We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure [email protected]
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
The control points (CP) play a significant role in the delivery of segmented based Intensity-Modulated Radiation Therapy (IMRT) delivery, particularly in dynamic mode. The number of segments is determined by control points and these segments will transfer from one to the other either during beam ON called dynamic delivery or during beam OFF called static delivery or step and shoot. This study was aimed at indirect estimation of the total monitor units (MU) to be delivered per field by exploiting the control points and also to find the MUs at any nth segment.
Materials and methods:
This study was performed in the Eclipse treatment planning software version 13.8.0. The details of control points, metre set weight per segment, leaf positions for each segment, field size, etc. were taken into consideration.
Results:
TPS calculated MU value and analytically estimated MU value were compared and the percentage of difference was estimated. The overall mean percentage of deviation was 1·03% between the TPS calculated method and the analytical method. The paired sample t-test was performed and, p-value <0·05, no significant difference was found. The analytical relationship determined to estimate the total number of MU delivered for any nth control point was also evaluated.
Conclusion:
The control points are a potential parameter in the conventional IMRT delivery. Through this study, we have addressed the indirect method to estimate the monitor units delivered per segment.