CLINICIAN’S CAPSULE
What is known about the topic?
Focused assessment with sonography (FAST) could be used by prehospital systems but the optimal training regimen has not been established.
What did this study ask?
Can paramedics interpret FAST in trauma as well as emergency physicians after a 1 hour didactic training session.
What did this study find?
FAST interpretation was comparable between the two groups with accuracy of 85.6% and 87.5% for paramedics and emergency physicians respectively.
Why does this study matter to clinicians?
Determination of effective, cost effective training programs is important when considering incorporating FAST examinations in prehospital systems.
BACKGROUND
The Focused Assessment with Sonography in Trauma (FAST) exam is a rapid test that uses ultrasound to survey the pericardial and abdominal regions sequentially for sonographic signs of hemorrhage.Reference Qi, Tian and Sun1,Reference Walcher, Weinlich and Conrad2 The FAST exam is particularly useful to medical practitioners as a quick clinical adjunct for determining the most appropriate procedure (laparotomy or thoracotomy) to obtain definitive hemorrhage control in the unstable trauma patient.Reference Walcher, Weinlich and Conrad2,Reference Wherrett, Boulanger and McLellan3 In the prehospital setting, the information from a FAST examination can help triage patients, direct patients to the most appropriate facilities, assist with management strategies, and potentially expedite time to definitive intervention.3,4
Despite significant literature to support the use of FAST in both hospital and prehospital settings by physicians, the utility of paramedic-performed FAST exams in the Canadian prehospital system is unclear.Reference O'Dochartaigh, Douma, Alexiu, Ryan and MacKenzie5 Current literature supports the use of prehospital FAST exams in prehospital systems where the prehospital team consists of physicians and paramedics. In these studies, FAST exams were performed or supervised by physicians.Reference von Vopelius-Feldt and Benger6 There are few studies that examine the accuracy of paramedic-only-performed FAST examinations.Reference Yates and Baylous7,Reference Press, Miller and Hassan8 In these studies, the accuracy of paramedic-performed FAST examinations and the optimal duration of a suitable paramedic ultrasound training program remain controversial.Reference Walcher, Kirschning and Müller9,Reference Heegaard, Plummer and Dries10
It can be assumed that paramedics with significantly less training may require more robust training regimens than physicians to achieve the same level of competency in ultrasound assessments such as the FAST exam. However, despite the potential benefits to the Canadian prehospital system, a potential barrier to implementation is the tremendous financial and operational burden, if paramedics require prolonged ultrasound training courses. In this study, we conducted a double-blinded observational study comparing the accuracy of paramedic-performed FAST with that of physician-performed tests on a sonographic Phantom, after a one-hour didactic training session.
METHODS
Study design
A prospective observational study was conducted in which Canadian aeromedical critical care paramedics were compared with emergency medicine staff with regards to their ability to accurately perform and interpret FAST examinations on a FAST phantom (CAE FAST Exam Real Time Ultrasound Training Model – model #BP-FAST1800), after a one-hour didactic training course. The trial used one ultrasound system (Philips lumify S4-1 broadband phased array 4-1 MHz transducer) for all scans performed in this study.
Study setting
The training sessions and evaluation of accuracy for performing FAST examinations took place at the Middlesex-London Paramedic Services Headquarters in London, Ontario, Canada.
Population
A letter asking for volunteers from the critical care paramedic contingent at two Ornge air ambulance bases (London and Toronto) was circulated through Ornge management. Only travel expenses were covered. Emergency medicine physicians were recruited from a major academic health sciences centre (London Health Sciences Centre). One day of training and evaluation was selected on May 21, 2019.
Intervention
A standardized one-hour didactic training course was taught to all paramedics and emergency physicians. The training course consisted of an introduction to the ultrasound system, including how to turn on the system and how to position the probe and change gain and depth setting on the ultrasound device. The training then focused on identifying anatomic and ultrasonographic landmarks necessary for performing an appropriate abdominal FAST exam. The three areas of focus were the hepatorenal space, splenorenal space, and retrovesicular/retrouterine pouch in the pelvis. Participants were taught to identify anechoic stripes in the appropriate anatomic locations as a sign of potential hemorrhage. Pericardial examination was not taught or performed.
Outcome measures
After the one-hour didactic session, participants were asked to complete a survey with questions pertaining to baseline demographic information, prior ultrasound training, and years of experience in their field. Each participant was then asked to perform two distinct scans on a specialized mannequin designed for FAST ultrasound training techniques. The mannequin utilized in this study was a realistic model of a human torso in which fluid could be injected into the abdomen to create a realistic ultrasound image of abdominal free fluid. Participants were required to scan the mannequin twice, once with 300 mL of fluid instilled and once with the abdomen free of fluid. Participants were blinded to the status of hemoperitoneum. The primary outcome of the study was the accuracy rate of FAST examination by paramedics compared with emergency room physicians. The accuracy of the FAST exam was reported as:
(number of true positives and true negatives)/(total number of scans performed).
Data analysis
Demographics, previous ultrasound experience, and FAST interpretation were analysed in a descriptive fashion using proportions with percentages. Time to complete each scan was summed across all scans for each participant to create a total scan time. Total scan time was reported using means, standard deviations, and 95% confidence intervals (CI) and was compared between groups using a standard t-test.
We compared the accuracy of paramedic-performed FAST exams with the accuracy of physician-performed FAST examinations, using a chi-square test. Differences in accuracy rate were deemed significant if p <0.05.
