Hostname: page-component-586b7cd67f-dsjbd Total loading time: 0 Render date: 2024-11-23T01:14:27.309Z Has data issue: false hasContentIssue false

Early dehiscence of a tricuspid valve annuloplasty ring in an adolescent with hypoplastic left heart syndrome presenting with unconjugated hyperbilirubinemia

Published online by Cambridge University Press:  22 November 2024

David Kielmayer*
Affiliation:
Department of Pediatric Cardiology, Kepler University Hospital, Linz, Austria
Eva Sames-Dolzer
Affiliation:
Division of Pediatric and Congenital Heart Surgery, Kepler University Hospital, Linz, Austria Medical Faculty, Johannes Kepler University Linz, Linz, Austria
Gerald Tulzer
Affiliation:
Department of Pediatric Cardiology, Kepler University Hospital, Linz, Austria Medical Faculty, Johannes Kepler University Linz, Linz, Austria
*
Corresponding author: David Kielmayer; Email: [email protected]
Rights & Permissions [Opens in a new window]

Abstract

We report a unique case of an adolescent patient with Fontan physiology presenting with unconjugated hyperbilirubinemia due to dehiscence of a tricuspid valve annuloplasty ring.

Type
Brief Report
Copyright
© The Author(s), 2024. Published by Cambridge University Press

Case report

A 14-year-old adolescent of Kyrgyz origin with hypoplastic left heart syndrome and Fontan circulation and an implanted dual chamber (DDD) pacemaker—sought medical attention in the emergency department due to jaundice for 10 days. He was discharged under best conditions from hospital a month earlier after tricuspidal valve repair via DeVega annuloplasty with a 30 mm Edwards® MC3 Tricuspid annuloplasty ring. On admission, discrete dermal and scleral icterus, a 3/6 systolic heart murmur, and an arrhythmical heart rhythm were found. There was no history of discoloured stool or reddish/brown urine and pruritus was denied. Furthermore, there was no history of fever, oedema, decreased physical capacity, increasing cyanosis, or respiratory distress. Blood pressure was 94/54 mmHg (right upper extremity), heart rate was 84 beats per minute, and saturation values averaged at 87%.

The patient was admitted to the Department of Pediatric Gastroenterology and Hepatology for further investigation where laboratory evaluation showed haemolysis without anaemia (haemoglobin: 13.1 g/dL) with a high reticulocyte count, reflecting an increased erythropoietic response, as well as an unconjugated hyperbilirubinemia (total serum bilirubin: 6.05 mg/dL, unconjugated bilirubin: 5.35 mg/dL) and a lactate dehydrogenase exceeding the laboratory measurable values.

Since extensive evaluation could not find an internal cause, the patient was re-introduced for a paediatric cardiologist consultation. The electrocardiogram showed an atrial sensed, ventricular paced rhythm with occasional ventricular extra beats. Echocardiography revealed moderate tricuspid valve regurgitation due to annuloplasty ring dehiscence from the septal aspect of the valve (Figure 1) as etiologic reason for mechanical haemolysis. This led to surgical indication of re-operation (Figure 2) with multidisciplinary decision for a mechanical valve replacement with a 30 mm SJM-MECJ-502 Mitral 33®. Postoperatively, he was admitted to the ICU, extubated early with only minimal need for catecholamines, and transferred to the paediatric cardiac intermediate care unit two days after the intervention. Haemolysis parameters, including total and unconjugated bilirubin (Figure 3), normalised, and jaundice disappeared within four days after valve replacement. On day five, he was discharged with only mild medication for heart failure and phenprocoumon for permanent anticoagulation.

Figure 1. ( a ) Two-dimensional transthoracic echocardiography apical view indicating dehiscence of the tricuspid annuloplasty ring from the septal valve aspect during diastole. ( b ) Colour Doppler image during diastole with blood inflow separated by the septal part of the ring and ( c ) tricuspid regurgitation during systole. F = extracardiac Fontan connection, LA = left atrium, RA = right atrium, RV = right ventricle, TVAL = anterior leaflet of tricuspid valve, *septal leaflet of tricuspid valve, empty arrow heads: area of dehiscence, filled arrows: tricuspid annuloplasty ring.

Figure 2. Intraoperative surgeons view showing the loose part of the annuloplasty ring held by the forceps.

Figure 3. Dynamics of total serum (blue line) and unconjugated (green line) bilirubin (mg/dL) from readmission showing dramatic drop of these parameters immediately after implantation of the mechanical tricuspid valve on day 21 indicated by the red vertical line. Follow-up total serum bilirubin consistently showed normal values until day 70 after readmission; unconjugated bilirubin was too low to be determined.

Discussion

Tricuspid regurgitation is a common complication seen in patients with hypoplastic left heart syndrome, resulting from various congenital and acquired factors like a highly variable subvalvar apparatus, annulus diameter or number of leaflets as well as leaflet dysplasia with preference for the septal one and consecutive movement restriction and higher incidence of cleft formation. Reference Bharucha and Viola1Reference Takahashi, Inage and Rebeyka3 Throughout the palliation stages, the systemic hypertrophied ventricle evolves a spherical ventricular geometry, which leads to shape-changes of atrioventricular valvular closure as well as to impairment of systolic and diastolic function, which furthermore contributes to tricuspid regurgitation. Reference Bharucha, Honjo and Seller4,Reference Gewillig and Brown5 In the past years, atrioventricular valve dysfunction in Fontan patients has become focus of scientific research, indicating higher morbidity and mortality, for example, increased risk of Fontan failure and transplantation in those with moderate or severe regurgitation. Reference King, Buratto and Celermajer6 Optimal timing and technique for atrioventricular valve intervention still is controversially discussed; it should be mentioned that repair should be aimed for before the development of definitive right ventricular dysfunction, since in these patients, tricuspid valves tend to continue to deteriorate. Reference Ohye, Gomez, Goldberg, Graves, Devaney and Bove7

Valve-sparing techniques are preferred; the most common ones are (partial) annuloplasty, commisuroplasty, and edge-to-edge repair. Nonetheless, careful preoperative imaging should always precede and contribute to surgical decision-making. Reference Tseng, Siddiqui and Di Maria8 Data on tricuspid annuloplasty ring dehiscence in the paediatric cohort are sparse; there are, however, numerous reports on this complication after mitral valve repair in adults.

