Article contents
First Develoment Step and Test of The SESAM / SATEM
Published online by Cambridge University Press: 02 July 2020
Extract
For the SESAM/SATEM project is for testing and as a first development step a new fully corrected imaging energy filter of the Ω type built. It is incorporated in a 200keV EFTEM equipped with an LaB6 cathode and a objective lens with a spherical and chromatic aberration coefficient of Cs = Cc = 1.2mm. This EFTEM will be used this year by the SESAM project group at the University Tiibungen, Germany and the MPI fur Metallforschung, Stuttgart. At the same time this kind of filter will be also mounted in the final SATEM.
The energy filter is similar to the design of the so-called FHI-filter suggested by Rose and Lanio [1,2] and built at the Fritz-Haber-Institut of the MPG by Krahl [3]. The total length of the proposed filter is half of the FHI-filter and nearly the same as the length of the LEO 912Ω filter.
- Type
- Electron Energy-Loss Spectroscopy (EELS) and Imaging
- Information
- Microscopy and Microanalysis , Volume 6 , Issue S2: Proceedings: Microscopy & Microanalysis 2000, Microscopy Society of America 58th Annual Meeting, Microbeam Analysis Society 34th Annual Meeting, Microscopical Society of Canada/Societe de Microscopie de Canada 27th Annual Meeting, Philadelphia, Pennsylvania August 13-17, 2000 , August 2000 , pp. 198 - 199
- Copyright
- Copyright © Microscopy Society of America
References
[1] Rose, H., Optik 51 (1978) 15Google Scholar
[2] Tanio, S., Diss. TH Darmstadt D17 (1986)Google Scholar
[3] Lanio, S., Rose, H., Krahl, D., Optik 73 (1986) 56Google Scholar
[4] Uhlemann, S., Rose, H., Optik 96 (1994) 163Google Scholar
- 2
- Cited by