Hostname: page-component-586b7cd67f-2brh9 Total loading time: 0 Render date: 2024-11-28T20:34:23.975Z Has data issue: false hasContentIssue false

Requirement of hollow process challenge device for monitoring hollow and complex instruments sterilization: a simulator for proper sterility assurance

Published online by Cambridge University Press:  10 June 2019

Debabrata Basu*
Affiliation:
Tata Medical Center, Kolkata, India
*
Author for correspondence: Debabrata Basu, MSc, DHM, Tata Medical Center, 14 Major Arterial Road (E-W), New Town, Rajarhat, Kolkata 700 160, India. Email: [email protected]
Rights & Permissions [Opens in a new window]

Abstract

Type
Letter to the Editor
Copyright
© 2019 by The Society for Healthcare Epidemiology of America. All rights reserved. 

To the Editor—The central sterile supply department (CSSD) acts as a backbone of any hospital services in which the sterilization of a wide range of materials is required for therapeutic, diagnostic, and infection control purposes. Depending upon the nature of materials, these items are sterilized using various sterilization processes such as steam sterilization, ethylene oxide sterilization, and hydrogen peroxide sterilization. The CSSD is committed to maintain a high level of standard in respect of quality assurance and quality control activities. To ensure the quality of equipment performance, several parameters are used to monitor the effectiveness of sterilization processes, including physical, chemical, and biological parameters. The overall vision is to ensure quality in all deliverables, not just sterile supplies.Reference Basu, Bhattacharya, Mahajan, Ramanan and Chandy1, Reference Basu, Bhattacharya, Mahajan, Ramanan and Chandy2

The CSSD supplies various sterile materials (eg, surgical instruments, reusable linens, and dressing materials) to clinical areas for patient care purposes. According to the Spaulding classification, most surgical instruments that touch blood and sterile tissues are categorically critical.Reference McDonnell and Burke3 Designwise, surgical instruments are of 2 types: solid and hollow. The sterilization of solid instruments is simple because the sterilant (eg, steam) can easily reach solid surfaces to destroy microorganisms. In contrast, the sterilization of hollow instruments (eg, suction tube or catheters) is difficult because air is entrapped easily inside the luminal instruments blocks steam from reaching all of the inner surfaces. Thus, for luminal instruments, satisfactory sterilization is not possible and the medical devices remain unsterile.

These problems can be overcome in 2 ways: inoculation test for culture or simulation test by hollow-process challenge device (HPCD). Both methods can be used to monitor the sterility of hollow and complex instruments.

The inoculation test is carried out by direct inoculation of complex hollow instruments at the most difficult penetration location, where usual biological indicators cannot reach. The inoculation test is the only way to check sterilization inside complex instruments and sealed areas. To check the efficacy of the sterilization procedure, the correct inoculation with a defined population is required at the worst-case location. For the steam process, suspension of G. stearothermophilus with a 40% ethanol concentration is required. This test cannot be used for routine monitoring, but it can be used for validation (EN ISO 17665-1, ISO 14937) or sterility testing purposes. For routine monitoring, HPCD testing (per EN867-5) is the better alternative because it can accurately simulate the luminal instruments by measuring their inner surfaces.4

The HPCD determines the hollow penetration resistance of a sterilizer, and a few HPCDs are required with different lengths and diameters, but wall thickness (eg, 0.5 mm) and tube materials (ie, polytetrafluoroethylene or stainless steel) are the same. The tube is open on one side for steam penetration and the other side is closed, with a capsule containing a biological indicator (per EN ISO 11138-1) or chemical indicator (per ISO 11140-1). The HPCDs are placed in the most challenging areas of the sterilizer (eg, in front of the door or above the drain strainer) in the worst-case scenario. After completion of the sterilization cycle, the results reveal that some of the chemical or biological indictor pass and others fail. The maximum size (inner diameter × inner length) that passes becomes the benchmark of the penetration capacity of the sterilizer.5

Worldwide, there are 2 types of steam sterilizers: the gravity type and the prevacuum type. The prevacuum sterilizer is more efficient than the gravity type with respect to the air removal process. Sterilization of solid instruments is easy (ie, only surface sterilization) because they can be sterilized using both gravity and prevacuum sterilization systems and because they can be monitored using any conventional method (ie, normal biological or chemical monitoring). Conversely, the sterilization of complex instruments is critical and can only be performed with a vacuum-assisted sterilization system due to the possibility of air pocket formation inside the luminal instruments. Under such conditions, no steam condenses and the sterilization process becomes suboptimal. Thus, for complex hollow instruments, HPCD is the only option for correct evaluation.6

