Hostname: page-component-78c5997874-xbtfd Total loading time: 0 Render date: 2024-11-14T19:22:39.980Z Has data issue: false hasContentIssue false

Cross-Section Sample Preparation of a Free-Standing Thin-Film Coupon for Transmission Electron Microscopy Analysis

Published online by Cambridge University Press:  28 July 2005

John M. Phelps
Affiliation:
Materials Reliability Division, National Institute of Standards and Technology, 325 Broadway, Boulder, CO 80303
Get access

Abstract

A method for preparing a cross-section sample from a free-standing thin-film coupon suitable for transmission electron microscopy analysis has been successfully developed. The free-standing coupon presented a significant challenge for the preparation of a cross-section sample. A stack consisting of a silicon wafer, thin-film coupon, and glass coverslip was epoxied together, providing both support for thin film and enough surface area to handle the specimen during the sample preparation procedure. The tripod polisher wedge technique was used to thin the cross-section sample to 2 to 5 μm in thickness. This produced a sample with the required structural integrity. Final thinning was accomplished using an ion mill. This technique showed no artifacts introduced from the sample preparation procedure.

Type
Research Article
Copyright
© 2005 Microscopy Society of America

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)