Get access

Extract

As features on an IC chip become smaller than the resolution power of an optical microscope and of the size of the grinding particles, the trend for preparing cross-sectional transmission electron microscopy (TEM) samples at specific locations (bits) is moving towards using a focused ion beam (FIB) machine. Details on how to use a FIB machine to prepare cross-sectional TEM samples have been outlined in many references.The general procedure is to first mark the specific location (bit) in the FIB machine and then grind the sample down to about 20 microns, 10 microns on each side of the feature of interest. After grinding, the sample is mounted on a pre-cut TEM grid and thinned with the FIB to about 0.1 micron in the region containing the feature of interest. There are several disadvantages to this method. First, the sample goes into the FIB machine at least twice—once for FIB marks on the location and once again for the final thinning.

Type
Precision Specimen Preparation
Copyright
Copyright © Microscopy Society of America

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1)Young, R.J., et al., Mat. Res. Soc. Symp. Proc, Vol. 199, P205 (1990).CrossRefGoogle Scholar
2)Basile, D.P., et al., Mat. Res. Soc. Symp. Proc, Vol. 254, P23 (1992).CrossRefGoogle Scholar
3)Hull, R. and Dunn, D., Mat. Res. Soc. Symp. Proc, Vol. 523, P141 (1998).CrossRefGoogle Scholar
4)Benedict, J.P., et al., Mat. Res. Soc. Symp. Proc, Vol. 199, P189 (1990).CrossRefGoogle Scholar