Published online by Cambridge University Press: 06 March 2019
A prototype instrument, which provides x-ray powder diffraction and x-ray fluorescence analysis in a compact unit, has been developed to support the needs of NASA for planetary exploration. The instrument uses a 9-watt Fe-anodc x-ray tube and CCD in a fixed geometry for recording powder patterns with a 2θ range of 35°. The fluorescence spectrum for elements below Fe is collected simultaneously with the diffraction data. A shuttered Cd-109 isotopic source with emissions at 22 and 80 keV is used to excite higher energy fluorescence.
The low-energy limit for discriminating single photon events was found to be ∼1.5 keV. Al-K could be distinguished from a pure sample, but the spectrum below 6 keV was degraded by the read noise of the CCD, which introduced spectral artifacts. Diffraction peaks from halite had a FWHM of ∼1°(2θ), with major contributions to the width from the use of slit collimation on the source and the low tilt angle of the sample.