No CrossRef data available.
Published online by Cambridge University Press: 06 March 2019
The effect of resin particles on copper x-ray fluorescence was studied. For any given resin size the relationship between copper concentration and x-ray intensity was linear. As the particle size decreased, the x-ray intensity increased for any given copper concentration. The general shape of the curves are similar to the ones predicted by Bernstein for a minor constituent in a power sample. This study indicates that the variation in intensity with particle size can be eliminated by using resins with a mean particle diameter of 56 microns or less.