RESULTS
Fourteen critical care paramedics and four emergency department (ED) physicians (two staff physicians and two residents) were voluntarily recruited. The critical care paramedics were mostly ultrasound-naive whereas the emergency physicians all had ultrasound training.
The correct interpretation of FAST scans was comparable between the paramedic and emergency physician groups, with an accuracy of 85.6% and 87.5%, respectively (∆1.79, 95% CI -33.85 to 21.82, p = 0.90). Average time per scan differed between groups but did not reach statistical significance; paramedics took longer to complete the FAST examination with a mean (standard deviation [SD]) time per scan of 5.18 (1.72) minutes, as compared with 3.67 (1.37) minutes for physicians (∆1.51 minutes, 95% CI -0.48 to 3.50, p = 0.128).
DISCUSSION
In this study, we found that ultrasound-naive critical care flight paramedics could accurately determine the presence or absence of simulated hemoperitoneum on abdominal FAST examination of an ultrasound phantom, after a one-hour training session. They accurately identified the presence or absence of free fluid on the FAST examination 85.6% of the time, compared with a control group of ED physicians, who were accurate 87.5% of the time.
The original papers describing the FAST examination focused on identifying free intraperitoneal or pericardial fluid in blunt trauma patients.Reference Osterwalder, Mathis and Hoffmann11 Originally, the clinical utility of the FAST examination was to identify trauma patients quickly who would benefit from emergent surgical management of either large volume hemoperitoneum or cardiac tamponade.Reference Osterwalder, Mathis and Hoffmann11 Ultrasound has now been expanded to include an examination of a variety of other clinically significant injuries in trauma that include: hemoperitoneum; pneumothorax; hemothorax; hemopericardium with or without tamponade; traumatic hypovolemia; and even rib, nose, and other fractures.Reference Osterwalder, Mathis and Hoffmann11,Reference Navaratnam and Davis12 Despite the increased utility of the extended FAST examination in trauma for a variety of conditions, there is disagreement regarding appropriate management strategies given particular positive findings. In this study, we chose to use 300 cc of fluid to simulate a clinically significant volume of hemoperitoneum. As such, our study suggests that a brief training program might be able to teach paramedics to interpret abdominal FAST examinations reliably only for large volume hemoperitoneum that would alter facility choice and/or triage priority.
Our study adds to the minimal literature concerning the optimal training duration and accuracy of FAST exams performed by paramedics.Reference Yates and Baylous7,Reference Press, Miller and Hassan8 Previous studies suggest that with a combination of didactic and practical education modalities, FAST can be effectively taught to paramedics in as little as a one-day course to reach a high degree of diagnostic accuracy. Unluer et al. trained four paramedics on FAST examinations for eight hours and reported a sensitivity of 84.6% for hemoperitoneum and specificity of 97.4%.Reference Unlüer, Yavaşi and Kara13 Walcher et al. trained a mixed group of five physicians and four paramedics on prehospital FAST exam, using healthy volunteers and patients with peritoneal dialysis for a full day. These nine study participants performed a total of 39 FAST examinations in the field, with both sensitivity and specificity of 100%.Reference Walcher, Kirschning and Müller9 Press et al. trained 33 paramedics and nurses to perform extended FAST exams (FAST and pleural assessments) on consenting in-flight trauma patients. Training consisted of a three-hour didactic and eight-hour practical over a two-month period.Reference Press, Miller and Hassan14 Our study is the first to suggest that FAST can be taught using a shorter duration course if the goal is only to detect a significant amount of free intra-abdominal blood.
In the past, shorter duration courses were associated with lower accuracy. Kim et al. trained six level 1 emergency medical technicians (EMTs) on FAST in a two-hour didactic and two-hour practical training program performed on EMT volunteers.Reference Kim, Shin, Song and Park15 Sensitivity was 61.3%, and specificity was 96.3% for the detection of peritoneal cavity fluid. West studied 10 paramedics and their accuracy in diagnosing free fluid in the abdomen in a cohort of patients requiring peritoneal dialysis. The paramedic success rate was only 60% for diagnosing free fluid, after a training duration of four hours. However, as discussed, the amount of fluid in the abdomen was significantly associated with increased sensitivity of paramedic-performed FAST.Reference West, Cusser, Etengoff, Landsgaard and LaBond16 Kim and colleagues found that paramedic FAST performed in the hospital showed low sensitivity for any hemoperitoneum, as compared with computed tomography (CT) scan findings. However, sensitivity increased to over 86%, as compared with CT findings of significant or moderate amounts of hemoperitoneum.Reference Kim, Shin, Song and Park15
LIMITATIONS
The major limitation of our study was the use of a mannequin and not human volunteers. Mannequin images simulate anatomic and ultrasonographic landmarks; however, they do not simulate anatomic variants, and images generated are evidently different, as compared with live volunteers. Another limitation of our study is that the overall number of paramedics and controls was relatively small, limiting the ability to apply these data to larger groups.
CONCLUSION
This study determined that critical care paramedics were able to detect large volumes of free fluid on a simulated abdomen after a short one-hour training course. Overall, this study is hypothesis generating. This suggests the potential use in prehospital programs to determine the most appropriate transport destination and aid in the triage of trauma patients while limiting the financial and logistical burden of ultrasound training. Further studies need to be performed to determine the optimal training time and effective training regimens for both small volume hemoperitoneum and extended FAST exams.
Competing interests
None. This study was approved by the Western Health Science Research Ethics Board.