To the authors’ knowledge, there are no reported cases of early tricuspid annuloplasty ring dehiscence presenting with mechanical haemolysis in adolescents with hypoplastic left heart syndrome and Fontan physiology, whereas in adults, haemolysis and hyperbilirubinemia due to failed annuloplasty rings have been reported sporadically. Reference Bhardwaj, Golwala, Song, Lantz, Chadderdon and Zahr9

Our patient recovered fast and was clinically asymptomatic after definitive atrioventricular valve replacement regarding jaundice and bilirubin values.

Despite the lack of guidelines on management and follow-up after atrioventricular valve repair in the reviewed literature, the authors stress that reoccurring postoperative valve regurgitation, with or without rise in haemolysis parameters, should always arouse clinical suspicion of valve repair failure. Three-dimensional transthoracic or conventional transesophageal echocardiography is a promising diagnostic imaging tools with additional benefit in these patients. The annuloplasty ring used in our patient was of rigid nature; experience in adults shows an increased risk of dehiscence with this kind of ring. Reference Pfannmüller, Doenst, Eberhardt, Seeburger, Borger and Mohr10 Applicability of these data on children with CHD, however, is problematic, as studies are lacking.

Acknowledgements

None.

Financial support

None.

Competing interests

None.

References

Bharucha, T, Viola, N. The tricuspid valve in hypoplastic left heart syndrome: echocardiography provides insight into anatomy and function. Front Pediatr 2013; 11: 1145161.CrossRefGoogle Scholar
Stamm, C. The morphologically tricuspid valve in hypoplastic left heart syndrome. Eur J Cardiothorac Surg 1997; 12 ( 4 ): 587592.CrossRefGoogle ScholarPubMed
Takahashi, K, Inage, A, Rebeyka, IM, et al. Real-time 3-dimensional echocardiography provides new insight into mechanisms of tricuspid valve regurgitation in patients with hypoplastic left heart syndrome. Circulation 2009; 120 ( 12 ): 10911098.CrossRefGoogle ScholarPubMed
Bharucha, T, Honjo, O, Seller, N, et al. Mechanisms of tricuspid valve regurgitation in hypoplastic left heart syndrome: a case-matched echocardiographic-surgical comparison study. Eur Heart J Cardiovasc Imaging 2013; 14 ( 2 ): 135141.CrossRefGoogle ScholarPubMed
Gewillig, M, Brown, SC. The fontan circulation after 45 years: update in physiology. Heart 2016; 102 ( 14 ): 10811086.CrossRefGoogle ScholarPubMed
King, G, Buratto, E, Celermajer, DS, et al. Natural and modified history of atrioventricular valve regurgitation in patients with fontan circulation. J Am Coll Cardiol 2022; 79 ( 18 ): 18321845.CrossRefGoogle ScholarPubMed
Ohye, R, Gomez, C, Goldberg, C, Graves, H, Devaney, E, Bove, E. Repair of the tricuspid valve in hypoplastic left heart syndrome. Cardiol Young 2006; 16 ( S3 ): 2126. DOI: 10.1017/S1047951106000722.CrossRefGoogle ScholarPubMed
Tseng, SY, Siddiqui, S, Di Maria, MV, et al. Atrioventricular valve regurgitation in single ventricle heart disease: a common problem associated with progressive deterioration and mortality. J Am Heart Assoc 2020; 9 ( 11 ): e015737.CrossRefGoogle ScholarPubMed
Bhardwaj, B, Golwala, H, Song, HK, Lantz, G, Chadderdon, S, Zahr, F. Haemolysis resolution after transcatheter valve in valve within a prior mitral annuloplasty ring: a case report. Eur Heart J Case Rep 2022; 6 ( 8 ): ytac 326.CrossRefGoogle ScholarPubMed
Pfannmüller, B, Doenst, T, Eberhardt, K, Seeburger, J, Borger, MA, Mohr, FW. Increased risk of dehiscence after tricuspid valve repair with rigid annuloplasty rings. J Thorac Cardiovasc Surg 2012; 143 ( 5 ): 10501055.CrossRefGoogle ScholarPubMed
Figure 0

Figure 1. (a) Two-dimensional transthoracic echocardiography apical view indicating dehiscence of the tricuspid annuloplasty ring from the septal valve aspect during diastole. (b) Colour Doppler image during diastole with blood inflow separated by the septal part of the ring and (c) tricuspid regurgitation during systole. F = extracardiac Fontan connection, LA = left atrium, RA = right atrium, RV = right ventricle, TVAL = anterior leaflet of tricuspid valve, *septal leaflet of tricuspid valve, empty arrow heads: area of dehiscence, filled arrows: tricuspid annuloplasty ring.

Figure 1

Figure 2. Intraoperative surgeons view showing the loose part of the annuloplasty ring held by the forceps.

Figure 2

Figure 3. Dynamics of total serum (blue line) and unconjugated (green line) bilirubin (mg/dL) from readmission showing dramatic drop of these parameters immediately after implantation of the mechanical tricuspid valve on day 21 indicated by the red vertical line. Follow-up total serum bilirubin consistently showed normal values until day 70 after readmission; unconjugated bilirubin was too low to be determined.