An Energy Dispersive System for the Analysis of Trace Elements in Human Blood Serum
Published online by Cambridge University Press: 06 March 2019
Abstract
A computerized systetn, consisting of a fluorescence x-ray source (Philips), an Si (Li) detector (Ortec), and a mini Computer (Xerox) is currentiy in operation at The University of Texas M. D. Anderson Hospital and Tumor Institute, Experimentel Pathology Section, for the analysis of trace elements in biological specimens. The elements of interest are Fe, Cu, and Zn for possible significance in the detection and study of cancer. The detection system is being used for comparatively routine analysis of these elements, which are present in blood serum in the ppM range. With appropriate sample preparation techniques, using 1 ml of serum, the limit of detectability for these elements is estimated to be 100 ppB (1 in 107). Further refinement is possible, and research in this direction continues.