Prospects for X-Ray Analysis of Rough Surfaces in the SEM
Published online by Cambridge University Press: 02 July 2020
Extract
Some of the key aspects affecting analysis of a rough surface are described in Fig. 1. If a hypothetical sample consisting of 96% Si, 1% O, 1% Na, 1% Al and 1% Fe is analysed at 20kV at constant beam current with an x-ray detector elevated at 30deg from horizontal then the measured intensities of the various characteristic lines will vary with tilt. While absorption of low energy x-rays is reduced as the sample tilts towards the detector, there is also a gradual loss of intensity due to increased backscatter. At the lower beam voltage of 5kV, absorption of x-rays at zero tilt is already small and Fig.2 shows that variation in intensity with positive tilts is substantially reduced, although absorption still becomes significant if the sample is tilted away from the detector.
- Type
- X-Ray Microanalysis Of Rough Surfaces
- Information
- Microscopy and Microanalysis , Volume 6 , Issue S2: Proceedings: Microscopy & Microanalysis 2000, Microscopy Society of America 58th Annual Meeting, Microbeam Analysis Society 34th Annual Meeting, Microscopical Society of Canada/Societe de Microscopie de Canada 27th Annual Meeting, Philadelphia, Pennsylvania August 13-17, 2000 , August 2000 , pp. 918 - 919