Understanding the contrast formation mechanism in the EM is one of the prerequisites for artefact-free reconstruction of biological structures from images. We found that the normally used correction of contrast formation applied to zero energy loss filtered images corrupted spatial resolution. Therefore the contribution of contrast formed by inelastic electrons was reconsidered, including partial coherence of inelastically scattered electrons and lens aberrations of the microscope. Based on this, a complete description of the zero-loss contrast transfer function (CTF) is now possible.
We used tobacco mosaic virus (TMV), a biological sample known at atomic resolution, for definition of optimum CTF-parameters to reconstruct defocus series from an EFTEM LEO 912. CTF theory as known so far describes image contrast in the weak phase approximation as a linear sum of amplitude and phase contrast. The contribution of amplitude contrast (ratio of amplitude to phase contrast A/P) was determined to be between 7% and 5 % for unfiltered images and 12-14 % for zero-loss filtered images. However, in a filter microscope we remove electrons from the image, so we expect a higher amplitude contrast than in non-filtered images.