We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure [email protected]
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
The RapidPlan is a radiotherapy planning tool that uses a dataset of approved plans to predict the dose distribution and automatically generates the dose–volume constraints for optimisation of the new plan. This study compares three strategies of model building for the treatment of prostate cancer with the 10-MV photon beam.
Methods:
Three models for prostate treatment were compared: Model 6X, Model10X and Model6Xrefined. Model6X is already used in our department and was trained on treatment plans based on the 6-MV photon beam. Model10X was trained on treatment plans based on the 10-MV photon beam and manually optimised by an experienced medical physicist. Finally, Model6Xrefined was trained on plans automatically created by the Model6X, but using the 10-MV photon beam. The three models were used to generate 25 new plans with the 10-MV photon beam.
Results:
Model10X generated plans with 2 Gy lower mean dose to bladder-PTV and rectum-PTV volumes and 8% lower V15Gy at bladder and rectum volumes, although the number of monitor units increased by 170 on average.
Conclusions:
The model trained on manually optimised plans generated plans with higher normal tissue sparing. However, model building is a time-consuming process, so a cost–benefit balance should be performed.