Hostname: page-component-669899f699-tpknm Total loading time: 0 Render date: 2025-04-24T12:06:55.870Z Has data issue: false hasContentIssue false

Analysis of ST segment alarms in children admitted to the paediatric and cardiac intensive care units and cardiac progressive care unit: a single-centre retrospective study

Published online by Cambridge University Press:  25 September 2024

Christine Peyton Gallagher*
Affiliation:
Pediatrics, University of Colorado Anschutz Medical Campus, Aurora, CO, USA Cardiac ICU, Boston Children’s Hospital, Boston, MA, USA
Sonja I. Ziniel
Affiliation:
Department of Pediatrics, Children’s Hospital Colorado, Aurora, CO, USA
Jennifer Newman
Affiliation:
Children’s Hospital Colorado, Aurora, CO, USA
Beth Wathen
Affiliation:
Children’s Hospital Colorado, Aurora, CO, USA
Paul Marusich
Affiliation:
Children’s Hospital Colorado, Aurora, CO, USA
Christopher Poppy
Affiliation:
Department of Pediatrics, Children’s Hospital Colorado, Aurora, CO, USA
Todd Carpenter
Affiliation:
Pediatrics, University of Colorado Anschutz Medical Campus, Aurora, CO, USA
Katja M. Gist
Affiliation:
Department of Pediatrics, Cincinnati Children’s Hospital Medical Center, University of Cincinnati College of Medicine, Cincinnati, OH, USA
*
Corresponding author: C. P. Gallagher; Email: [email protected]

Abstract

ST segment monitoring in the adult population allows for the early detection of myocardial ischaemia. In children admitted to the paediatric intensive care unit (PICU), cardiac intensive care unit (CICU), and cardiac progressive care unit (CPCU), it is unclear if continuous ST segment alarm monitoring is necessary in all patients. All patients admitted to the PICU, CICU, and CPCU during the study period were included. Children with any ST segment alarms were compared with those without an alarm during their stay. The electrocardiogram confirmed true ST segment alarms were compared with all other ST segment alarms. Demographic and clinical data were extracted from the medical record. Medical interventions and procedures occurring around ST segment alarms were recorded for multivariable analysis assessing for the association of true ST segment. Logistic regression was used to evaluate the associations with ST segment alarms during hospital stays. ST segment alarms occurred in 36% of hospital stays, and only 3.4% were considered true. True alarms were significantly more common among patients with a cardiac-related diagnosis, located in both cardiac units, and having received an intervention with any vasoactive medication. In the multivariable logistic regression, patients 11 years or older, hypotension, supraventricular tachycardia, and initiation/escalation of any vasoactive were independently associated with a true ST segment alarm. True ST segment alarms were infrequent, occurring in 1.2% of stays during the study period. Alarm monitoring may be beneficial in those with an underlying cardiac diagnosis.

Type
Original Article
Copyright
© The Author(s), 2024. Published by Cambridge University Press

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

Article purchase

Temporarily unavailable

References

Flanders, SA. Continuous ST-segment monitoring: raising the bar. Crit Care Nurs Clin North Am 2006; 18: 169177.CrossRefGoogle ScholarPubMed
Fox, L, Kirkendall, C, Craney, M. Continuous ST-segment monitoring in the intensive care unit. Crit Care Nurse 2010; 30: 3343.CrossRefGoogle ScholarPubMed
Pelter, MM, Kozik, TM, Al-Zaiti, SS, Carey, MG. Acute coronary syndrome ST-segment monitoring. Am J Crit Care 2014; 23: 503504.CrossRefGoogle ScholarPubMed
Sandau, KE, Funk, M, Auerbach, A, et al. Update to practice standards for electrocardiographic monitoring in hospital settings: a scientific statement from the American heart association. Circulation 2017; 136: e273e344.CrossRefGoogle ScholarPubMed
Wimmer, NJ, Scirica, BM, Stone, PH. The clinical significance of continuous ECG (ambulatory ECG or holter) monitoring of the ST-segment to evaluate ischemia: a review. Prog Cardiovasc Dis 2013; 56: 195202.CrossRefGoogle ScholarPubMed
A AACN Practice Alert: ST-Segment Monitoring. American Association of Critical Care Nurses, Aliso Viejo, CA, 2016, pp 9265694109. https://www.aacn.org/clinical-resources/practice-alerts/st-segment-monitoring.Google Scholar
Christensen, M, Dodds, A, Sauer, J, Watts, N. Alarm setting for the critically ill patient: a descriptive pilot survey of nurses’ perceptions of current practice in an Australian regional critical care unit. Intensive Crit Care Nurs 2014; 30: 204210.CrossRefGoogle Scholar
Drew, BJ, Harris, P, Zegre-Hemsey, JK, Mammone, T, et al. Insights into the problem of alarm fatigue with physiologic monitor devices: a comprehensive observational study of consecutive intensive care unit patients. PLoS One 2014; 9: e110274.CrossRefGoogle ScholarPubMed
Karnik, A, Bonafide, CP. A framework for reducing alarm fatigue on pediatric inpatient units. Hosp Pediatr 2015; 5: 160163.CrossRefGoogle ScholarPubMed
Joint Commission on Accreditation of Healthcare Organizations. The joint commission announces 2014 National patient safety goal. Jt Comm Perspect 2013; 33: 34.Google Scholar
Lewandowska, K, Weisbrot, M, Cieloszyk, A, Medrzycka-Dabrowska, W, Krupa, S, Ozga, D. Impact of alarm fatigue on the work of nurses in an intensive care environment-A systematic review. Int J Environ Res Public Health 2020; 17: 8409.CrossRefGoogle Scholar
Alten, J, Cooper, DS, Klugman, D, et al. Preventing cardiac arrest in the pediatric cardiac intensive care unit through multicenter collaboration. JAMA Pediatr 2022; 176: 10271036.CrossRefGoogle ScholarPubMed