Today, more and more infectious diseases pose serious threats to human health. A patient can be infected before admission to the hospital (ie, community-acquired infection or CAI) or may be infected after hospital admission (ie, hospital-acquired infection or HAI). Many HAIs are associated with prolonged hospital stays: ventilator-associated pneumonia (VAP), catheter-associated urinary tract infection (CAUTI), central-line–associated bloodstream infection (CLABSI), surgical-site infection (SSI), and so on. Although CSSD services are basically responsible for preventing SSI (eg, laparoscopic port infection, infection from implant procedure), which is an element of HAI.Reference Lobdell, Stamou and Sanchez7, Reference Sasmal, Mishra, Rath, Meher and Mohapatra8 However, minimizing SSI also depends on various factors, and sterilization monitoring is an important aspect of in these prevention efforts. The main difficulty in India is to the use of costly disposable medical devices (eg, cardiac catheter, Octopus, suction tube, robotic and laparoscopic instruments, dental and ophthalmic instruments), which drives up the procedure cost.9 Reprocessing medical devices and proper monitoring are required to minimize the risk of infection. To ensure sterility, a proper simulator that can validate the entire sterilization process is recommended. For monitoring proper steam penetration inside hollow or complex instruments (ie, luminal instruments), HPCD is clearly the best option.10

References

Basu, D, Bhattacharya, S, Mahajan, A, Ramanan, VR, Chandy, M. The importance of the central sterile supply department in infection prevention and control. Infect Control Hosp Epidemiol 2014;35:13121314.CrossRefGoogle ScholarPubMed
Basu, D, Bhattacharya, S, Mahajan, A, Ramanan, VR, Chandy, M. Sterilization indicators in central sterile supply department: quality assurance and cost implications. Infect Control Hosp Epidemiol 2015;36:484486.CrossRefGoogle ScholarPubMed
McDonnell, G, Burke, P. Disinfection: is it time to reconsider Spaulding? J Hosp Infect 2011;78:163170.CrossRefGoogle ScholarPubMed
Steri-record direct inoculation syringe to test sterility inside complex instruments. gke website. https://www.gke.eu/en/product-catalogues-and-data-sheets.html?file=files/gke_2013/content/en/downloads/Data%20sheets/Biological%20indicators/Direct%20Inoculation%20Syringe.pdf. Published 2013. Accessed April 18, 2019.Google Scholar
Current reference devices for hollow instruments loads as defined in standards are not a valid steam penetration test. gke website. http://www.gke.de/en/publications.html?file=files/gke_2013/content/en/downloads/Publications/Others/Current%20reference%20devices%20for%20hollow%20instruments%20loads.pdf. Published 2013. Accessed May 11, 2018.Google Scholar
Effectiveness of hospital central sterilization processing vs clinic-based sterilization protocols: a noninferiority trial. Dalhousie University Library website. https://dalspace.library.dal.ca/bitstream/handle/10222/54542/Stefanuto-Peter-MSc-OMFS-September-2014.pdf?sequence=5. Published 2014. Accessed May 11, 2018.Google Scholar
Lobdell, KW, Stamou, S, Sanchez, JA. Hospital-acquired infections. Surg Clin N Am 2012;92:6577.CrossRefGoogle ScholarPubMed
Sasmal, PK, Mishra, TS, Rath, S, Meher, S, Mohapatra, D. Port site infection in laparoscopic surgery: a review of its management. World J Clin Cases 2015;3:864871.CrossRefGoogle ScholarPubMed
Cost comparison of medical treatments: India vs the world. India Profile Medical Tourism website. http://www.indiaprofile.com/medical-tourism/cost-comparison.html. Accessed April 10, 2018.Google Scholar
Performance of hollow load process challenge devices (HLPCDs) for the determination of air removal and steam penetration in porous load steam sterilization processes. University of Glascow Enlighten website. http://eprints.gla.ac.uk/136299/1/136299.pdf. Accessed May 10, 2018.Google Scholar