We have now used the device to close fistulas in 3 patients, aged from 3 to 14 years, who presented with ratios of pulmonary-to-systemic flow from 1.5 to 3. In 2 patients, fistulas arising from the proximal right and left coronary arteries, with maximal diameters of 9 and 10 millimetres, respectively, had their narrowest diameter, of 6 millimetres, proximal to the entrance into the right atrium via a saccular aneurysm. The third fistula, with a maximal diameter of 16 millimetres, and with its origin from the circumflex coronary artery, entered the right atrium with nearly unrestricted flow, its narrowest diameter being 8 millimetres. For interventional closure, we chose plugs twice the diameter of the narrowest segment of the fistula, thus using 2 devices of 12 millimetres and one of 16 millimetres diameter. An arteriovenous loop was established through the fistula by snaring an exchange guide wire. Using a 7 or 8 French guide catheter inserted through the femoral vein, all plugs were placed at the narrowest segment of the fistula, leading to immediate complete closure of 2 fistulas. The third patient, with a fistula of the circumflex coronary artery, who received the largest plug initially had residual flow, but the fistula was found to be completely occluded at 12 months follow-up examination. We have demonstrated, therefore, safe and effective usage of the new vascular plug for transcatheter closure of moderate- to large-sized coronary arterial fistulas. The plug offers an alternative to cardiac surgery, or occlusion using coils.

Type
Original Article
Copyright
© 2007 Cambridge University Press

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Sunder KR, Balakrishnan KG, Tharakan JA, et al. Coronary artery fistula in children and adults: a review of 25 cases with long-term observations. Int J Cardiol 1997; 58: 4753.Google Scholar
Liberthson RR, Sagar K, Berkoben JP, Weintraub RM, Levine FH. Congenital coronary arteriovenous fistula. Report of 13 patients, review of the literature and delineation of management. Circulation 1979; 59: 849854.Google Scholar
Björck G, Crafoord C. Arteriovenous aneurysm on the pulmonary artery simulating patent ductus arteriosus botalli. Thorax 1947; 2: 6568.Google Scholar
Reidy JF, Sowton E, Ross DN. Transcatheter occlusion of coronary to bronchial anastomosis by detachable balloon combined with coronary angioplasty at same procedure. Br Heart J 1983; 49: 284287.Google Scholar
Reidy JF, Anjos RT, Qureshi SA, Baker EJ, Tynan MJ. Transcatheter embolization in the treatment of coronary artery fistulas. J Am Coll Cardiol 1991; 18: 187192.Google Scholar
Qureshi SA, Tynan M. Catheter closure of coronary artery fistulas. J Interv Cardiol 2001; 14: 299307.Google Scholar
Armsby LR, Keane JF, Sherwood MC, Forbess JM, Perry SB, Lock JE. Management of coronary artery fistulae. Patient selection and results of transcatheter closure. J Am Coll Cardiol 2002; 39: 10261032.Google Scholar
Hill SL, Hijazi ZM, Hellenbrand WE, Cheatham JP. Evaluation of the AMPLATZER vascular plug for embolization of peripheral vascular malformations associated with congenital heart disease. Catheter Cardiovasc Interv 2006; 67: 113119.Google Scholar
Kessler J, Trerotola SO. Use of the Amplatzer Vascular Plug for embolization of a large retroperitoneal shunt during transjugular intrahepatic portosystemic shunt creation for gastric variceal bleeding. J Vasc Interv Radiol 2006; 17: 135140.Google Scholar
Farra H, Balzer DT. Transcatheter occlusion of a large pulmonary arteriovenous malformation using the Amplatzer vascular plug. Pediatr Cardiol 2005; 26: 683685.Google Scholar
Gillespie MJ, Golden A, Sivarajan VB, Rome JJ. Transcatheter closure of patent ductus venosus with the Amplatzer vascular plug in twin brothers. Pediatr Cardiol 2006; 27: 142145.Google Scholar
Subramanyam K, Kapoor R, Radhakrishnan S, et al. Use of amplatzer vascular plugs in the closure of coronary arteriovenous fistula. Indian Heart Journal 2005; 57: Abstract No. 231: 288.Google Scholar
Boening A, Scheewe J, Fischer G, Cremer J. Surgical therapy of a coronary artery fistula draining into the left atrium. Pediatr Cardiol 2001; 22: 412414.Google Scholar
Piechaud JF. Closing down: transcatheter closure of intracardiac defects and vessel embolisations. Heart 2004; 90: 15051510.Google Scholar