We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure [email protected]
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
CyberKnife is the most advanced form of stereotactic body radiotherapy (SBRT) system that uses a robotic arm to deliver highly focused beams of radiation; however, a limitation is that it only irradiates from ceiling to floor direction. In patients with posterior lungs tumour who are positioned supine, normal lung tissue may suffer undesirable radiation injuries. This study compares the treatment planning between the prone set-up and the supine set-up for lung cancer in CyberKnife SBRT to decrease normal lung dose to avoid radiation side effects.
Materials and methods:
A human phantom was used to generate 108 plans (54 for prone and 54 for supine) using the CyberKnife planning platform. The supine and prone plans were compared in terms of the dosimetric characteristics, delivery efficiency and plan efficiency.
Results:
For posterior targets, the area of low-dose exposure to normal lungs was smaller in the prone set-up than in the supine set-up. V10 of the lungs was 7·53% and 10·47% (p < 0·001) in the anterior region, and 10·78% and 8·03% (p < 0·001) in the posterior region in the supine and prone set-up plans, respectively.
Conclusions:
The comparison between the prone set-up and the supine set-up was investigated with regard to target coverage and dose to organs at risk. Our results may be deployed in CyberKnife treatment planning to monitor normal tissue dose by considering patient positioning. This may assist in the design of better treatment plans and prevention of symptomatic radiation pneumonitis in lung cancer patients.