Internal Conversion in Energy Dispersive X-ray Analysis of Actinide-Containing Materials
Published online by Cambridge University Press: 09 May 2007
Abstract
The use of X-ray elemental analysis tools like energy dispersive X-ray (EDS) is described in the context of the investigation of nuclear materials. These materials contain radioactive elements, particularly alpha-decaying actinides that affect the quantitative EDS measurement by producing interferences in the X-ray spectra. These interferences originating from X-ray emission are the result of internal conversion by the daughter atoms from the alpha-decaying actinides. The strong interferences affect primarily the L X-ray lines from the actinides (in the typical energy range used for EDS analysis) and would require the use of the M lines. However, it is typically at the energy of the actinide's M lines that the interferences are dominant. The artifacts produced in the X-ray analysis are described and illustrated by some typical examples of analysis of actinide-bearing material.