Development of a Field Emission VP-SEM and Some Initial Applications

Published online by Cambridge University Press:  02 July 2020

Masako Nishimura
Affiliation:
Hitachi Science Systems, Ltd., 882 Ichige, Hitachinaka, Ibaraki, 312-8504, Japan
Sukehiro Itoh
Affiliation:
Hitachi Science Systems, Ltd., 882 Ichige, Hitachinaka, Ibaraki, 312-8504, Japan
Steve Joens
Affiliation:
Hitachi Science Systems, Ltd., 882 Ichige, Hitachinaka, Ibaraki, 312-8504, Japan
Get access

Abstract

The use of variable pressure SEMs (VP-SEMs) is increasing in various fields of science and industry, allowing microscopy in a variable pressure environment of 1 ∼ 270 Pa utilizing backscattered electrons for imaging. The VP-SEM allows microscopy of insulated samples without the need for sample preparation. Charging artifacts can be minimized as well. When the VP-SEM is operated with a cooling stage, which allows cooling of samples at −20° and above, vaporization of water from samples is reduced. This permits microscopy of wet samples at close to the natural state for extended periods of time.

Poor S/N ratio and deterioration of resolution, both of which are due to collisions among residual gas molecules and primary/backscattered electrons, have limited the performance of VP-SEMs. For resolving these limitations, we have completed the development of a new field emission VP-SEM which operates with a stable Schottky field emission source, a new environmental secondary electron detector (ESED), and a multi-stage differential pumping system. Fig. 1 shows a sectional view of the column with the differential pumping system. This design allows stable gun vacuum conditions with variable specimen chamber pressure 10 through 3,000 Pa, permitting a pressure difference from the gun by 1011 Pa without problems.

Type
Corporate Session (Organized by M. Kersker)
Copyright
Copyright © Microscopy Society of America 2001

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

references

1.Farely, A. N. and Shah, J. S. ; Instr. Phys. Conf. Ser. 93: 241 (1988).Google Scholar
2.Mohan, A.et al.; Scanning 20: 436 (1998).CrossRefGoogle Scholar
3.Ushiki, T.et al.; Proc. 55th Japan Electron Microscopy, 151 (1999).Google Scholar
4.Ito, S.et al.; Proc. 56lh Japan Electron Microscopy, 345 (2000).Google Scholar