Initially, Wielopolski (Wielopolski et al. 1981) used x-rays from either a 125I or a 105Cd source to estimate lead (Pb) in tibial cortical bone in intact legs, post-mortem. This system utilized the Lα and Lβ x-rays of Pb with energies of 10.5 and 12.6 keV, respectively. The minimum detection limit (MDL) was considered to be comparable with existing K-line x-ray fluorescence (KXRF) instruments, namely, 20-30 ppm. The feasibility of partially polarized radiation (Barkla, 1906) was assessed from a 125I source on Pb (No3)2 (10,000 ppm) dissolved in wa:er. Reduction of the detector total count rate was observed by a factor of two; and reduction in background by the same factor was appaent. This preliminary study suggested that, with improved design, tlie MDL might be lowered by a factor of five (Wielopolski et al., 1981).