We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure [email protected]
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
One of the stereotactic radiosurgery techniques is Gamma Knife radiosurgery, in which intracranial lesions that are inaccessible or inappropriate for surgery are treated using 201 cobalt-60 sources in one treatment session. In this conformal technique, the penumbra width, which results in out-of-field dose in tumour-adjacent normal tissues should be determined accurately. The aim of this study is to calculate the penumbra widths of single and 201 beams for different collimator sizes of Gamma Knife machine model 4C using EGSnrc/BEAMnrc Monte Carlo simulation code and comparison the results with EBT3 film dosimetry data.
Methods and materials
In this study, simulation of Gamma Knife machine model 4C was performed based on the Monte Carlo codes of EGSnrc/BEAMnrc. To investigate the physical penumbra width (80−20%), the single beam and 201 beams profiles were obtained using EGSnrc/DOSXYZnrc code and EBT3 films located at isocentre point in a spherical Plexiglas head phantom.
Results
Based on the results, the single beam penumbra widths obtained from simulation data for 4, 8, 14 and 18 mm collimator sizes along X axis were 0·75, 0·77, 0·90 and 0·92 mm, respectively. The data for 201 beams obtained from simulation were 2·61, 4·80, 7·92 and 9·81 mm along X axis and 1·31, 1·60, 1·91 and 2·14 mm along Z axis and from film dosimetry were 3·21, 4·90, 8·00 and 10·61 mm along X axis and 1·22, 1·69, 2·01 and 2·25 mm along Z axis, respectively.
Conclusion
The differences between measured and simulated penumbra widths are in an acceptable range. However, for more precise measurement in the penumbra region in which dose gradient is high, Monte Carlo simulation is recommended.
Recommend this
Email your librarian or administrator to recommend adding this to your organisation's collection.