We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure [email protected]
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
Varian RapidArc is a volumetric modulated arc therapy (VMAT) that obtains a conformal dose around the desired structure by employing variable gantry speed, dose rate and dynamic multileaf collimator (DMLC) speed as the gantry rotates about machine isocenter. This study is meant to build upon previous research by Ling et al. by completing the tests with an in vivo dosimetric device attached to the linac gantry and a 2D ionisation chamber array with an isocentric gantry mount.
Materials and methods
Two PTW detectors, seven29 array with gantry mount and DAVID, were attached to the linear accelerator gantry, allowing each device to remain perpendicular to the beam at all gantry angles. Three tests for RapidArc evaluation were performed on these devices including: dose rate and gantry speed variation, DMLC speed and dose rate variation and DMLC position accuracy. The reproducibility of the arc data was also reported.
Results
A picket fence plan varying dose rates (111 to 600 MU/minute) and gantry speeds (5·5 to 4·3°/second) was delivered consisting of seven sections of different combinations. These measurements were compared with static gantry, open field measurements and found to be within 2·39% for the DAVID device and 0·84% for the seven29. A four-section picket fence of varying DMLC speeds (0·46, 0·92, 1·84 and 2·76 cm/second) was similarly evaluated and found to be within 1·99% and 3·66% for the DAVID and seven29, respectively. For DMLC position accuracy, a picket fence arc plan was compared with a static picket fence and found to agree within 0.38% and 2.91%. Reproducibility for these three RapidArc plans was found to be within 0·30% and 2·70% for the DAVID and seven29.
Conclusion
The DAVID and seven29 detectors were able to perform the RapidArc quality assurance tests efficiently and accurately and the results were reproducible. Periodic verification of DMLC movement, dose rate variation and gantry speed variation relating to RapidArc delivery can be completed in a timelier manner using this equipment.